MANSBACH HEALTH TOOLS LLC
12815 River Rd, Potomac, Maryland, United States
FEI 3035111095
FDA has one inspection on file for MANSBACH HEALTH TOOLS LLC in Potomac, Maryland (FEI 3035111095). It ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 11 citations from this inspection. They are listed below. Three FOIA requests have asked FDA for this inspection's 483.
No warning letter, court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 1, in 2026
- Inspections with 483 observations shown
- 1 of 1
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 3 requests, latest Aug 21, 2026
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices11 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
- ISO 13485:2016 Clause 7.3.7
Design and development validation not complete prior to release of product
Design and development validation was not adequately completed prior to release for use of the product to the customer.
- ISO 13485:2016 Clause 8.2.4
Internal audits not conducted to ensure QMS meets requirements and is effective
Internal audits have not been adequately conducted at planned intervals to determine whether the QMS conforms to planned and documented arrangements and applicable requirements and is effectively implemented and maintained.
- ISO 13485:2016 Clause 7.3.10
Design and development files not maintained
A design and development file was not adequately maintained for each medical device type or medical device family.
- ISO 13485:2016 Clause 7.1
Required elements not determined in planning product realization
In product realization planning your firm did not adequately determine, as appropriate, records needed to provide evidence that the realization processes and resulting product meet requirements.
- ISO 13485:2016 Clause 4.1.2
Risk-based approach not applied to QMS processes
A risk-based approach to the control of the appropriate processes needed for the quality management system has not been applied.
Show 6 more citations
- ISO 13485:2016 Clause 7.4.1
Criteria for the evaluation and selection of suppliers not established
Criteria for the evaluation and selection of suppliers were not established.
- ISO 13485:2016 Clause 8.2.1
Feedback process procedures not documented
Procedures for the feedback process have not been documented.
Devices subject to device identification GUDID data submission requirements.
The information required by the Global Unique Device Identification Database for each medical device version or model required to bear a unique device identifier (UDI) was not provided.
- ISO 13485:2016 Clause 7.5.6
Procedures for validation of processes not documented
Procedures for validation of processes were not documented.
- ISO 13485:2016 Clause 7.5.1
Production controls not adequate
Production controls do not include documentation of procedures and methods for the control of production.
- ISO 13485:2016 Clause 5.6.2
Input to management review is incomplete
Management review input does not adequately include information arising from feedback, reporting to regulatory authorities and follow-up actions from previous reviews.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Three requests to FDA under the Freedom of Information Act named this site, most recently on Aug 21, 2026. Each one asked for a Form 483.
Inspection ending Jun 2, 2026VAI
3 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Jul 30, 2026. FDAgent holds no copy that FDA posted.
- Aug 21, 2026FDA-FOIA-2026-8129Form 483· 1 of 4 sitesRecords released
- Aug 10, 2026FDA-FOIA-2026-7654Form 483· 1 of 5 sitesRecords released
- Jul 16, 2026FDA-FOIA-2026-6771Form 483Records released
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 1 section of 21 CFR in FDA’s published citations for this facility’s 1 inspection.
- 21 CFR 830.300Devices subject to device identification data submission requirements1 of 1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3035111095 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3035111095. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.