Philips North America LLC
22100 Bothell Everett Hwy, Bothell, Washington, United States
FEI 1000524572
FDA has 10 inspections on file for Philips North America LLC in Bothell, Washington (FEI 1000524572), from 2008 to 2022. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 3 citations from this inspection: 2 under 21 CFR Part 803, the device adverse event reporting rules, and 1 under 21 CFR Part 820, the device quality system rules. They are listed below. Five of the six FOIA requests for this site asked FDA for 483s from three later inspections that FDA's Data Dashboard doesn't list yet, the latest ending Nov 21, 2025.
The most recent OAI inspection ended on Dec 3, 2018. FDA issued a warning letter on (CDRH). FDA's compliance record also lists 2 other actions under this FEI, the latest Injunction, . 7 import refusals are filed under this FEI, the latest on .
Four regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 10, 2008–2022
- 3 OAI · 5 VAI · 2 NAI
- Inspections with 483 observations shown
- 8 of 10
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 6 requests, latest Jan 8, 2026
- Warning letters
- · CDRH ↗
- Import refusals
- 7 refusals
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
10 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Lack of System for Event Evaluations
The written MDR Procedure does not include an internal system which provides for timely and effective identification and evaluation of events that may be subject to medical device reporting requirements.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Procedures for product review,disposition lack of/inadequate
Procedures that define the responsibility for review and the authority for the disposition of nonconforming product have not been adequately established.
Submission of Supplement to Individual Report Within One Month
An individual supplemental report was not submitted to FDA within one month following receipt of information that was not provided when the initial report was submitted.
Periodic equipment inspection lack of or inadequate procedu
Procedures for conducting periodic inspections to ensure adherence to equipment maintenance schedules have not been established.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: DevicesCitations not published
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices8 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Process changes - review, evaluation and revalidation
A validated process was not reviewed and evaluated and revalidated when changes or process deviations occurred.
Software validation for automated processes
Software used as part of the quality system has not been validated for its intended use according to an established protocol.
Design validation- Lack of or inadequate procedures
Procedures for design validation have not been adequately established.
Lack of System for Timely Submission of Reports
The written MDR procedure does not include an internal system which provides for timely transmission of complete medical device reports to FDA.
Show 3 more citations
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
Design verification - Lack of or inadequate procedures
Procedures for design verification have not been adequately established.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Evaluation and Selection, Suppliers, Contractors, etc.
Potential suppliers were not evaluated and selected based on their ability to meet specified requirements.
Lack of or inadequate procedures - Acceptance activities
Procedures for acceptance activities have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Acceptable supplier records, inadequate records
Records of acceptable suppliers, contractors, and consultants have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Incorrect translation to production specifications
The device design was not correctly translated into production specifications.
Evaluation, timeliness, identification
Complaints representing events that are MDR reportable were not promptly reviewed, evaluated, and investigated by a designated individual.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Bioresearch Monitoring1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Sponsors' general responsibilities
For an investigational study, proper monitoring was not ensured.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices11 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Quality Audit/Reaudit - conducted
Quality audits were not conducted at sufficient regular intervals, as prescribed by internal procedures to verify that the quality system is effective in fulfilling your quality system objectives.
Software validation for automated processes
Software used as part of the quality system has not been adequately validated for its intended use according to an established protocol.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Show 6 more citations
Type of Reportable Event
An individual medical device manufacturer report submitted per FDA Form 3500A did not indicate in Block H the type of reportable event.
Statistical techniques - Lack of or inadequate procedures
Procedures for identifying valid statistical techniques required for establishing, controlling, and verifying the acceptability of process capability and product characteristics have not been adequately established.
Records of complaint investigation
Records of complaint investigations do not include required information.
Approval, inadequate purchasing data
Purchasing data that clearly describe or reference specified requirements for purchased or otherwise received product and services have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Not established, maintained
Written procedures for the collection, maintenance, and auditing of device tracking information were not established.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Six requests to FDA under the Freedom of Information Act named this site, most recently on Jan 8, 2026.
Inspection ending Nov 21, 2025
1 request. Not listed in FDA’s Data Dashboard yet. FDA released records in answer to a request for its Form 483 on Mar 11, 2026. FDAgent holds no copy that FDA posted.
- Jan 8, 2026FDA-FOIA-2026-267Form 483· 1 of 5 sitesRecords released
Inspection ending Nov 6, 2024
3 requests. Not listed in FDA’s Data Dashboard yet. None of FDA’s log entries for these requests say records were released.
- Jun 25, 2025FDA-FOIA-2025-8008Form 483· 1 of 2 sitesClosed over fees
- Mar 14, 2025FDA-FOIA-2025-2371Form 483· 1 of 4 sitesWithdrawn
- Dec 20, 2024FDA-FOIA-2024-11208Form 483· 1 of 2 sitesClosed
Inspection ending Nov 3, 2023
1 request. Not listed in FDA’s Data Dashboard yet. None of FDA’s log entries for these requests say records were released.
- Jan 5, 2024FDA-FOIA-2024-172Form 483Closed
Inspection date not given
1 request. None of FDA’s log entries for these requests say records were released.
- Dec 3, 2020FDA-FOIA-2020-8444EIR· 1 of 2 sitesWithdrawn
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 16 sections of 21 CFR in FDA’s published citations for 8 of this facility’s 10 inspections, with how many of them cited each.
- 21 CFR 820.100Corrective and preventive action6 of 8
- 21 CFR 820.50Purchasing controls3 of 8
- 21 CFR 820.70Production and process controls3 of 8
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?2 of 8
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?2 of 8
- 21 CFR 820.30Design controls2 of 8
- 21 CFR 820.198Complaint files2 of 8
- 21 CFR 803.52If I am a manufacturer, what information must I submit in my individual adverse event reports?1 of 8
Show all 16
- 21 CFR 803.56If I am a manufacturer, in what circumstances must I submit a supplemental or followup report and what are the requirements for such reports?1 of 8
- 21 CFR 812.40General responsibilities of sponsors1 of 8
- 21 CFR 820.22Quality audit1 of 8
- 21 CFR 820.75Process validation1 of 8
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 8
- 21 CFR 820.90Nonconforming product1 of 8
- 21 CFR 820.250Statistical techniques1 of 8
- 21 CFR 821.25Device tracking system and content requirements: manufacturer requirements1 of 8
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- InjunctionCDRH · DevicesFDA record
- InjunctionCDRH · DevicesFDA record
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Import refusals
FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 1000524572, one row per refused product line.
7 refusals from 2016-10-25 to 2023-04-24 · charged FRNMFGREG on 6, NOT LISTED on 5 · no FDA sample analysis
FDA firm profile- 2023-04-24
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)74MKJ
NOT LISTED 502(o), 801(a)(3); MISBRANDING
NO PMA 501(f)(1)(B), 801(a)(3); ADULTERATION
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
EXPIRED 501(c); 801(a)(3) Adulteration
Division of Southeast Imports799-4257990-5/11/1
- 2020-10-08
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)74MKJ
NOT LISTED 502(o), 801(a)(3); MISBRANDING
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
NO 510(K) 801(a)(3); 502(o) Misbranding
Division of Southeast Imports799-8850087-0/11/1
- 2020-07-31
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)74MKJ
NOT LISTED 502(o), 801(a)(3); MISBRANDING
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
Division of Southeast ImportsSCS-1232716-1/11/1
- 2020-01-15
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)74MKJ
NOT LISTED 502(o), 801(a)(3); MISBRANDING
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
REGISTERED 502(o), 801(a)(3); MISBRANDING
Division of Southeast Imports799-4461541-6/11/1
- 2019-02-01
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)74MKJ
NOT LISTED 502(o), 801(a)(3); MISBRANDING
Division of Southeast Imports799-8490279-8/11/1A
Show 2 more
- 2017-08-01
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)74MKJ
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
Division of Southeast Imports799-1458176-1/11/1
- 2016-10-25
SYSTEM, NETWORK AND COMMUNICATION, PHYSIOLOGICAL MONITORS74MSX
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
REGISTERED 502(o), 801(a)(3); MISBRANDING
Division of Southeast Imports799-7726510-2/1/1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1000524572 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1000524572. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.