Marlen Manufacturing & Development Co.

5150 Richmond Rd, Bedford, Ohio, United States

FEI 1520020

FDA has 5 inspections on file for Marlen Manufacturing & Development Co. in Bedford, Ohio (FEI 1520020), from 2009 to 2026. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 4 citations from this inspection. They are listed below. Two FOIA requests have asked FDA for this inspection's 483.

The most recent OAI inspection ended on Jul 8, 2009. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.

Six regulations were cited at more than one inspection; they are listed below.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
5, 2009–2026
1 OAI · 2 VAI · 2 NAI
Inspections with 483 observations shown
5 of 5
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
2 requests, latest Jul 10, 2026
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

5 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. ISO 13485:2016 Clause 7.1

      Product realization risk management processes not documented

      One or more processes for risk management in product realization have not been adequately documented.

    2. Devices subject to device identification GUDID data submission requirements.

      The information required by the Global Unique Device Identification Database for each medical device version or model required to bear a unique device identifier (UDI) was not provided.

    3. ISO 13485:2016 Clause 7.5.1

      Production controls not adequate

      Production controls do not include documentation of procedures and methods for the control of production, implementation of monitoring and measurement of process parameters and product characteristics and implementation of defined operations for labeling and packaging.

    4. ISO 13485:2016 Clause 7.5.6

      Procedures for validation of processes not documented

      Procedures for validation of processes were not adequately documented.

  2. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.

    2. Calibration, Inspection, etc. Procedures Lack of or Inadequ

      Procedures to ensure equipment is routinely calibrated have not been adequately established.

  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate procedures for labeling

      Procedures to control labeling activities have not been adequately established.

    2. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    3. Calibration, Inspection, etc. Procedures Lack of or Inadequ

      Procedures to ensure equipment is routinely calibrated, inspected and maintained have not been adequately established.

    4. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been adequately established to control product that does not conform to specified requirements.

    5. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.

    Show 1 more citation
    1. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

  4. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been established.

    2. Documentation

      Process validation activities and results have not been adequately documented.

  5. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices11 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Quantity manufactured

      The device history record does not include the quantity of devices manufactured.

    2. Production monitoring and control

      During production, process parameters are not fully monitored and controlled.

    3. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been validated and approved according to established procedures.

    4. Maintained

      The acceptance status of product was not clearly identified throughout labeling of the product.

    5. Complete files maintained

      Complete complaint files are not maintained.

    Show 6 more citations
    1. Calibration, Inspection, etc. Procedures Lack of or Inadequ

      Procedures to ensure that equipment is routinely calibrated were not implemented.

    2. Lack of Written MDR Procedures

      Written MDR procedures have not been developed.

    3. Incoming inspection, testing, verification

      Incoming product was not adequately inspected or tested to verify conformance to specifications.

    4. Design changes - Lack of or Inadequate Procedures

      Procedures were not established for the documentation of design changes before their implementation.

    5. Management review - Lack of or inadequate procedures

      Procedures for management review are not implemented.

    6. Nonconforming product, Lack of or inadequate procedures

      Procedures were not followed for the control of products that do not conform to specifications.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Two requests to FDA under the Freedom of Information Act named this site, most recently on Jul 10, 2026. Each one asked for a Form 483.

  • Inspection ending May 6, 2026VAI

    2 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Sep 2, 2026. FDAgent holds no copy that FDA posted.

    1. Jul 10, 2026FDA-FOIA-2026-6611Form 483· 1 of 5 sitesRecords released
    2. Jun 2, 2026FDA-FOIA-2026-5475Form 483Records released

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Regulations FDA cited here

The 15 sections of 21 CFR in FDA’s published citations for this facility’s 5 inspections, with how many of them cited each.

Show all 15

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 1520020 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 1520020. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.