Welch Allyn, Inc.
4341 State Street Rd, Skaneateles Falls, New York, United States
FEI 1316463
FDA has 7 inspections on file for Welch Allyn, Inc. in Skaneateles Falls, New York (FEI 1316463), from 2012 to 2026. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published one citation from this inspection, under 21 CFR Part 820, the device quality system rules. It is listed below. One of the two FOIA requests for this site has asked FDA for this inspection's 483.
12 import refusals are filed under this FEI, the latest on . No warning letter or court action is filed under this FEI.
Three regulations were cited at more than one inspection: 21 CFR 820.100(a), 21 CFR 820.90(a) and 21 CFR 820.198(a).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 7, 2012–2026
- 0 OAI · 5 VAI · 2 NAI
- Inspections with 483 observations shown
- 5 of 7
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 2 requests, latest Mar 18, 2026
- Warning letters
- None on file
- Import refusals
- 12 refusals
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
7 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Process control procedures, Lack of or inadequate procedures
Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.
Documentation of software validation
Software validation activities and results for computers or automated data processing systems used as part of production have not been adequately documented.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Evaluation of suppliers, contractors, etc., requirements
Requirements that must be met by suppliers have not been adequately established.
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been adequately established.
Evaluation, timeliness, identification
Complaints representing events that are MDR reportable were not promptly reviewed, evaluated, and investigated by a designated individual.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Nonconforming product control
Products that do not conform to specifications are not adequately controlled.
Show 2 more citations
Records of complaint investigation
Records of complaint investigations do not include required information.
Complaints
Complaint files are not adequately maintained.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Two requests to FDA under the Freedom of Information Act named this site, most recently on Mar 18, 2026. Each one asked for a Form 483.
Inspection ending Jan 9, 2026VAI
1 request. In the inspection history above. FDA released records in answer to a request for its Form 483 on Aug 27, 2026. FDAgent holds no copy that FDA posted.
- Mar 18, 2026FDA-FOIA-2026-2720Form 483· 1 of 5 sitesRecords released
Inspection ending Aug 17, 2023VAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Oct 19, 2023FDA-FOIA-2023-9272Form 483Closed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 6 sections of 21 CFR in FDA’s published citations for 5 of this facility’s 7 inspections, with how many of them cited each.
- 21 CFR 820.100Corrective and preventive action3 of 5
- 21 CFR 820.90Nonconforming product2 of 5
- 21 CFR 820.198Complaint files2 of 5
- 21 CFR 820.50Purchasing controls1 of 5
- 21 CFR 820.70Production and process controls1 of 5
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 5
Import refusals
FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 1316463, one row per refused product line.
12 refusals from 2021-09-28 to 2025-07-23 · charged FRNMFGREG on 10, NOT LISTED on 8 · no FDA sample analysis
FDA firm profile- 2025-07-23
TESTER, AUDITORY IMPEDANCE77ETY
NOT LISTED 502(o), 801(a)(3); MISBRANDING
Division of West Coast Imports799-6790215-7/11/1
- 2025-06-04
OTOSCOPE77ERA
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
REGISTERED 502(o), 801(a)(3); MISBRANDING
Division of Southwest Imports8RG-2574190-7/11/4
- 2025-06-04
RETINOSCOPE, BATTERY-POWERED86HKM
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
REGISTERED 502(o), 801(a)(3); MISBRANDING
Division of Southwest Imports8RG-2574190-7/11/3
- 2025-06-04
OTOSCOPE77ERA
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
REGISTERED 502(o), 801(a)(3); MISBRANDING
Division of Southwest Imports8RG-2574190-7/11/2
- 2025-06-04
OPHTHALMOSCOPE, AC-POWERED86HLI
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
REGISTERED 502(o), 801(a)(3); MISBRANDING
Division of Southwest Imports8RG-2574190-7/11/1
Show 7 more
- 2025-02-05
ELECTROCARDIOGRAPH74DPS
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
Division of West Coast Imports799-5190411-0/11/1
- 2024-09-19
TESTER, AUDITORY IMPEDANCE77ETY
NOT LISTED 502(o), 801(a)(3); MISBRANDING
NO 510(K) 801(a)(3); 502(o) Misbranding
Division of West Coast Imports799-3236988-8/11/1
- 2024-05-22
RETINOSCOPE, BATTERY-POWERED86HKM
NOT LISTED 502(o), 801(a)(3); MISBRANDING
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
Division of Southeast Imports799-0440948-0/11/1
- 2023-12-13
MONITOR, PHYSIOLOGICAL, PATIENT (WITHOUT ARRHYTHMIA DETECTION OR ALARMS)74MWI
NOT LISTED 502(o), 801(a)(3); MISBRANDING
Division of West Coast Imports799-8018577-1/11/1
- 2023-08-31
ELECTROCARDIOGRAPH74DPS
NOT LISTED 502(o), 801(a)(3); MISBRANDING
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
Division of Northeast ImportsBUU-1575770-7/11/1
- 2021-12-06
ELECTROCARDIOGRAPH74DPS
NOT LISTED 502(o), 801(a)(3); MISBRANDING
NOT LISTED 502(o), 801(a)(3); MISBRANDING
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
Division of Southeast ImportsAEK-0039297-5/11/1
- 2021-09-28
OPHTHALMOSCOPE, BATTERY-POWERED86HLJ
NOT LISTED 502(o), 801(a)(3); MISBRANDING
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
Division of Southeast Imports799-4718906-3/11/1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1316463 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1316463. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.