ZOLL Medical Corporation

269 Mill Rd, Chelmsford, Massachusetts, United States

FEI 1220908

FDA has 15 inspections on file for ZOLL Medical Corporation in Chelmsford, Massachusetts (FEI 1220908), from 2009 to 2025. The latest ended on and was classified OAI, official action indicated. FDA has issued 3 warning letters under this FEI, the latest on (CDRH).

OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.

FDA published 5 citations from this inspection: 4 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 803, the device adverse event reporting rules. They are listed below. Two FOIA requests have asked FDA for this inspection's 483.

7 import refusals are filed under this FEI, the latest on . No court action is filed under this FEI.

10 regulations were cited at more than one inspection; they are listed below.

Sources:FDA firm profile ↗Warning letter ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
OAI
Inspections on file
15, 2009–2025
4 OAI · 5 VAI · 6 NAI
Inspections with 483 observations shown
9 of 15
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
2 requests, latest Oct 15, 2025
Warning letters
· CDRH ↗
3 on file
Import refusals
7 refusals
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

15 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design validation- Lack of or inadequate procedures

      Procedures for design validation have not been adequately established.

    2. Evaluation of suppliers, contractors, etc., requirements

      Requirements that must be met by suppliers have not been established.

    3. Design Validation - Risk analysis not performed/inadequate

      Risk analysis is inadequate.

    4. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    5. Individual Report of Malfunction

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
  3. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  4. NAINo objectionable conditions that need follow-upDevices · Project Evaluation: DevicesNone
  5. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  6. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices1 citation published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.

  7. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Bioresearch MonitoringCitations not published
  8. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices7 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    2. Production and Process Change Procedures, lack of or Inad.

      Procedures for changes to a specification, method, process, and procedure have not been adequately established.

    3. Design output - Lack of or inadequate procedures

      Procedures for design output have not been adequately established.

    4. Lack of or inadequate receiving acceptance procedures

      Procedures for acceptance of incoming product have not been adequately established.

    5. Supplier notification of changes

      There is no agreement with suppliers to notify you of changes in the product or service.

    Show 2 more citations
    1. Training - Lack of or inadequate procedures

      Procedures for training and identifying training needs have not been adequately established.

    2. Investigation of device failures

      Complaints involving the possible failure of a device, labeling, and packaging to meet any of its specifications were not investigated where necessary.

  9. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Training records

      Personnel training is not documented.

    2. Notification of FDA re: non-compliant distributor

      FDA was not notified, nor were reasonable steps taken concerning a distributor who has not collected, maintained, or furnished any record or information required by the device tracking regulations.

    3. Design validation - user needs and intended uses

      Design validation did not ensure the device conforms to defined user needs and intended uses.

    4. Lack of or inadequate receiving acceptance procedures

      Procedures for acceptance of incoming product have not been adequately established.

  10. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Bioresearch Monitoring1 citation published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Sponsor device disposition records inadequate

      Records of disposition of a device which describe the batch number or code marks of any devices returned to the sponsor, repaired, and disposed of by the investigator or another person and the reasons for and method of disposition reports are not all accurate and complete.

  11. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    2. Records not current

      Current records for each tracked device while in distribution for use were not complete and up-to-date.

    3. Design history file

      The design history file does not demonstrate that the design was developed following the requirements of 21 CFR 820.

    4. Quality assurance procedures: 6-month & annual audits

      A quality assurance program that includes audit procedures for each device subject to tracking and audits performed at 6-month intervals for the first 3 years of distribution and annual audits thereafter using a statistically relevant sampling has not been established.

    5. Production and Process Change Procedures, lack of or Inad.

      Procedures for changes to a procedure have not been adequately established.

  12. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices7 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Quality assurance procedures: 6-month & annual audits

      A quality assurance program that includes audit procedures for each device subject to tracking and audits performed at 6-month intervals for the first 3 years of distribution and annual audits thereafter using a statistically relevant sampling has not been followed.

    2. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    3. Written tracking program

      An adequate written device tracking program was not implemented for each device subject to a tracking order.

    4. MDR not evaluated per 820.198

      A submitted MDR event was not evaluated in accordance with the requirements of the regulations regarding Complaint Files.

    5. Design validation - user needs and intended uses

      Design validation did not ensure the device conforms to defined user needs and intended uses.

    Show 2 more citations
    1. Nonconforming product control

      Products that do not conform to specifications are not adequately controlled.

    2. Training - Lack of or inadequate procedures

      Procedures for training and identifying training needs have not been adequately established.

  13. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  14. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate procedures

      The procedures for implementing corrective and preventive actions were not complete.

    2. Lack of System for Event Evaluations

      The written MDR Procedure does not include an internal system which provides for the timely and effective identification of events that may be subject to medical device reporting requirements.

    3. Investigation of device failures

      Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.

    4. Essential design outputs

      Design outputs that are essential for the proper functioning of the device were not identified.

  15. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Training - Lack of or inadequate procedures

      Procedures for identifying training needs were not implemented.

    2. Investigation procedures

      The corrective and preventive action procedures addressing the investigation of the cause of nonconformities relating to product, processes, and the quality system were not implemented.

    3. Lack of System for Event Evaluations

      The written MDR Procedure does not include an internal system which provides for the timely and effective evaluation of events that may be subject to medical device reporting requirements.

    4. Design validation - user needs and intended uses

      Design validation did not ensure that devices conform to defined user/patient needs and intended uses.

    5. Supplier notification of changes

      There is no clear agreement from the suppliers that they will notify you of changes in the product or service.

    Show 1 more citation
    1. Implementing changes

      Corrective and preventive action procedures addressing implementation and recording of changes in methods and procedures to correct and prevent identified quality problems were not documented, complete, and implemented.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Two requests to FDA under the Freedom of Information Act named this site, most recently on Oct 15, 2025. Each one asked for a Form 483.

  • Inspection ending Apr 15, 2025OAI

    2 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Jul 11, 2025. FDAgent holds no copy that FDA posted.

    1. Oct 15, 2025FDA-FOIA-2025-12015Form 483· 1 of 5 sitesOpen
    2. May 15, 2025FDA-FOIA-2025-4437Form 483· 1 of 5 sitesRecords released

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Regulations FDA cited here

The 13 sections of 21 CFR in FDA’s published citations for 9 of this facility’s 15 inspections, with how many of them cited each.

Show all 13
  • 21 CFR 820.80Receiving, in-process, and finished device acceptance2 of 9
  • 21 CFR 803.18What are the requirements for establishing and maintaining MDR files or records that apply to me?1 of 9
  • 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?1 of 9
  • 21 CFR 812.140Records1 of 9
  • 21 CFR 820.90Nonconforming product1 of 9

Warning letters and compliance actions

FDA Data Dashboard compliance actions

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched some of these, so the link opens FDA’s record of the action.

Import refusals

FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 1220908, one row per refused product line.

7 refusals from 2018-02-26 to 2026-05-15 · charged FRNMFGREG on 5, NOT LISTED on 5 · no FDA sample analysis

FDA firm profile
  • 2026-05-15

    DC-DEFIBRILLATOR, LOW-ENERGY, (INCLUDING PADDLES)74LDD

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast ImportsAEK-0440165-8/10001/1

  • 2024-11-13

    VENTILATOR, EMERGENCY, POWERED (RESUSCITATOR)73BTL

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    Division of Southwest ImportsMC2-8016312-2/11/1

  • 2021-10-26

    AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)74MKJ

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    NO PMA/PDP 501(f)(1)(A); 801(a)(3); ADULTERATION

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-5298553-8/11/1

  • 2021-03-02

    AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)74MKJ

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    Division of Southeast ImportsMC2-6830930-9/41/1

  • 2020-01-22

    VENTILATOR, CONTINUOUS (RESPIRATOR)73CBK

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    NO 510(K) 801(a)(3); 502(o) Misbranding

    Division of West Coast Imports534-1017536-7/11/1

Show 2 more
  • 2019-01-02

    VENTILATOR, CONTINUOUS (RESPIRATOR)73CBK

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    Division of West Coast Imports534-1016342-1/11/1

  • 2018-02-26

    MONITOR, CARDIAC (INCL. CARDIOTACHOMETER & RATE ALARM)74DRT

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    Division of Southeast ImportsUPS-7397863-8/11/1

Other FEIs recorded as ZOLL Medical Corporation

Matched on the exact firm name FDA recorded. Sites recorded under other names are not listed, and a shared name does not show common ownership.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 1220908 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 1220908. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.