ECLOSION
1029 Jonathan Ave, Fairlawn, Ohio, United States
FEI 3030198312
FDA has one inspection on file for ECLOSION in Fairlawn, Ohio (FEI 3030198312). It ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 10 citations from this inspection. They are listed below. Three FOIA requests have asked FDA for this inspection's 483.
No warning letter, court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 1, in 2026
- Inspections with 483 observations shown
- 1 of 1
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 3 requests, latest Sep 10, 2026
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices10 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
- ISO 13485:2016 Clause 8.3.3
Records not maintained for nonconforming product actions detected after delivery
Records were not maintained of actions taken to address nonconforming product detected after delivery or use.
- ISO 13485:2016 Clause 7.3.10
Design and development files not maintained
A design and development file was not maintained for each medical device type or medical device family.
Devices subject to device identification GUDID data submission requirements.
The information required by the Global Unique Device Identification Database for each medical device version or model required to bear a unique device identifier (UDI) was not provided.
- ISO 13485:2016 Clause 7.4.1
Procedures for purchasing process not documented
Procedures to ensure that purchased product conforms to specified purchasing information have not been documented.
- ISO 13485:2016 Clause 7.1
Processes needed for product realization not planned and developed
Processes needed for product realization were not planned and developed.
Show 5 more citations
Lack of Written MDR Procedures
Written MDR procedures have not been developed, maintained and implemented.
- ISO 13485:2016 Clause 8.5.2
Corrective action procedures not documented
Corrective action procedures defining the requirements for reviewing nonconformities, determining the causes of nonconformities, evaluating the need for action to ensure that nonconformities do not recur, planning and documenting action needed and implementing such action, verifying that the corrective action does not adversely affect the ability to meet applicable regulatory requirements or the safety and performance of the medical device and reviewing the effectiveness of corrective action taken have not been documented.
- ISO 13485:2016 Clause 8.5.3
Preventive action procedures not documented
Preventive action procedures describing the requirements for determining potential nonconformities and their causes, evaluating the need for action to prevent occurrence of potential nonconformities, planning and documenting action needed and implementing such action, verifying that the action does not adversely affect the ability to meet applicable regulatory requirements or the safety and performance of the medical device and reviewing the effectiveness of the preventive action taken have not been documented.
- ISO 13485:2016 Clause 7.4.3
Purchased product verification activities not established or implemented
The inspection or other activities necessary for ensuring that purchased product meets specified purchasing requirements were not established and implemented.
- ISO 13485:2016 Clause 4.2.4
Lack of document control
Documents required by the quality management system have not been controlled.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Three requests to FDA under the Freedom of Information Act named this site, most recently on Sep 10, 2026. Each one asked for a Form 483.
Inspection ending Aug 13, 2026VAI
3 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Oct 7, 2026. FDAgent holds no copy that FDA posted.
- Sep 10, 2026FDA-FOIA-2026-8792Form 483Records released
- Sep 4, 2026FDA-FOIA-2026-8623Form 483· 1 of 5 sitesOpen
- Sep 4, 2026FDA-FOIA-2026-8616Form 483· 1 of 3 sitesRecords released
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 2 sections of 21 CFR in FDA’s published citations for this facility’s 1 inspection.
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 1
- 21 CFR 830.300Devices subject to device identification data submission requirements1 of 1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3030198312 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3030198312. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.