21 CFR 821.25: Device tracking system and content requirements: manufacturer requirements
FDA's inspection data lists 16 Form 483 citations of 21 CFR 821.25 at 8 facilities, from August 2009 to January 29, 2016.
Counts: FDA Data Dashboard citations, last synced Nov 23, 2026, 04:25 UTC, and FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Regulation text: eCFR. FDAgent is independent of FDA.
- Form 483 citations
- 16
- At 12 inspections
- Facilities cited
- 8
- Distinct FEIs
- Most recent citation
- Jan 29, 2016
- Curonix LLC
- Warning letters
- 0
- State a violation of it
The regulation
The current text of 21 CFR 821.25 from the Electronic Code of Federal Regulations.
§ 821.25 Device tracking system and content requirements: manufacturer requirements.
(a) A manufacturer of a tracked device shall adopt a method of tracking for each such type of device that it distributes that enables a manufacturer to provide FDA with the following information in writing for each tracked device distributed:
(1) Except as required by order under section 518(e) of the act, within 3 working days of a request from FDA, prior to the distribution of a tracked device to a patient, the name, address, and telephone number of the distributor, multiple distributor, or final distributor holding the device for distribution and the location of the device;
(2) Within 10 working days of a request from FDA for tracked devices that are intended for use by a single patient over the life of the device, after distribution to or implantation in a patient:
(i) The unique device identifier (UDI), lot number, batch number, model number, or serial number of the device or other identifier necessary to provide for effective tracking of the devices;
(ii) The date the device was shipped by the manufacturer;
(iii) The name, address, telephone number, and social security number (if available) of the patient receiving the device, unless not released by the patient under § 821.55(a);
(iv) The date the device was provided to the patient;
(v) The name, mailing address, and telephone number of the prescribing physician;
(vi) The name, mailing address, and telephone number of the physician regularly following the patient if different than the prescribing physician; and
(vii) If applicable, the date the device was explanted and the name, mailing address, and telephone number of the explanting physician; the date of the patient's death; or the date the device was returned to the manufacturer, permanently retired from use, or otherwise permanently disposed of.
(3) Except as required by order under section 518(e) of the act, within 10 working days of a request from FDA for tracked devices that are intended for use by more than one patient, after the distribution of the device to the multiple distributor:
(i) The unique device identifier (UDI), lot number, batch number, model number, or serial number of the device or other identifier necessary to provide for effective tracking of the devices;
(ii) The date the device was shipped by the manufacturer;
(iii) The name, address, and telephone number of the multiple distributor;
(iv) The name, address, telephone number, and social security number (if available) of the patient using the device, unless not released by the patient under § 821.55(a);
(v) The location of the device;
(vi) The date the device was provided for use by the patient;
(vii) The name, address, and telephone number of the prescribing physician; and
(viii) If and when applicable, the date the device was returned to the manufacturer, permanently retired from use, or otherwise permanently disposed of.
(b) A manufacturer of a tracked device shall keep current records in accordance with its standard operating procedure of the information identified in paragraphs (a)(1), (a)(2) and (a)(3)(i) through (a)(3)(iii) of this section on each tracked device released for distribution for as long as such device is in use or in distribution for use.
(c) A manufacturer of a tracked device shall establish a written standard operating procedure for the collection, maintenance, and auditing of the data specified in paragraphs (a) and (b) of this section. A manufacturer shall make this standard operating procedure available to FDA upon request. A manufacturer shall incorporate the following into the standard operating procedure:
(1) Data collection and recording procedures, which shall include a procedure for recording when data which is required under this part is missing and could not be collected and the reason why such required data is missing and could not be collected;
(2) A method for recording all modifications or changes to the tracking system or to the data collected and maintained under the tracking system, reasons for any modification or change, and dates of any modification or change. Modification and changes included under this requirement include modifications to the data (including termination of tracking), the data format, the recording system, and the file maintenance procedures system; and
(3) A quality assurance program that includes an audit procedure to be run for each device product subject to tracking, at not less than 6-month intervals for the first 3 years of distribution and at least once a year thereafter. This audit procedure shall provide for statistically relevant sampling of the data collected to ensure the accuracy of data and performance testing of the functioning of the tracking system.
(d) When a manufacturer becomes aware that a distributor, final distributor, or multiple distributor has not collected, maintained, or furnished any record or information required by this part, the manufacturer shall notify the FDA district office responsible for the area in which the distributor, final distributor, or multiple distributor is located of the failure of such persons to comply with the requirements of this part. Manufacturers shall have taken reasonable steps to obtain compliance by the distributor, multiple distributor, or final distributor in question before notifying FDA.
(e) A manufacturer may petition for an exemption or variance from one or more requirements of this part according to the procedures in § 821.2 of this chapter.
21 CFR Part 821: Medical Device Tracking Requirements, Subpart BRead on eCFR
Citations by year
Form 483 citations of 21 CFR 821.25 by the year the inspection ended. FDA's citation data starts with inspections ending in October 2008, and the current year runs to the latest sync, so both are partial years (lighter bars).
Peak: 5 in 2011
| Year | Form 483 citations |
|---|---|
| 2009 | 3 |
| 2010 | 2 |
| 2011 | 5 |
| 2012 | 2 |
| 2013 | 1 |
| 2014 | 2 |
| 2016 | 1 |
How FDA words this citation
FDA's short descriptions and citation text on these 16 citations, most used first. These are FDA's standard summaries in its citation data, not the text of any one Form 483.
Short descriptions
- Quality assurance procedures: 6-month & annual audits6 citations
- Not established, maintained4 citations
- Records not current2 citations
- Written tracking program2 citations
- Missing data1 citations
Show 1 more
- Notification of FDA re: non-compliant distributor1 citations
Citation text
- “A quality assurance program that includes audit procedures for each device subject to tracking and audits performed at 6-month intervals for the first 3 years of distribution and annual audits thereafter using a statistically relevant sampling has not been followed.”3 citations
- “Written procedures for the collection, maintenance, and auditing of device tracking information were not established .”3 citations
- “A quality assurance program that includes audit procedures for each device subject to tracking and audits performed at 6-month intervals for the first 3 years of distribution and annual audits thereafter using a statistically relevant sampling is not complete.”2 citations
- “A quality assurance program that includes audit procedures for each device subject to tracking and audits performed at 6-month intervals for the first 3 years of distribution and annual audits thereafter using a statistically relevant sampling has not been established.”1 citations
Show 2 more
- “An adequate written device tracking program was not established and implemented for each device subject to a tracking order.”1 citations
- “An adequate written device tracking program was not implemented for each device subject to a tracking order.”1 citations
Facilities cited most recently
The latest inspections where FDA cited 21 CFR 821.25, by inspection end date, with FDA's short description. Each links to the facility's FDA record on FDAgent.
Show 4 more
- ZOLL Medical Corporation2011-03-18 · FEI 1220908Quality assurance procedures: 6-month & annual audits
Often cited at the same inspection
Of the 12 inspections where FDA cited 21 CFR 821.25, the share where it also cited each of these sections.
- 21 CFR 820.100Corrective and preventive action8 (67%)
- 21 CFR 820.30Design controls6 (50%)
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?4 (33%)
- 21 CFR 820.70Production and process controls4 (33%)
- 21 CFR 820.198Complaint files3 (25%)
- 21 CFR 803.52If I am a manufacturer, what information must I submit in my individual adverse event reports?2 (17%)
Ask FDAgent about 21 CFR 821.25
FDAgent reads FDA's Form 483 citations, observations and warning letters and answers in plain English, with links to the records.
Related
About this data
- Citation counts come from FDA's Data Dashboard, which lists the regulation behind each Form 483 observation for inspections since October 2008.
- Observation excerpts come from Form 483 PDFs in FDA's electronic reading room, tied to this section only by exact wording as described above.
- A warning letter counts for this section when one of its numbered violation sentences cites 21 CFR 821.25; FDAgent quotes that sentence exactly.
- FDAgent rebuilds this page every week from FDA's public data and is independent of FDA.