Datascope Corp.
1300 Macarthur Blvd, Mahwah, New Jersey, United States
FEI 3001418283
FDA has 3 inspections on file for Datascope Corp. in Mahwah, New Jersey (FEI 3001418283), from 2010 to 2018. The latest ended on and was classified NAI, no action indicated. FDA issued a warning letter on (CDRH).
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
FDA published 9 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.
No court action or import refusal is filed under this FEI.
Two regulations were cited at more than one inspection: 21 CFR 820.100(a) and 21 CFR 820.50(a)(1).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 3, 2010–2018
- 0 OAI · 1 VAI · 2 NAI
- Inspections with 483 observations shown
- 2 of 3
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices9 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Product rework procedures, Lack of or inadequate procedures
Procedures for rework of nonconforming product have not been adequately established.
Design verification - output does not meet input requirement
Design verification does not confirm that design output meets design input requirements.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Design validation- Lack of or inadequate procedures
Procedures for design validation have not been established.
Show 4 more citations
DHR content
The device history record does not demonstrate that the device was manufactured in accordance with the device master record.
Documentation
Corrective and preventive action activities and/or results have not been documented.
Document review, approval by designated individual
Documents were not reviewed and not approved by designated individual(s) prior to issuance .
Evaluation and Selection, Suppliers, Contractors, etc.
Potential suppliers were not evaluated and selected based on their ability to meet specified requirements.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures
The procedures for implementing corrective and preventive actions were not complete.
Evaluation and Selection, Suppliers, Contractors, etc.
Potential suppliers were not sufficiently evaluated and selected on the basis of their ability to meet specified requirements.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
Regulations FDA cited here
The 6 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 3 inspections, with how many of them cited each.
- 21 CFR 820.50Purchasing controls2 of 2
- 21 CFR 820.100Corrective and preventive action2 of 2
- 21 CFR 820.30Design controls1 of 2
- 21 CFR 820.40[Reserved]1 of 2
- 21 CFR 820.90Nonconforming product1 of 2
- 21 CFR 820.184Device history record1 of 2
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Other FEIs recorded as Datascope Corp.
Matched on the exact firm name FDA recorded. Sites recorded under other names are not listed, and a shared name does not show common ownership.
Datascope Corp.FEI 3007965489Wayne, New Jersey, United States
Last inspected 2023-06-15OAI
Datascope Corp.FEI 1000222374Fairfield, New Jersey, United States
Last inspected 2018-06-11OAI
Datascope Corp.FEI 3009540546Dayton, New Jersey, United States
Last inspected 2013-10-17OAI
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3001418283 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3001418283. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.