21 CFR 820.186: Quality system record
FDA's inspection data lists 53 Form 483 citations of 21 CFR 820.186 at 49 facilities, from September 2009 to September 25, 2025. 2 warning letters state a violation of it. eCFR shows this section removed on February 4, 2026.
Counts: FDA Data Dashboard citations, last synced Nov 23, 2026, 04:25 UTC, and FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Regulation text: eCFR. FDAgent is independent of FDA.
- Form 483 citations
- 53
- At 53 inspections
- Facilities cited
- 49
- Distinct FEIs
- Most recent citation
- Sep 25, 2025
- EuropTec Polymer AG
- Warning letters
- 2
- State a violation of it
The regulation
Removed from the CFR on 2026-02-04. The text shown is the last text in force before then (eCFR version of 2016-12-30).
§ 820.186 Quality system record.
Each manufacturer shall maintain a quality system record (QSR). The QSR shall include, or refer to the location of, procedures and the documentation of activities required by this part that are not specific to a particular type of device(s), including, but not limited to, the records required by § 820.20. Each manufacturer shall ensure that the QSR is prepared and approved in accordance with § 820.40.
21 CFR Part 820: Quality Management System Regulation, Subpart MRead on eCFR
Citations by year
Form 483 citations of 21 CFR 820.186 by the year the inspection ended. FDA's citation data starts with inspections ending in October 2008, and the current year runs to the latest sync, so both are partial years (lighter bars).
Peak: 10 in 2015
| Year | Form 483 citations |
|---|---|
| 2009 | 1 |
| 2010 | 5 |
| 2011 | 9 |
| 2012 | 5 |
| 2013 | 3 |
| 2014 | 5 |
| 2015 | 10 |
| 2016 | 3 |
| 2017 | 3 |
| 2018 | 1 |
| 2019 | 1 |
| 2021 | 2 |
| 2023 | 3 |
| 2025 | 2 |
How FDA words this citation
FDA's short descriptions and citation text on these 53 citations, most used first. These are FDA's standard summaries in its citation data, not the text of any one Form 483.
Short descriptions
- QSR53 citations
Citation text
- “The quality system record has not been adequately maintained.”25 citations
- “The quality system record has not been maintained.”21 citations
- “The quality system record does not include or refer to the location of quality system procedures and documentation of quality system activities.”2 citations
- “The quality system record does not include or refer to the location of and locations of all quality system procedures and documentation of quality system activities.”1 citations
Show 2 more
- “The quality system record has not been adequately maintained Electronic records are used, but they do not meet systems validation, audit trail, and signature manifestation requirements to ensure that they are trustworthy, reliable and generally equivalent to paper records.”1 citations
- “The quality system record has not been maintained Electronic records are used, but they do not meet retention, audit trail, employee accountability/responsibility policy, systems documentation control, and open systems control requirements to ensure that they are trustworthy, reliable and generally equivalent to paper records.”1 citations
Facilities cited most recently
The latest inspections where FDA cited 21 CFR 820.186, by inspection end date, with FDA's short description. Each links to the facility's FDA record on FDAgent.
Show 4 more
Warning letters stating a violation of 21 CFR 820.186
2 warning letters FDA posted state a violation of this section in a numbered violation sentence. The most recent are below, each linking to FDAgent's page for the letter.
“Failure to adequately maintain quality system records, as required by 21 CFR § 820.186.”
- W.H.P.M, Inc.Issued 2022-06-30
“Failure to maintain a quality system record (QSR) which includes, or refers to the location of, procedures and the documentation of activities required by this part that are not specific to a particular type of device(s), including, but not limited to, the records required by 21 CFR 820.20, as required by 21 CFR 820.186.”
- Tianjin Bolang Science-Technology Development Co., Ltd.Issued 2021-06-17
Often cited at the same inspection
Of the 53 inspections where FDA cited 21 CFR 820.186, the share where it also cited each of these sections.
- 21 CFR 820.100Corrective and preventive action30 (57%)
- 21 CFR 820.30Design controls30 (57%)
- 21 CFR 820.198Complaint files28 (53%)
- 21 CFR 820.22Quality audit22 (42%)
- 21 CFR 820.50Purchasing controls22 (42%)
- 21 CFR 820.20Management responsibility20 (38%)
Ask FDAgent about 21 CFR 820.186
FDAgent reads FDA's Form 483 citations, observations and warning letters and answers in plain English, with links to the records.
Related
About this data
- Citation counts come from FDA's Data Dashboard, which lists the regulation behind each Form 483 observation for inspections since October 2008.
- Observation excerpts come from Form 483 PDFs in FDA's electronic reading room, tied to this section only by exact wording as described above.
- A warning letter counts for this section when one of its numbered violation sentences cites 21 CFR 820.186; FDAgent quotes that sentence exactly.
- FDAgent rebuilds this page every week from FDA's public data and is independent of FDA.