Maquet Cardiovascular, LLC
45 Barbour Pond Dr, Wayne, New Jersey, United States
FEI 2242352
FDA has 8 inspections on file for Maquet Cardiovascular, LLC in Wayne, New Jersey (FEI 2242352), from 2009 to 2024. The latest ended on and was classified OAI, official action indicated.
OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.
FDA published 11 citations from this inspection: 8 under 21 CFR Part 820, the device quality system rules, and 3 under 21 CFR Part 803, the device adverse event reporting rules. They are listed below. One FOIA request has asked FDA for this inspection's 483.
FDA issued a warning letter on (CDRH). 1 import refusal is filed under this FEI, on . No court action is filed under this FEI.
19 regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- OAI
- Inspections on file
- 8, 2009–2024
- 6 OAI · 2 VAI · 0 NAI
- Inspections with 483 observations shown
- 8 of 8
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 1 request, latest Oct 11, 2024
- Warning letters
- · CDRH ↗
- Import refusals
- 1 refusal
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
8 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices11 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Report of Death or Serious Injury
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.
Submission of Supplement to Individual Report Within One Month
An individual supplemental report was not submitted to FDA within one month following receipt of information that was not provided when the initial report was submitted.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Design Validation - Risk analysis not performed/inadequate
Risk analysis is inadequate.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Show 6 more citations
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.
Lack of or inadequate receiving acceptance procedures
Procedures for acceptance of incoming product have not been adequately established.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Design validation- Lack of or inadequate procedures
Procedures for design validation have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Design validation - user needs and intended uses
Design validation did not ensure the device conforms to defined user needs and intended uses.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices10 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Statistical techniques - Lack of or inadequate procedures
Procedures for identifying valid statistical techniques required for establishing, controlling, and verifying the acceptability of process capability and product characteristics have not been established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Training records
Personnel training is not documented.
Design Validation - Risk analysis not performed/inadequate
Risk analysis is inadequate.
Show 5 more citations
Submission of Supplement to Individual Report Within One Month
An individual supplemental report was not submitted to FDA within one month following receipt of information that was not provided when the initial report was submitted.
Design validation - user needs and intended uses
Design validation did not ensure the device conforms to defined user needs and intended uses.
Report of Death or Serious Injury
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.
Type of Reportable Event
An individual medical device manufacturer report submitted per Form FDA 3500A format did not indicate in Block H the type of reportable event.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices16 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Environmental control Lack of or inadequate procedures
Procedures to control environmental conditions have not been established.
Personnel requirements, Lack of or inadequate requirements
Requirements have not been adequately established to address personnel cleanliness and clothing.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Statistical techniques - Lack of or inadequate procedures
Procedures for identifying valid statistical techniques required for establishing, controlling, and verifying the acceptability of process capability and product characteristics have not been adequately established.
Contamination control, Lack of or inadequate procedures
Procedures to prevent contamination of equipment or product by substances that may have an adverse effect on product quality have not been established.
Show 11 more citations
Review and evaluation for investigation
Not all complaints have been reviewed and evaluated to determine whether an investigation is necessary.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been adequately established.
Design validation- Lack of or inadequate procedures
Procedures for design validation have not been established.
Submission of Supplement to Individual Report Within One Month
A supplemental report was not submitted to FDA within one month following receipt of information that was not provided when the initial report was submitted.
Procedures not adequately established or maintained
Document control procedures have not been adequately established.
Acceptable supplier records, inadequate records
Records of acceptable suppliers and contractors have not been adequately established.
Report of Death or Serious Injury
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices17 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design Validation - Risk analysis not performed/inadequate
Risk analysis is inadequate.
Nonconforming product control
Products that do not conform to specifications are not adequately controlled.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been established.
Documentation - review in response to changes or deviations
There is no documentation of the revalidation of a process performed in response to changes or process deviations.
Evaluation of suppliers, contractors, etc., requirements
Requirements that must be met by suppliers have not been adequately established.
Show 12 more citations
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.
Report of Death or Serious Injury
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.
Management review - Lack of or inadequate procedures
Procedures for management review have not been adequately established.
Lack/Inad procedure-Monitoring/Control of Validated Proces
Procedures for monitoring and control of process parameters for a validated process have not been established.
Records of MDR Investigation
Investigation records of MDR reportable complaints do not include required information.
Report of risk to health
A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been established.
Design review - Lack of or inadequate procedures
Procedures for design review have not been established.
Design transfer - Lack of or inadequate procedures
Procedures for design transfer have not been established.
Submission of Supplement to Individual Report Within One Month
A supplemental report was not submitted to FDA within one month following receipt of information that was not provided when the initial report was submitted.
Document review, approval by designated individual
Documents were not reviewed and not approved by designated individual(s) prior to issuance .
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices14 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been established.
Lack of System for Determining MDR Events
The written MDR procedure does not include an internal system which provides for a standardized review process/procedure for determining when an event meets the criteria for reporting.
Design Validation - Risk analysis not performed/inadequate
Risk analysis is inadequate.
Personnel
Personnel do not have the necessary education, training, and experience to perform their jobs.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.
Show 9 more citations
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Evaluation, timeliness, identification
Complaints representing events that are MDR reportable were not promptly reviewed, evaluated, and investigated by a designated individual.
Availability
Required records were not made readily available for review and copying by the FDA.
Report of Death or Serious Injury
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.
Process changes - review, evaluation and revalidation
A validated process was not revalidated when changes or process deviations occurred.
Lack/Inad procedure-Monitoring/Control of Validated Proces
Procedures for monitoring and control of process parameters for a validated process have not been established.
Management review - defined interval, sufficient frequency
Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system.
Design validation acceptance criteria
Acceptance criteria were not established prior to the performance of validation activities.
Procedures for product review,disposition lack of/inadequate
Procedures that define the responsibility for review and the authority for the disposition of nonconforming product have not been established.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices13 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been established.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.
Rationale documented for no investigation
Records for complaints where no investigation was made do not include required information.
Personnel
Personnel do not have the necessary background and experience to perform their jobs.
Process changes - review, evaluation and revalidation
A validated process was not reviewed and evaluated and revalidated when changes or process deviations occurred.
Show 8 more citations
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.
Design validation - user needs and intended uses
Design validation did not ensure the device conforms to defined user needs and intended uses.
Design validation acceptance criteria
Acceptance criteria were not established prior to the performance of validation activities.
Evaluation, timeliness, identification
Complaints representing events that are MDR reportable were not promptly reviewed, evaluated, and investigated by a designated individual.
Documentation of validated process performance
There is inadequate documentation of the date performed and the individual performing the process for a validated process.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been established.
Report of Death or Serious Injury
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Time to report - 10 days
A report of the required information regarding device correction and removal actions was not sent to FDA within 10 days of initiating the correction or removal.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated and approved according to established procedures.
Contamination control, Lack of or inadequate procedures
Procedures have not been defined to prevent contamination of equipment or product by certain substances.
Design validation - user needs and intended uses
Design validation did not ensure that devices conform to defined user/patient needs and intended uses.
Environmental control Lack of or inadequate procedures
Appropriate procedures have not been defined for controlling environmental conditions.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Approval, inadequate purchasing data
Purchasing information that clearly includes/describes the specified requirements for purchased or otherwise received products and services was not complete.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
One request to FDA under the Freedom of Information Act named this site, on Oct 11, 2024. It asked for a Form 483.
Inspection ending Jul 11, 2024OAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Oct 11, 2024FDA-FOIA-2024-9038Form 483Closed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 18 sections of 21 CFR in FDA’s published citations for this facility’s 8 inspections, with how many of them cited each.
- 21 CFR 820.30Design controls7 of 8
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?6 of 8
- 21 CFR 820.100Corrective and preventive action6 of 8
- 21 CFR 820.198Complaint files6 of 8
- 21 CFR 820.75Process validation5 of 8
- 21 CFR 803.56If I am a manufacturer, in what circumstances must I submit a supplemental or followup report and what are the requirements for such reports?4 of 8
- 21 CFR 820.25Personnel4 of 8
- 21 CFR 820.50Purchasing controls4 of 8
Show all 18
- 21 CFR 820.90Nonconforming product3 of 8
- 21 CFR 806.10Reports of corrections and removals2 of 8
- 21 CFR 820.20Management responsibility2 of 8
- 21 CFR 820.40[Reserved]2 of 8
- 21 CFR 820.70Production and process controls2 of 8
- 21 CFR 820.250Statistical techniques2 of 8
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 8
- 21 CFR 803.52If I am a manufacturer, what information must I submit in my individual adverse event reports?1 of 8
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 8
- 21 CFR 820.180General requirements1 of 8
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Import refusals
FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 2242352, one row per refused product line.
1 refusal on 2024-02-16 · charged FRNMFGREG on 1 · no FDA sample analysis
FDA firm profile- 2024-02-16
DEVICE, ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES79GEI
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
Division of Southeast ImportsAEK-4699714-2/11/1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 2242352 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 2242352. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.