Skip to content

21 CFR 803.56: If I am a manufacturer, in what circumstances must I submit a supplemental or followup report and what are the requirements for such reports?

FDA's inspection data lists 58 Form 483 citations of 21 CFR 803.56 at 50 facilities, from November 2009 to May 8, 2025.

Counts: FDA Data Dashboard citations, last synced Nov 23, 2026, 04:25 UTC, and FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Regulation text: eCFR. FDAgent is independent of FDA.

Form 483 citations
58
At 58 inspections
Facilities cited
50
Distinct FEIs
Most recent citation
May 8, 2025
Sol-Millennium Medical Inc.
Warning letters
0
State a violation of it

The regulation

The current text of 21 CFR 803.56 from the Electronic Code of Federal Regulations.

§ 803.56 If I am a manufacturer, in what circumstances must I submit a supplemental or followup report and what are the requirements for such reports?

If you are a manufacturer, when you obtain information required under this part that you did not provide because it was not known or was not available when you submitted the initial report, you must submit the supplemental information to us within 30 calendar days of the day that you receive this information. You must submit the supplemental or followup report in accordance with the requirements of § 803.12(a). On a supplemental or followup report, you must:

(a) Indicate that the report being submitted is a supplemental or followup report;

(b) Submit the appropriate identification numbers of the report that you are updating with the supplemental information (e.g., your original manufacturer report number and the user facility or importer report number of any report on which your report was based), if applicable; and

(c) Include only the new, changed, or corrected information.

21 CFR Part 803: Medical Device Reporting, Subpart ERead on eCFR

Citations by year

Form 483 citations of 21 CFR 803.56 by the year the inspection ended. FDA's citation data starts with inspections ending in October 2008, and the current year runs to the latest sync, so both are partial years (lighter bars).

Peak: 10 in 2012

Form 483 citations of 21 CFR 803.56 by inspection year
YearForm 483 citations
20092
20109
20118
201210
20138
20142
20154
20163
20172
20182
20201
20212
20221
20232
20241
20251

How FDA words this citation

FDA's short descriptions and citation text on these 58 citations, most used first. These are FDA's standard summaries in its citation data, not the text of any one Form 483.

Short descriptions

  • Submission of Supplement to Individual Report Within One Month58 citations

Citation text

  • “A supplemental report was not submitted to FDA within one month following receipt of information that was not provided when the initial report was submitted.”51 citations
  • “An individual supplemental report was not submitted to FDA within one month following receipt of information that was not provided when the initial report was submitted.”7 citations

Facilities cited most recently

The latest inspections where FDA cited 21 CFR 803.56, by inspection end date, with FDA's short description. Each links to the facility's FDA record on FDAgent.

  • Sol-Millennium Medical Inc.2025-05-08 · FEI 3014312726Submission of Supplement to Individual Report Within One Month
  • Maquet Cardiovascular, LLC2024-07-11 · FEI 2242352Submission of Supplement to Individual Report Within One Month
  • Datascope Corp.2023-06-15 · FEI 3007965489Submission of Supplement to Individual Report Within One Month
  • Sakura Seiki Co., Ltd. Chikuma Factory2023-03-24 · FEI 3009685834Submission of Supplement to Individual Report Within One Month
  • Datascope Corp.2022-01-21 · FEI 3007965489Submission of Supplement to Individual Report Within One Month
  • Philips North America LLC2021-08-20 · FEI 1000524572Submission of Supplement to Individual Report Within One Month
Show 4 more

Often cited at the same inspection

Of the 58 inspections where FDA cited 21 CFR 803.56, the share where it also cited each of these sections.

Ask FDAgent about 21 CFR 803.56

FDAgent reads FDA's Form 483 citations, observations and warning letters and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

Topics that count this section:Device CAPA and complaints

About this data

  • Citation counts come from FDA's Data Dashboard, which lists the regulation behind each Form 483 observation for inspections since October 2008.
  • Observation excerpts come from Form 483 PDFs in FDA's electronic reading room, tied to this section only by exact wording as described above.
  • A warning letter counts for this section when one of its numbered violation sentences cites 21 CFR 803.56; FDAgent quotes that sentence exactly.
  • FDAgent rebuilds this page every week from FDA's public data and is independent of FDA.