FDA has 2 inspections on file for Philosys Co. Ltd in Gunsan, Korea (the Republic of) (FEI 3010172486), from 2018 to 2023. The latest ended on and was classified OAI, official action indicated.

OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.

FDA published 2 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.

FDA issued a warning letter on (CDRH). 5 import refusals are filed under this FEI, the latest on . FDA’s Import Alert 89-04 Red List names this firm at this address. FDA printed no FEI on the entry. No court action is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
OAI
Inspections on file
2, 2018–2023
1 OAI · 1 VAI · 0 NAI
Inspections with 483 observations shown
2 of 2
Form 483s posted by FDA
None held for this FEI
Import refusals
5 refusals
Import alerts
89-04Red List
Matched on FDA’s firm name and street address

Inspection history

2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Training - Lack of or inadequate procedures

      Procedures for training and identifying training needs have not been adequately established.

    2. QSR

      The quality system record has not been maintained.

  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Production and Process Change Procedures, lack of or Inad.

      Procedures for changes to a process have not been adequately established.

    2. Adverse events--all info not in file

      MDR event files do not contain or reference all adverse event information in the possession of the reporting entity, including documentation of the deliberations and decision making process used to determine if an event was or was not reportable.

Regulations FDA cited here

The 4 sections of 21 CFR in FDA’s published citations for this facility’s 2 inspections, with how many of them cited each.

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Import refusals

FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 3010172486, one row per refused product line.

5 refusals from 2021-03-16 to 2022-02-23 · charged NO 510(K) on 2, DEVICEGMPS on 2 · no FDA sample analysis

FDA firm profile
  • 2022-02-23

    COVID-19 TEST HOME COLLECTION KIT DEVICES83QLW

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    Division of West Coast Imports799-6922980-5/11/1

  • 2021-03-30

    CORONAVIRUS ANTIGEN DETECTION TEST SYSTEM.83QKP

    NO 510(K) 801(a)(3); 502(o) Misbranding

    Division of West Coast ImportsEFJ-0001823-5/21/1

  • 2021-03-30

    CORONAVIRUS ANTIGEN DETECTION TEST SYSTEM.83QKP

    NO 510(K) 801(a)(3); 502(o) Misbranding

    Division of West Coast ImportsEFJ-0001823-5/11/1

  • 2021-03-16

    GLUCOSE OXIDASE, GLUCOSE75CGA

    DEVICEGMPS 501(h), 801(a)(1); ADULTERATION

    Division of Northern Border ImportsSCS-4484749-1/21/1

  • 2021-03-16

    SYSTEM, TEST, BLOOD GLUCOSE, OVER THE COUNTER75NBW

    DEVICEGMPS 501(h), 801(a)(1); ADULTERATION

    Division of Northern Border ImportsSCS-4484749-1/11/1

Import alerts

FDA import alert pages, last synced Nov 23, 2026, 04:17 UTC. On a Red List, FDA can detain the firm's listed products without examining them.

  • 89-04Red ListDetention Without Physical Examination of Devices from Firms that Have not met Device Quality System RequirementsImport alert pageAlert on fda.gov

    Date published 2020-09-21

    FDA’s entry
    Philosys Co. Ltd · 28-5 Gwangwol-Gil, Okgu-Eup, Gunsan, Jeonrabuk KOREA (THE REPUBLIC OF)
    FDA’s inspection record for FEI 3010172486
    Philosys Co. Ltd · 28-5 Gwangwol-Gil, Okgu-Eup, Gunsan, Korea (the Republic of)

    Matched on exact firm name, street address and country. FDA printed no FEI on this entry.

    16 products

    • 75 C - - GAGlucose Oxidase, Glucose
    • 75 C - - GAGlucose Oxidase, Glucose
    • 75 C - - GAGlucose Oxidase, Glucose
    Show 13 more products
    • 75 C - - GAGlucose Oxidase, Glucose
    • 75 C - - GAGlucose Oxidase, Glucose
    • 75 C - - GAGlucose Oxidase, Glucose
    • 75 C - - GAGlucose Oxidase, Glucose
    • 75 C - - GAGlucose Oxidase, Glucose
    • 75 N - - BWSystem, Test, Blood Glucose, Over the Counter
    • 75 N - - BWSystem, Test, Blood Glucose, Over the Counter
    • 75 N - - BWSystem, Test, Blood Glucose, Over the Counter
    • 75 N - - BWSystem, Test, Blood Glucose, Over the Counter
    • 75 N - - BWSystem, Test, Blood Glucose, Over the Counter
    • 75 N - - BWSystem, Test, Blood Glucose, Over the Counter
    • 75 N - - BWSystem, Test, Blood Glucose, Over the Counter
    • 75 N - - BWSystem, Test, Blood Glucose, Over the Counter

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3010172486 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3010172486. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.