Skip to content

Datascope Corp.

45 Barbour Pond Dr, Wayne, New Jersey, United States

FEI 3007965489

FDA has 2 inspections on file for Datascope Corp. in Wayne, New Jersey (FEI 3007965489), from 2022 to 2023. The latest ended on and was classified OAI, official action indicated.

OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.

FDA published 6 citations from this inspection: 4 under 21 CFR Part 820, the device quality system rules, and 2 under 21 CFR Part 803, the device adverse event reporting rules. They are listed below. One of the two FOIA requests for this site asked FDA for the 483 from a later inspection, ending Aug 21, 2025, that FDA's Data Dashboard doesn't list yet.

FDA's compliance record also lists one other action under this FEI: Injunction, . No warning letter or import refusal is filed under this FEI.

Three regulations were cited at more than one inspection: 21 CFR 803.50(a)(2), 21 CFR 803.56 and 21 CFR 820.100(a).

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
OAI
Inspections on file
2, 2022–2023
2 OAI · 0 VAI · 0 NAI
Inspections with 483 observations shown
2 of 2
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
2 requests, latest Feb 27, 2026
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Statistical techniques - Lack of or inadequate procedures

      Procedures for identifying valid statistical techniques required for establishing, controlling, and verifying the acceptability of process capability and product characteristics have not been adequately established.

    2. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    3. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been adequately established to control product that does not conform to specified requirements.

    4. Lack of or inadequate receiving acceptance procedures

      Procedures for acceptance of incoming product have not been adequately established.

    5. Submission of Supplement to Individual Report Within One Month

      An individual supplemental report was not submitted to FDA within one month following receipt of information that was not provided when the initial report was submitted.

    Show 1 more citation
    1. Individual Report of Malfunction

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

  2. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices13 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate procedures - Acceptance activities

      Procedures for acceptance activities have not been established.

    2. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been established.

    3. Report of Death or Serious Injury

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.

    4. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been established.

    5. Design Validation - Risk analysis not performed/inadequate

      Risk analysis was not performed.

    Show 8 more citations
    1. Investigation of device failures

      Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.

    2. Calibration, Inspection, etc. Procedures Lack of or Inadequ

      Procedures to ensure equipment is routinely calibrated and maintained have not been adequately established.

    3. Individual Report of Malfunction

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

    4. Design verification - Lack of or inadequate procedures

      Procedures for design verification have not been established.

    5. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.

    6. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been established.

    7. Submission of Supplement to Individual Report Within One Month

      An individual supplemental report was not submitted to FDA within one month following receipt of information that was not provided when the initial report was submitted.

    8. Design validation- Lack of or inadequate procedures

      Procedures for design validation have not been established.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Two requests to FDA under the Freedom of Information Act named this site, most recently on Feb 27, 2026. Each one asked for a Form 483.

  • Inspection ending Aug 21, 2025

    1 request. Not listed in FDA’s Data Dashboard yet. FDA released records in answer to a request for its Form 483 on Mar 23, 2026. FDAgent holds no copy that FDA posted.

    1. Jan 2, 2026FDA-FOIA-2026-30Form 483· 1 of 4 sitesRecords released
  • Inspection ending May 15, 2023

    1 request. Not listed in FDA’s Data Dashboard yet. FDA released records in answer to a request for its Form 483 on Apr 7, 2026. FDAgent holds no copy that FDA posted.

    1. Feb 27, 2026FDA-FOIA-2026-2026Form 483Records released

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Regulations FDA cited here

The 10 sections of 21 CFR in FDA’s published citations for this facility’s 2 inspections, with how many of them cited each.

  • 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?2 of 2
  • 21 CFR 803.56If I am a manufacturer, in what circumstances must I submit a supplemental or followup report and what are the requirements for such reports?2 of 2
  • 21 CFR 820.80Receiving, in-process, and finished device acceptance2 of 2
  • 21 CFR 820.100Corrective and preventive action2 of 2
  • 21 CFR 820.30Design controls1 of 2
  • 21 CFR 820.50Purchasing controls1 of 2
  • 21 CFR 820.72Inspection, measuring, and test equipment1 of 2
  • 21 CFR 820.90Nonconforming product1 of 2
Show all 10

Warning letters and compliance actions

FDA Data Dashboard compliance actions

Other FEIs recorded as Datascope Corp.

Matched on the exact firm name FDA recorded. Sites recorded under other names are not listed, and a shared name does not show common ownership.

  • Datascope Corp.FEI 3001418283Mahwah, New Jersey, United States

    Last inspected 2018-10-03NAI

  • Datascope Corp.FEI 1000222374Fairfield, New Jersey, United States

    Last inspected 2018-06-11OAI

  • Datascope Corp.FEI 3009540546Dayton, New Jersey, United States

    Last inspected 2013-10-17OAI

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3007965489 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3007965489. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.