Heartsine Technologies Inc
121 Friends Ln Ste 400, Newtown, Pennsylvania, United States
FEI 3004018643
FDA has 5 inspections on file for Heartsine Technologies Inc in Newtown, Pennsylvania (FEI 3004018643), from 2009 to 2018. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
The most recent OAI inspection ended on Aug 22, 2013. FDA has issued 2 warning letters under this FEI, the latest on (CDRH). 1 import refusal is filed under this FEI, on . No court action is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 5, 2009–2018
- 1 OAI · 1 VAI · 3 NAI
- Inspections with 483 observations shown
- 3 of 5
- Form 483s posted by FDA
- None held for this FEI
- Import refusals
- 1 refusal
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
5 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Project Evaluation: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: DevicesCitations not published
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Software validation for automated processes
Software used as part of the quality system has not been validated for its intended use according to an established protocol.
Management review - Lack of or inadequate procedures
Procedures for management review have not been adequately established.
Design validation - user needs and intended uses
Design validation did not ensure the device conforms to defined user needs and intended uses.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Not established, maintained
Written procedures for the collection, maintenance, and auditing of device tracking information were not implemented and complete.
Show 1 more citation
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Approval, inadequate purchasing data
Purchasing information that clearly includes/describes the specified requirements for purchased or otherwise received products and services was not complete.
- NAINo objectionable conditions that need follow-upDevices · Compliance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Quality assurance procedures: 6-month & annual audits
A quality assurance program that includes audit procedures for each device subject to tracking and audits performed at 6-month intervals for the first 3 years of distribution and annual audits thereafter using a statistically relevant sampling has not been followed.
General
Complaint handling procedures for receiving complaints have not been defined.
Missing data
Procedures for explaining why device tracking information is missing have not been established.
Records not current
Current records for each tracked device while in distribution for use were not maintained.
Regulations FDA cited here
The 8 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 5 inspections, with how many of them cited each.
- 21 CFR 821.25Device tracking system and content requirements: manufacturer requirements2 of 3
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?1 of 3
- 21 CFR 820.20Management responsibility1 of 3
- 21 CFR 820.30Design controls1 of 3
- 21 CFR 820.50Purchasing controls1 of 3
- 21 CFR 820.70Production and process controls1 of 3
- 21 CFR 820.100Corrective and preventive action1 of 3
- 21 CFR 820.198Complaint files1 of 3
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched some of these, so the link opens FDA’s record of the action.
Import refusals
FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 3004018643, one row per refused product line.
1 refusal on 2020-11-06 · charged NOT LISTED on 1, NO REGISTR on 1 · no FDA sample analysis
FDA firm profile- 2020-11-06
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)74MKJ
NOT LISTED 502(o), 801(a)(3); MISBRANDING
NO REGISTR 536(a); Failure to file initial report
Division of Northern Border Imports799-9236757-1/11/1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3004018643 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3004018643. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.