FEI 3004492321

FDA has one inspection on file for Indiana University Hospital in Indianapolis, Indiana (FEI 3004492321). It ended on and was classified VAI, voluntary action indicated. FDA has also posted a Form 483 dated Oct 25, 2016 that FDAgent has not matched to an inspection on the Data Dashboard.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA posted the Form 483 from that inspection. FDAgent read its 2 observations from the PDF; they are listed below.

No warning letter, court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
1, in 2016
Inspections with 483 observations shown
1 of 1
1 posted 483 not matched to an inspection
Form 483s posted by FDA
↗
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC · FDA reading room, last synced Nov 23, 2026, 04:33 UTC

  1. Not classified yetFDA posted this Form 483; FDAgent has not matched it to a Data Dashboard inspectionNo inspection matchedForm 483 pagePDF
  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices2 citations published

    Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent’s.

    1. Observation 1

      Inadequate MDR procedures and reporting

      OBSERVATION 1 The written MDR Procedure does not include an internal system which provides for the timely and effective identification, communication and evaluation of events that may be subject to medical device reporting requirements. Specifically, your facility's Safe Medical Devices, policy #EC 4.04 approved 4/30/2013, Patient Incident and Significant Event Management, policy #ADM 1.51 approved 5/31/2012, and Sentinel and Adverse Event Management, policy #RM 1.01 approved 2/28/2014, procedures are inadequate in that: Adverse events involving the use of a medical device that may have caused or contributed to the 1. death of a patient were not reported to the FDA and the device manufacturer via a MedWatch 3500A form within ten (10) working days of being identified. On [illegible] your facility initiated a retrospective review of cases involving the use of b(3) [illegible] and subsequently identified that two (2) patients had [illegible] infection that resulted in complications which may have potentially led to the deaths of those patients. Your facility could not provide documentation showing that Medical Device Reports were submitted to the FDA and device manufacturer for these incidents. 2. Adverse events involving the use of a medical device that may have caused or contributed to the serious injury of a patient were not reported to the FDA via a MedWatch 3500A form within ten (10) working days of being identified. Your firm utilizes the NCC MERP index for Categorizing Medication Errors to determine when events involving patient harm occur. The index classifies Categories E through H as resulting in patient harm and, in the timeframe from 11/1/2014 to 12/15/2015, your facility has documented twelve (12) incidents that were classified as Category E, indicating "an error occurred that may have contributed to or resulted in temporary harm to the patient and required intervention." Your facility has no documentation of Medical Device Reports being submitted to the device manufacturers for these incidents. 3. The procedures do not include provisions for: a. Reporting adverse events via the mandatory Medwatch 3500A form to the FDA or device manufacturers. During the inspection, your organization's personnel stated that your policy is to report all adverse events via the voluntary reporting Medwatch portal. However, your facility has not maintained documentation that would serve as objective evidence of this activity. b. Including required information that is reasonably known to your facility in a Medical Device Report to the FDA or device manufacturer. This information would include: Patient information, adverse event information, device information, initial reporter information, and user facility information.

    2. Observation 2

      Missing MDR event files and records

      OBSERVATION 2 MDR event files have not been established and maintained. Specifically, your facility has not established and maintained MDR event files to include: 1. Copies or references for: a. All information that is reasonably known to your facility regarding potentially reportable adverse events. b. All information that is utilized in the evaluation of potentially reportable adverse events. 2. The decisions for whether or not an adverse event must be reported to the FDA or medical device manufacturer via Medwatch Form 3500A.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3004492321 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3004492321. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.