Laser Dental Innovations
1219 Quail Creek Cir, San Jose, California, United States
FEI 3003610527
FDA has 2 inspections on file for Laser Dental Innovations in San Jose, California (FEI 3003610527), from 2017 to 2019. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 9 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.
The most recent OAI inspection ended on Dec 14, 2017. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
Five regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 2, 2017–2019
- 1 OAI · 1 VAI · 0 NAI
- Inspections with 483 observations shown
- 2 of 2
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices9 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Quality audits - Lack of or inadequate procedures
Procedures for quality audits have not been adequately established.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Procedures not adequately established or maintained
Document control procedures have not been adequately established and maintained.
QSR
The quality system record has not been adequately maintained.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Show 4 more citations
Equipment suitability & capability
Certain inspection, measuring and test equipment is not suitable for its intended purposes and capable of producing valid results.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been adequately established.
Design Validation - Risk analysis not performed/inadequate
Risk analysis is inadequate and is incomplete.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices12 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Procedures not adequately established or maintained
Document control procedures have not been adequately established.
Lack of or inadequate procedures - Acceptance activities
Procedures for acceptance activities have not been adequately established.
Design history file
The design history file does not demonstrate that the design was developed following the approved design plan.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Quality Audits - defined intervals
Quality audits were not performed at defined intervals and at sufficient frequency to determine whether the quality system activities and results comply with quality system procedures.
Show 7 more citations
DHR - not or inadequately maintained
A device history record has not been maintained.
Production and Process Change Procedures, lack of or Inad.
Procedures for changes to a specification, method, process and procedure have not been established.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely calibrated, inspected, checked and maintained have not been established.
Sampling methods - Lack of or inadequate procedures
Procedures to ensure sampling methods are adequate for their intended use have not been adequately established.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Regulations FDA cited here
The 14 sections of 21 CFR in FDA’s published citations for this facility’s 2 inspections, with how many of them cited each.
- 21 CFR 820.22Quality audit2 of 2
- 21 CFR 820.30Design controls2 of 2
- 21 CFR 820.40[Reserved]2 of 2
- 21 CFR 820.50Purchasing controls2 of 2
- 21 CFR 820.72Inspection, measuring, and test equipment2 of 2
- 21 CFR 820.100Corrective and preventive action2 of 2
- 21 CFR 820.198Complaint files2 of 2
- 21 CFR 820.25Personnel1 of 2
Show all 14
- 21 CFR 820.70Production and process controls1 of 2
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 2
- 21 CFR 820.90Nonconforming product1 of 2
- 21 CFR 820.184Device history record1 of 2
- 21 CFR 820.186Quality system record1 of 2
- 21 CFR 820.250Statistical techniques1 of 2
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3003610527 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3003610527. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.