21 CFR 803.18: What are the requirements for establishing and maintaining MDR files or records that apply to me?
FDA's inspection data lists 64 Form 483 citations of 21 CFR 803.18 at 55 facilities, from November 2008 to July 31, 2025. 4 warning letters state a violation of it.
Counts: FDA Data Dashboard citations, last synced Nov 23, 2026, 04:25 UTC, and FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Regulation text: eCFR. FDAgent is independent of FDA.
- Form 483 citations
- 64
- At 58 inspections
- Facilities cited
- 55
- Distinct FEIs
- Most recent citation
- Jul 31, 2025
- Euromi S.A.
- Warning letters
- 4
- State a violation of it
The regulation
The current text of 21 CFR 803.18 from the Electronic Code of Federal Regulations.
§ 803.18 What are the requirements for establishing and maintaining MDR files or records that apply to me?
(a) If you are a user facility, importer, or manufacturer, you must establish and maintain MDR event files. You must clearly identify all MDR event files and maintain them to facilitate timely access.
(b)(1) For purposes of this part, “MDR event files” are written or electronic files maintained by user facilities, importers, and manufacturers. MDR event files may incorporate references to other information (e.g., medical records, patient files, engineering reports), in lieu of copying and maintaining duplicates in this file. Your MDR event files must contain:
(i) Information in your possession or references to information related to the adverse event, including all documentation of your deliberations and decision making processes used to determine if a device-related death, serious injury, or malfunction was or was not reportable under this part;
(ii) Copies of all reports submitted under this part (whether paper or electronic), and of all other information related to the event that you submitted to us or other entities such as an importer, distributor, or manufacturer; and
(iii) Copies of all electronic acknowledgments FDA sends you in response to electronic MDR submissions.
(2) If you are a user facility, importer, or manufacturer, you must permit any authorized FDA employee, at all reasonable times, to access, to copy, and to verify the records required by this part.
(c) If you are a user facility, you must retain an MDR event file relating to an adverse event for a period of 2 years from the date of the event. If you are a manufacturer or importer, you must retain an MDR event file relating to an adverse event for a period of 2 years from the date of the event or a period of time equivalent to the expected life of the device, whichever is greater. If the device is no longer distributed, you still must maintain MDR event files for the time periods described in this paragraph (c).
(d)(1) If you are a device distributor, you must establish and maintain device complaint records (files). Your records must contain any incident information, including any written, electronic, or oral communication, either received or generated by you, that alleges deficiencies related to the identity (e.g., labeling), quality, durability, reliability, safety, effectiveness, or performance of a device. You must also maintain information about your evaluation of the allegations, if any, in the incident record. You must clearly identify the records as device incident records and file these records by device name. You may maintain these records in written or electronic format. You must back up any file maintained in electronic format.
(2) You must retain copies of the required device incident records for a period of 2 years from the date of inclusion of the record in the file or for a period of time equivalent to the expected life of the device, whichever is greater. You must maintain copies of these records for this period even if you no longer distribute the device.
(3) You must maintain the device complaint files established under this section at your principal business establishment. If you are also a manufacturer, you may maintain the file at the same location as you maintain your complaint file under part 820 of this chapter. You must permit any authorized FDA employee, at all reasonable times, to access, to copy, and to verify the records required by this part.
(e) If you are a manufacturer, you may maintain MDR event files as part of your complaint file, under part 820 of this chapter, if you prominently identify these records as MDR reportable events. We will not consider your submitted MDR report to comply with this part unless you evaluate an event in accordance with the quality management system requirements described in part 820 of this chapter. You must document and maintain in your MDR event files an explanation of why you did not submit or could not obtain any information required by this part, as well as the results of your evaluation of each event.
21 CFR Part 803: Medical Device Reporting, Subpart ARead on eCFR
Citations by year
Form 483 citations of 21 CFR 803.18 by the year the inspection ended. FDA's citation data starts with inspections ending in October 2008, and the current year runs to the latest sync, so both are partial years (lighter bars).
Peak: 10 in 2009
| Year | Form 483 citations |
|---|---|
| 2008 | 1 |
| 2009 | 10 |
| 2010 | 8 |
| 2011 | 9 |
| 2012 | 4 |
| 2013 | 5 |
| 2014 | 1 |
| 2015 | 7 |
| 2016 | 9 |
| 2018 | 1 |
| 2021 | 2 |
| 2023 | 2 |
| 2024 | 4 |
| 2025 | 1 |
How FDA words this citation
FDA's short descriptions and citation text on these 64 citations, most used first. These are FDA's standard summaries in its citation data, not the text of any one Form 483.
Short descriptions
- Adverse events--all info not in file18 citations
- Event files -- failure to establish and maintain12 citations
- Distributor responsibilities for complaint records8 citations
- MDR not evaluated per 820.1987 citations
- Results of evaluation not in mfr's MDR file6 citations
Show 3 more
- Files do not contain copies of MDR reports4 citations
- Distributor's device incident records3 citations
- Info not obtained or submitted2 citations
Citation text
- “MDR event files do not contain or reference all adverse event information in the possession of the reporting entity, including documentation of the deliberations and decision making process used to determine if an event was or was not reportable.”18 citations
- “MDR event files have not been established and maintained.”12 citations
- “Device complaint records are not established and maintained by the device distributor, including any relevant incident information and information regarding the evaluation of the allegations.”6 citations
- “The results of the evaluation of each event are not documented and maintained in the manufacturer's MDR event file.”6 citations
Show 2 more
- “A submitted MDR event was not evaluated in accordance with the requirements of the regulations regarding Complaint Files.”3 citations
- “MDR event files do not contain copies of all MDR forms and other information related to the event that was submitted to FDA.”3 citations
Facilities cited most recently
The latest inspections where FDA cited 21 CFR 803.18, by inspection end date, with FDA's short description. Each links to the facility's FDA record on FDAgent.
- Knox Medical Diagnostics, Inc.2024-08-21 · FEI 3016721526Results of evaluation not in mfr's MDR file
- Globus Medical, Inc.2024-03-07 · FEI 3004142400Files do not include FDA's Acknowledgement of electronic MDR submissions
Show 4 more
What the observations say
Excerpts from Form 483s that FDA posted in its reading room. A Form 483 usually doesn't name the regulation, so FDAgent ties an observation to 21 CFR 803.18 only when it opens with one of FDA's standard citation texts for this section, word for word apart from case, spacing and punctuation. 1 observation ties this way. The highlighted sentence is the matched wording.
Indiana University Hospital2016-02-03 · Observation 2
MDR event files have not been established and maintained. Specifically, your facility has not established and maintained MDR event files to include: 1. Copies or references for: a. All information that is reasonably known to your facility regarding potentially reportable adverse events. b. All information that is utilized in the evaluation of potentially reportable adverse events. 2. The decisions for whether or not an adverse …
This observation opens with FDA's standard wording for 21 CFR 803.18(a).
Warning letters stating a violation of 21 CFR 803.18
4 warning letters FDA posted state a violation of this section in a numbered violation sentence. The most recent are below, each linking to FDAgent's page for the letter.
“Your organization’s MDR event files do not contain all documentation of its deliberations and decision-making processes used to determine if a device-related death, serious injury, or malfunction was or was not reportable under 21 CFR Part 803, as required under 21 CFR 803.18(b)(1)(i).”
- O.N. Diagnostics, LLCIssued 2026-09-01
“Your procedure does not describe how your firm will address documentation and record keeping requirements, as required by 21 CFR 803.17(b), including: a. Documentation of adverse event related information maintained as MDR event files. b. Documentation of the deliberations and decision-making processes used to determine if a device-related death, serious injury, or malfunction was or was not reportable, as required under 21 CFR 803.18(b)(1)(i).”
- Robbins Instruments, LLCIssued 2025-01-21
“The procedure does not describe how your firm will address documentation and recordkeeping requirements, as required by 21 CFR 803.17(b), including documentation of the deliberations and decision-making processes used to determine if a device-related death, serious injury, or malfunction was or was not reportable, as required under 21 CFR 803.18(b)(1)(i).”
- Rontis Hellas S.A.Issued 2024-10-28
“Failure to maintain copies of all electronic acknowledgments FDA sends you in response to electronic MDR submissions in your MDR event file, as required by 21 CFR 803.18(b)(1)(iii).”
- Globus Medical, Inc.Issued 2024-07-15
Often cited at the same inspection
Of the 58 inspections where FDA cited 21 CFR 803.18, the share where it also cited each of these sections.
- 21 CFR 820.100Corrective and preventive action24 (41%)
- 21 CFR 820.198Complaint files24 (41%)
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?22 (38%)
- 21 CFR 803.30If I am a user facility, what reporting requirements apply to me?10 (17%)
- 21 CFR 820.30Design controls10 (17%)
- 21 CFR 820.70Production and process controls8 (14%)
Ask FDAgent about 21 CFR 803.18
FDAgent reads FDA's Form 483 citations, observations and warning letters and answers in plain English, with links to the records.
Related
Topics that count this section:Device CAPA and complaints
About this data
- Citation counts come from FDA's Data Dashboard, which lists the regulation behind each Form 483 observation for inspections since October 2008.
- Observation excerpts come from Form 483 PDFs in FDA's electronic reading room, tied to this section only by exact wording as described above.
- A warning letter counts for this section when one of its numbered violation sentences cites 21 CFR 803.18; FDAgent quotes that sentence exactly.
- FDAgent rebuilds this page every week from FDA's public data and is independent of FDA.