A.M. Surgical, Inc.
402 Amherst St Ste 300, Nashua, New Hampshire, United States
FEI 3001617730
FDA has 7 inspections on file for A.M. Surgical, Inc. in Nashua, New Hampshire (FEI 3001617730), from 2008 to 2023. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published one citation from this inspection, under 21 CFR Part 820, the device quality system rules. It is listed below.
No warning letter, court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 7, 2008–2023
- 0 OAI · 5 VAI · 2 NAI
- Inspections with 483 observations shown
- 5 of 7
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
7 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
DMR - not or inadequately maintained
A device master record has not been adequately maintained.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Statistical techniques - Lack of or inadequate procedures
Procedures for identifying valid statistical techniques required for establishing, controlling, and verifying the acceptability of process capability and product characteristics have not been established.
Supplier oversight
The type and extent of control to be exercised over the product and contractors was not clearly defined.
Evaluation of suppliers, contractors, etc., requirements
Requirements that must be met by suppliers and contractors have not been established.
Design output - review and approval
Design output was not reviewed and approved before release.
Show 1 more citation
QSR
The quality system record has not been adequately maintained.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Remain legible
Labels were not printed so as to remain legible during processing, storage, handling, distribution, and/or use.
Complaints
Complaint files are not adequately maintained.
Documentation
Process validation results have not been adequately documented.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Production processes
Production processes were not to ensure that a device conforms to its specifications.
Quality Audit/Reaudit - conducted
Quality audits have not been performed.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Device identification, control numbers
The device history record does not include certain device identifications and control numbers used.
Regulations FDA cited here
The 11 sections of 21 CFR in FDA’s published citations for 5 of this facility’s 7 inspections, with how many of them cited each.
- 21 CFR 820.22Quality audit1 of 5
- 21 CFR 820.30Design controls1 of 5
- 21 CFR 820.50Purchasing controls1 of 5
- 21 CFR 820.70Production and process controls1 of 5
- 21 CFR 820.75Process validation1 of 5
- 21 CFR 820.120Device labeling1 of 5
- 21 CFR 820.181Device master record1 of 5
- 21 CFR 820.184Device history record1 of 5
Show all 11
- 21 CFR 820.186Quality system record1 of 5
- 21 CFR 820.198Complaint files1 of 5
- 21 CFR 820.250Statistical techniques1 of 5
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3001617730 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3001617730. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.