21 CFR 803.56: If I am a manufacturer, in what circumstances must I submit a supplemental or followup report and what are the requirements for such reports?
FDA's inspection data lists 58 Form 483 citations of 21 CFR 803.56 at 50 facilities, from November 2009 to May 8, 2025.
Counts: FDA Data Dashboard citations, last synced Nov 23, 2026, 04:25 UTC, and FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Regulation text: eCFR. FDAgent is independent of FDA.
- Form 483 citations
- 58
- At 58 inspections
- Facilities cited
- 50
- Distinct FEIs
- Most recent citation
- May 8, 2025
- Sol-Millennium Medical Inc.
- Warning letters
- 0
- State a violation of it
The regulation
The current text of 21 CFR 803.56 from the Electronic Code of Federal Regulations.
§ 803.56 If I am a manufacturer, in what circumstances must I submit a supplemental or followup report and what are the requirements for such reports?
If you are a manufacturer, when you obtain information required under this part that you did not provide because it was not known or was not available when you submitted the initial report, you must submit the supplemental information to us within 30 calendar days of the day that you receive this information. You must submit the supplemental or followup report in accordance with the requirements of § 803.12(a). On a supplemental or followup report, you must:
(a) Indicate that the report being submitted is a supplemental or followup report;
(b) Submit the appropriate identification numbers of the report that you are updating with the supplemental information (e.g., your original manufacturer report number and the user facility or importer report number of any report on which your report was based), if applicable; and
(c) Include only the new, changed, or corrected information.
21 CFR Part 803: Medical Device Reporting, Subpart ERead on eCFR
Citations by year
Form 483 citations of 21 CFR 803.56 by the year the inspection ended. FDA's citation data starts with inspections ending in October 2008, and the current year runs to the latest sync, so both are partial years (lighter bars).
Peak: 10 in 2012
| Year | Form 483 citations |
|---|---|
| 2009 | 2 |
| 2010 | 9 |
| 2011 | 8 |
| 2012 | 10 |
| 2013 | 8 |
| 2014 | 2 |
| 2015 | 4 |
| 2016 | 3 |
| 2017 | 2 |
| 2018 | 2 |
| 2020 | 1 |
| 2021 | 2 |
| 2022 | 1 |
| 2023 | 2 |
| 2024 | 1 |
| 2025 | 1 |
How FDA words this citation
FDA's short descriptions and citation text on these 58 citations, most used first. These are FDA's standard summaries in its citation data, not the text of any one Form 483.
Short descriptions
- Submission of Supplement to Individual Report Within One Month58 citations
Citation text
- “A supplemental report was not submitted to FDA within one month following receipt of information that was not provided when the initial report was submitted.”51 citations
- “An individual supplemental report was not submitted to FDA within one month following receipt of information that was not provided when the initial report was submitted.”7 citations
Facilities cited most recently
The latest inspections where FDA cited 21 CFR 803.56, by inspection end date, with FDA's short description. Each links to the facility's FDA record on FDAgent.
- Sol-Millennium Medical Inc.2025-05-08 · FEI 3014312726Submission of Supplement to Individual Report Within One Month
- Maquet Cardiovascular, LLC2024-07-11 · FEI 2242352Submission of Supplement to Individual Report Within One Month
- Datascope Corp.2023-06-15 · FEI 3007965489Submission of Supplement to Individual Report Within One Month
- Sakura Seiki Co., Ltd. Chikuma Factory2023-03-24 · FEI 3009685834Submission of Supplement to Individual Report Within One Month
- Datascope Corp.2022-01-21 · FEI 3007965489Submission of Supplement to Individual Report Within One Month
- Philips North America LLC2021-08-20 · FEI 1000524572Submission of Supplement to Individual Report Within One Month
Show 4 more
- Maquet Cardiovascular, LLC2021-02-26 · FEI 2242352Submission of Supplement to Individual Report Within One Month
- Olympus Corporation of the Americas2020-01-17 · FEI 1000635022Submission of Supplement to Individual Report Within One Month
- Sharps Compliance, Inc.2018-11-15 · FEI 3007508552Submission of Supplement to Individual Report Within One Month
- American Medical Systems Innovations2018-02-27 · FEI 2937094Submission of Supplement to Individual Report Within One Month
Often cited at the same inspection
Of the 58 inspections where FDA cited 21 CFR 803.56, the share where it also cited each of these sections.
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?33 (57%)
- 21 CFR 820.198Complaint files31 (53%)
- 21 CFR 820.100Corrective and preventive action30 (52%)
- 21 CFR 820.30Design controls22 (38%)
- 21 CFR 820.90Nonconforming product16 (28%)
- 21 CFR 820.75Process validation14 (24%)
Ask FDAgent about 21 CFR 803.56
FDAgent reads FDA's Form 483 citations, observations and warning letters and answers in plain English, with links to the records.
Related
Topics that count this section:Device CAPA and complaints
About this data
- Citation counts come from FDA's Data Dashboard, which lists the regulation behind each Form 483 observation for inspections since October 2008.
- Observation excerpts come from Form 483 PDFs in FDA's electronic reading room, tied to this section only by exact wording as described above.
- A warning letter counts for this section when one of its numbered violation sentences cites 21 CFR 803.56; FDAgent quotes that sentence exactly.
- FDAgent rebuilds this page every week from FDA's public data and is independent of FDA.