ZOLL Medical Corporation
269 Mill Rd, Chelmsford, Massachusetts, United States
FEI 1220908
FDA has 15 inspections on file for ZOLL Medical Corporation in Chelmsford, Massachusetts (FEI 1220908), from 2009 to 2025. The latest ended on and was classified OAI, official action indicated. FDA has issued 3 warning letters under this FEI, the latest on (CDRH).
OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.
FDA published 5 citations from this inspection: 4 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 803, the device adverse event reporting rules. They are listed below. Two FOIA requests have asked FDA for this inspection's 483.
7 import refusals are filed under this FEI, the latest on . No court action is filed under this FEI.
10 regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Warning letter ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- OAI
- Inspections on file
- 15, 2009–2025
- 4 OAI · 5 VAI · 6 NAI
- Inspections with 483 observations shown
- 9 of 15
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 2 requests, latest Oct 15, 2025
- Import refusals
- 7 refusals
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
15 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design validation- Lack of or inadequate procedures
Procedures for design validation have not been adequately established.
Evaluation of suppliers, contractors, etc., requirements
Requirements that must be met by suppliers have not been established.
Design Validation - Risk analysis not performed/inadequate
Risk analysis is inadequate.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Project Evaluation: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Bioresearch MonitoringCitations not published
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Production and Process Change Procedures, lack of or Inad.
Procedures for changes to a specification, method, process, and procedure have not been adequately established.
Design output - Lack of or inadequate procedures
Procedures for design output have not been adequately established.
Lack of or inadequate receiving acceptance procedures
Procedures for acceptance of incoming product have not been adequately established.
Supplier notification of changes
There is no agreement with suppliers to notify you of changes in the product or service.
Show 2 more citations
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been adequately established.
Investigation of device failures
Complaints involving the possible failure of a device, labeling, and packaging to meet any of its specifications were not investigated where necessary.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Training records
Personnel training is not documented.
Notification of FDA re: non-compliant distributor
FDA was not notified, nor were reasonable steps taken concerning a distributor who has not collected, maintained, or furnished any record or information required by the device tracking regulations.
Design validation - user needs and intended uses
Design validation did not ensure the device conforms to defined user needs and intended uses.
Lack of or inadequate receiving acceptance procedures
Procedures for acceptance of incoming product have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Bioresearch Monitoring1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Sponsor device disposition records inadequate
Records of disposition of a device which describe the batch number or code marks of any devices returned to the sponsor, repaired, and disposed of by the investigator or another person and the reasons for and method of disposition reports are not all accurate and complete.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Records not current
Current records for each tracked device while in distribution for use were not complete and up-to-date.
Design history file
The design history file does not demonstrate that the design was developed following the requirements of 21 CFR 820.
Quality assurance procedures: 6-month & annual audits
A quality assurance program that includes audit procedures for each device subject to tracking and audits performed at 6-month intervals for the first 3 years of distribution and annual audits thereafter using a statistically relevant sampling has not been established.
Production and Process Change Procedures, lack of or Inad.
Procedures for changes to a procedure have not been adequately established.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Quality assurance procedures: 6-month & annual audits
A quality assurance program that includes audit procedures for each device subject to tracking and audits performed at 6-month intervals for the first 3 years of distribution and annual audits thereafter using a statistically relevant sampling has not been followed.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Written tracking program
An adequate written device tracking program was not implemented for each device subject to a tracking order.
MDR not evaluated per 820.198
A submitted MDR event was not evaluated in accordance with the requirements of the regulations regarding Complaint Files.
Design validation - user needs and intended uses
Design validation did not ensure the device conforms to defined user needs and intended uses.
Show 2 more citations
Nonconforming product control
Products that do not conform to specifications are not adequately controlled.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures
The procedures for implementing corrective and preventive actions were not complete.
Lack of System for Event Evaluations
The written MDR Procedure does not include an internal system which provides for the timely and effective identification of events that may be subject to medical device reporting requirements.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.
Essential design outputs
Design outputs that are essential for the proper functioning of the device were not identified.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Training - Lack of or inadequate procedures
Procedures for identifying training needs were not implemented.
Investigation procedures
The corrective and preventive action procedures addressing the investigation of the cause of nonconformities relating to product, processes, and the quality system were not implemented.
Lack of System for Event Evaluations
The written MDR Procedure does not include an internal system which provides for the timely and effective evaluation of events that may be subject to medical device reporting requirements.
Design validation - user needs and intended uses
Design validation did not ensure that devices conform to defined user/patient needs and intended uses.
Supplier notification of changes
There is no clear agreement from the suppliers that they will notify you of changes in the product or service.
Show 1 more citation
Implementing changes
Corrective and preventive action procedures addressing implementation and recording of changes in methods and procedures to correct and prevent identified quality problems were not documented, complete, and implemented.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Two requests to FDA under the Freedom of Information Act named this site, most recently on Oct 15, 2025. Each one asked for a Form 483.
Inspection ending Apr 15, 2025OAI
2 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Jul 11, 2025. FDAgent holds no copy that FDA posted.
- Oct 15, 2025FDA-FOIA-2025-12015Form 483· 1 of 5 sitesOpen
- May 15, 2025FDA-FOIA-2025-4437Form 483· 1 of 5 sitesRecords released
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 13 sections of 21 CFR in FDA’s published citations for 9 of this facility’s 15 inspections, with how many of them cited each.
- 21 CFR 820.30Design controls7 of 9
- 21 CFR 820.100Corrective and preventive action6 of 9
- 21 CFR 820.25Personnel4 of 9
- 21 CFR 820.50Purchasing controls3 of 9
- 21 CFR 820.198Complaint files3 of 9
- 21 CFR 821.25Device tracking system and content requirements: manufacturer requirements3 of 9
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?2 of 9
- 21 CFR 820.70Production and process controls2 of 9
Show all 13
- 21 CFR 820.80Receiving, in-process, and finished device acceptance2 of 9
- 21 CFR 803.18What are the requirements for establishing and maintaining MDR files or records that apply to me?1 of 9
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?1 of 9
- 21 CFR 812.140Records1 of 9
- 21 CFR 820.90Nonconforming product1 of 9
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesLetter on fda.govFDA record
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched some of these, so the link opens FDA’s record of the action.
Import refusals
FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 1220908, one row per refused product line.
7 refusals from 2018-02-26 to 2026-05-15 · charged FRNMFGREG on 5, NOT LISTED on 5 · no FDA sample analysis
FDA firm profile- 2026-05-15
DC-DEFIBRILLATOR, LOW-ENERGY, (INCLUDING PADDLES)74LDD
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
REGISTERED 502(o), 801(a)(3); MISBRANDING
Division of Southeast ImportsAEK-0440165-8/10001/1
- 2024-11-13
VENTILATOR, EMERGENCY, POWERED (RESUSCITATOR)73BTL
NOT LISTED 502(o), 801(a)(3); MISBRANDING
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
Division of Southwest ImportsMC2-8016312-2/11/1
- 2021-10-26
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)74MKJ
NOT LISTED 502(o), 801(a)(3); MISBRANDING
NO PMA/PDP 501(f)(1)(A); 801(a)(3); ADULTERATION
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
Division of Southeast Imports799-5298553-8/11/1
- 2021-03-02
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)74MKJ
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
Division of Southeast ImportsMC2-6830930-9/41/1
- 2020-01-22
VENTILATOR, CONTINUOUS (RESPIRATOR)73CBK
NOT LISTED 502(o), 801(a)(3); MISBRANDING
NO 510(K) 801(a)(3); 502(o) Misbranding
Division of West Coast Imports534-1017536-7/11/1
Show 2 more
- 2019-01-02
VENTILATOR, CONTINUOUS (RESPIRATOR)73CBK
NOT LISTED 502(o), 801(a)(3); MISBRANDING
REGISTERED 502(o), 801(a)(3); MISBRANDING
Division of West Coast Imports534-1016342-1/11/1
- 2018-02-26
MONITOR, CARDIAC (INCL. CARDIOTACHOMETER & RATE ALARM)74DRT
NOT LISTED 502(o), 801(a)(3); MISBRANDING
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
Division of Southeast ImportsUPS-7397863-8/11/1
Other FEIs recorded as ZOLL Medical Corporation
Matched on the exact firm name FDA recorded. Sites recorded under other names are not listed, and a shared name does not show common ownership.
ZOLL Medical CorporationFEI 1218058Pawtucket, Rhode Island, United States
Last inspected 2022-03-10NAI
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1220908 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1220908. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.