Meridian Medical Technologies, LLC a Subsidiary of Kindeva Drug Delivery, L.P.
1945 Craig Rd, Saint Louis, Missouri, United States
FEI 1950222DUNS 089384390
Company: Meridian Medical Technologies, LLC · Linked by exact D-U-N-S match (049504624) in FDA registration data
FDA has 19 inspections on file for Meridian Medical Technologies, LLC a Subsidiary of Kindeva Drug Delivery, L.P. in Saint Louis, Missouri (FEI 1950222), from 2009 to 2026. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 9 citations from this inspection. They are listed below. One of the five FOIA requests for this site has asked FDA for this inspection's 483.
The most recent OAI inspection ended on Mar 24, 2017. FDA issued a warning letter on (CDER). No court action or import refusal is filed under this FEI.
Nine regulations were cited at more than one inspection; they are listed below.
As of Nov 21, 2026, the site is registered with FDA as a drug establishment for analysis, through Dec 31, 2027. Its registrant, Meridian Medical Technologies, LLC (D-U-N-S 049504624), has registered 1 other site.
Sources:FDA firm profile ↗FDA drug establishment registrations ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 19, 2009–2026
- 7 OAI · 9 VAI · 3 NAI
- Inspections with 483 observations shown
- 8 of 19
- 3 posted 483s not matched to an inspection
- Requested under FOIA
- 5 requests, latest Aug 12, 2026
- Warning letters
- · CDER ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
- Drug establishment registration
- Through
Inspection history
19 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC · FDA reading room, last synced Nov 23, 2026, 04:33 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices9 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Written complaint record to be maintained at facility
A written record of each complaint is not maintained in a file designated for drug product complaints at the facility where the drug product was manufactured, processed or packed.
Scientifically sound laboratory controls
Laboratory controls do not include the establishment of scientifically sound and appropriate specifications and test procedures designed to assure that in-process materials and drug products conform to appropriate standards of identity, strength, quality and purity.
Calibration/Inspection/Checking not done
Routine checking of automatic equipment is not performed according to a written program designed to assure proper performance.
Investigations of discrepancies, failures
There is a failure to thoroughly review any unexplained discrepancy and the failure of a batch or any of its components to meet any of its specifications whether or not the batch has been already distributed.
Control procedures to monitor and validate performance
Control procedures are not established which monitor the output and validate the performance of those manufacturing processes that may be responsible for causing variability in the characteristics of in-process material and the drug product.
Show 4 more citations
Procedures for sterile drug products
Procedures designed to prevent microbiological contamination of drug products purporting to be sterile are not written and followed.
Contamination, chemical or physical change, deterioration
An NDA-Field Alert Report was not submitted within three working days of receipt of information concerning bacteriological contamination and significant chemical, physical, or other change or deterioration in a distributed drug product.
- ISO 13485:2016 Clause 7.1
Required elements not determined in planning product realization
In product realization planning your firm did not adequately determine, as appropriate, quality objectives and requirements for the product and required activities specific to the product together with criteria for product acceptance.
- ISO 13485:2016 Clause 8.2.1
Information gathered in feedback process not used as input into risk management
Information gathered in the feedback process does not adequately serve as potential input into risk management for monitoring and maintaining product requirements as well as the product realization or improvement processes.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices9 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Procedures for sterile drug products
Procedures designed to prevent microbiological contamination of drug products purporting to be sterile are not written and followed.
Design validation - Risk analysis
Results of the design risk analysis were not adequately documented.
Design output - Lack of or inadequate procedures
Procedures for design output have not been adequately established.
Acceptance criteria for sampling & testing
Acceptance criteria for the sampling and testing conducted by the quality control unit is not adequate to assure that batches of drug products meet each appropriate specification and appropriate statistical quality control criteria as a condition for their approval and release.
Investigations of discrepancies, failures
There is a failure to thoroughly review any unexplained discrepancy and the failure of a batch or any of its components to meet any of its specifications whether or not the batch has been already distributed.
Show 4 more citations
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Scientifically sound laboratory controls
Laboratory controls do not include the establishment of scientifically sound and appropriate test procedures designed to assure that in-process materials and drug products conform to appropriate standards of identity, strength, quality and purity.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Control procedures to monitor and validate performance
Control procedures are not established which monitor the output and validate the performance of those manufacturing processes that may be responsible for causing variability in the characteristics of in-process material and the drug product.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices11 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate receiving acceptance procedures
Procedures for acceptance of incoming product have not been adequately established.
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been adequately established.
Design verification - output does not meet input requirement
Design verification does not confirm that design output meets design input requirements.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been established.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Show 6 more citations
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Design validation- Lack of or inadequate procedures
Procedures for design validation have not been adequately established.
Valid stability test methods
The written stability program for drug products does not include meaningful test methods.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.
Design transfer - Lack of or inadequate procedures
Procedures for design transfer have not been adequately established.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
- Not classified yetFDA posted this Form 483; FDAgent has not matched it to a Data Dashboard inspectionNo inspection matchedForm 483 pagePDF
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices14 citations published
Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent’s.
- Observation 1
Failure to thoroughly investigate discrepancies and batch failures
There is a failure to thoroughly review any unexplained discrepancy and the failure of a batch or any of its components to meet any of its specifications whether or not the batch has been already distributed. Specifically, A. Your investigation of two complaints of 'Failure to Activate' for the EpiPen 0.3 mg Auto-injector finished product lot number 5FA665 did not extend to all potentially impacted lots. Your firm confirmed these complaints through investigations conducted under QAR Detail #PR ID: 22268 opened 5/9/16 and QAR Detail #PR ID: 28587 opened 2/23/17. These investigations identified manufacturing defects that resulted in distorted (b) (4) in the Power Pak (b) (4) of the Power Pak assembly. This assembly is used in both EpiPen and EpiPen Jr. Auto-injectors. Power Pak assembly lots (b) (4) (b) (4) were identified as potentially affected lots that were included in finished product lots. Your firm inspected all Power Pak (b) (4) for defects that were part of Power Pak assembly lots (b) (4) Your investigation implicated (b) (4) additional Power Pak assembly lots that were not utilized in finished product, (b) (4) You inspected lot (b) (4) and found up to 11 instances of flash near the (b) (4) and rejected this lot. Lot (b) (4) was also rejected but you did not evaluate or inspect the lot to determine if additional (b) (4) manufacturing defects were present. Your firm did not implement an inspection of complaint devices alleging similar failure modes for "failure to activate" / "difficult to activate" to determine if these devices had a distorted (b) (4) From your investigation, (b) (4) finished product lots remain on the market that contain potentially defective Power Pak (b) (4). B. Your firm did not perform '(b) (4) testing" on the six (6) implicated lots identified in QAR PR 15815 prior to release of the lots. This QAR was opened when EpiPen Jr. lot 5GR446 failed functionality testing for low delivered volume due to (b) (4) defect. Your firm listed the most probable root cause of the (b) (4) defect as the presence of a non-homogeneous area in material. Your investigation for the other finished product using these (b) (4) was limited to batch record and testing record review. No additional physical tests were performed. C. Your investigation PR 24844 did not fully address the issue of the large number of low fills for atropine in Lot 6M1454 of Atropine and Pralidoxime Chloride Injection (ATNAA). For this lot, your firm identified (b) (4) units with low fill in the front chamber (Atropine). The (b) (4) other batches inspected during the (b) (4) had a low fill range of (b) (4) units per batch. No new CAPAs were identified for the Lot 6M1454 investigation and your investigation concluded the filler was operating within its capabilities.
- Observation 2
Lack of control procedures for manufacturing process variability
Control procedures are not established which monitor the output and validate the performance of those manufacturing processes that may be responsible for causing variability in the characteristics of in-process material and the drug product. Specifically, A. Your procedures and practices allow your employees to repeatedly inspect lots of ATNAA to cull out defective units. These repeated 100% visual inspections however, do not remove all of the defective units. 1. Based upon review of production data, (b) (4)% of the batches from the ATNAA / DuoDote process ((b) (4) out of (b) (4) batches inspected between 1 JAN 2015 and 21 FEB 2017) are not complying with the specified criteria for percent defect for critical, major, or minor defects during the 100% manual inspection. Your firm performed an additional 100% inspection for (b) (4) batches out of (b) (4), approximately (b) (4)% of the time. Additionally, there has been at least one batch, Lot 6M1133, which was subsequently rejected because it did not meet the criteria for a (b) (4); after the performance of a (b) (4) 100% inspection. 2. ANTAA Lot 6M1454, exceeded the alert limits for critical and minor defects during the initial 100% visual inspection. It passed the initial Acceptable Quality Limit (Visual (b) (4)) QA performs on (b) (4) of the "acceptable" units. Since this lot exceeded your defect alert limits, your firm performed a (b) (4) on (b) (4) of the "acceptable" units. During the (b) (4) the QA Inspector found one (1) unit with a low fill in the front chamber which is a critical defect. Therefore, your firm performed another 100% visual inspection and found 20 units with critical defects, 30 with major defects, and 907 with minor defects. (b) (4) was performed on the "acceptable" units and three minor defects were found. This batch was released on 19 JAN 2017 by your Quality Assurance Unit. B. Your manual visual inspection programs are deficient since: 1. There is no evaluation of the efficiency / effectiveness of an individual performing repetitive inspection over time for your basic units. 2. The (b) (4) certification process for detection of defects in your basic units is not currently performed at the end of the shift in order to evaluate the impact of fatigue, if any, upon defect detection capability. 3. The process for incoming (b) (4) introduced into EpiPen manufacturing does not specify the maximum amount of time (b) (4) that an inspector is allowed to perform visual inspection. Current established practice does not limit inspection time to prevent errors caused by fatigue and loss of concentration. The operators do not have required break times and are not qualified for this inspection before being assigned and performing the (b) (4) evaluation task.
Show observations 3 to 14
- Observation 3
Inadequate written complaint handling procedures
Procedures describing the handling of all written and oral complaints regarding a drug product are not established, written and followed. Specifically, A. Your firm has not documented rationale why (b) (4) complaints of a similar nature on the same lot are necessary to identify a trend per your procedure Product Complaint Handling SOP-QLC-QLE-00702. B. Complaint classifications, listed in GPB-QS1073 Prioritization of Pfizer Product Quality Complaints, are assigned prioritization classifications that are not commensurate with the associated clinical risk (e.g. no dose delivered) based on the alleged complaint. Three complaint prioritization classifications (Expedite, High, Normal) are assigned to complaint classifications. These prioritization classifications dictate the speed and thoroughness of the investigations conducted by your firm. There are at least ten product complaint classifications that do not adequately reflect the associated risk including, but not limited to, 'Spontaneous Activation' classified as Normal, 'Container Broken/Cracked/Leaking Prior to Use' classified as Normal, and '(b) (4) (b) (4) classified as Normal. C. No safety or risk assessment has been completed to establish the acceptability of the mean complaint rates received by Pfizer Corporate, which is utilized by your firm to establish the upper and lower confidence limits of complaint.
- Observation 4
Inadequate statistical techniques for process capability
Procedures for identifying valid statistical techniques required for establishing, controlling, and verifying the acceptability of process capability and product characteristics have not been adequately established. Specifically, insufficient statistical assessments were in place for establishing, controlling, and verifying process capability and product characteristics. A. Your firm provided no rationale for the acceptability of your sampling plan at lot release based on the risk associated with releasing defective product. The AQL (Acceptable Quality Limits) sampling plan and associated (b) (4) is not commensurate with the product risk based on intended use and design inputs of the product. B. Your firm does not currently distinguish the failure modes of rejected components/units that are collected in reject bins on the EpiPen manufacturing assembly line. In addition, your firm does not currently track or trend the rejects from the EpiPen assembly manufacturing line and there are no action limits associated with the number or type of rejects. C. Your firm failed to routinely evaluate ongoing state of control of the equipment and process. For example, 1. The capability and suitability of equipment to produce conforming units was not assessed. Specifically, process capability evaluations did not reflect inherent capability of the process equipment to produce conforming units. Instead, you only calculated process capability after a large number of defects were inspected out of the batch. 2. Your firm failed to adequately analyze data to determine process points in which excessive variation occurred and defective units were produced. 3. Your firm lacked sufficient ongoing trending of numerous in-process attributes that are critical to the quality of the finished product. For example, there was insufficient ongoing trending to periodically assess (e.g. process trending reports, Annual Review) process performance to promptly detect atypical variability that may affect product quality.
- Observation 5
Failure to annually review quality data for trends
Records are not maintained so that data therein can be reviewed at least annually to evaluate the quality standards of each drug product to determine the need for changes in specifications or manufacturing or control procedures. Specifically, your analysis of the quality data used to identify existing and recurring quality problems does not include an analysis of complaints based on finished product lots, manufacturing dates or component lots to identify existing or recurring quality problems. Your analysis is presented and discussed (b) (4) during management review meetings, is currently conducted by totaling complaints by (b) (4)
- Observation 6
Inadequate design input procedures for reliability
Procedures for design input have not been adequately established. Specifically, A. The design inputs/requirements of the EpiPen, EpiPen Jr., and (b) (4) products are not appropriate for the intended use of the products in regards to the reliability of the delivery system. The reliability inputs/requirements of the device according to your PRD/TRD 16-001 Rev 1 do not include a system level reliability design input/requirement. 1. Your inputs/requirements list an AQL (b) (4) for critical tests, which equates to establishing a certain confidence of no more than (b) (4) failures per (b) (4) products. (b) (4) failures of delivered dose, or other critical tests, per (b) (4) EpiPen, EpiPen Jr., or (b) (4) products does not give assurance of the level of reliability commensurate with the risk of failure of a critical test as defined by your firm (e.g. delivered dose, (b) (4), etc.). 2. Your pharmaceutical / drug delivery performance inputs/requirements list the inputs/requirements for certain essential performance inputs/requirements being specified to an AQL of (b) (4), which establishes a certain confidence of no more than (b) (4) failures of an essential performance input/requirement, such as (b) (4), per (b) (4) products. 3. No rationale is provided for the acceptability of these design inputs/requirements based on the emergency use, life-saving intended use of the product. B. The design inputs/requirements of the EpiPen, EpiPen Jr., and (b) (4) are conflicting. In PRD/TRD 16-001 Rev 1 the design inputs/requirements related to the 'Pharmaceutical / drug delivery performance' of the EpiPen, EpiPen Jr., and (b) (4) have specifications that are outside of the principal design attributes listed. As an example, the principal design attribute for the EpiPen for dose volume is (b) (4) mL. However, under input/requirement the delivered dose volume for the EpiPen is described as ((b) (4) - (b) (4))mL @ AQL ((b) (4)), Not less than ((b) (4)) or greater than ((b) (4)) mL at AQL ((b) (4)). C. The design inputs/requirements of the EpiPen, EpiPen Jr., and (b) (4) do not directly trace to a risk assessment of the product. It is unclear if the specifications listed under the 'Must' column of the document titled PRD/TRD 16-001 Rev 1 are appropriate for the intended use of the device and the user/patient needs, especially considering that many of the specifications are split into different levels of acceptability based on the confidence of a sampling plan. D. Design inputs/requirement specifications of the EpiPen, EpiPen Jr., and (b) (4) utilize AQLs to describe the necessary confidence level associated with each range of a specification; however, AQLs do not consider the needs of the user/patient. As an example, in PRD/TRD 16-001 Rev 1 the (b) (4) input/requirement specification is described with two different AQLs for two ranges of values associated with (b) (4). E. Your firm has not established design inputs/requirements regarding the ability of the device to successfully inject through clothing as specified by the intended use of the product as defined by your firm's Injection Sites and Site Condition of the PRD/TRD 16-001 Rev 1 document.
- Observation 7
Inadequate design output procedures
Procedures for design output have not been adequately established. Specifically, A. Your firm has not established design outputs that allow an adequate evaluation of conformance to design input requirements. The design outputs as defined by your firm in the Design History File document titled DHF 17-001 Rev 1.0 include batch record tests, controlled drawings, QC test reports, and SOPs that do not establish acceptance criteria for any system level reliability requirements after (b) (4) given the intended use of the product. B. The design outputs acceptance criteria do not reflect the risk associated with the failure of the specifications. The design output documents for the design input requirements allow for failures of the design attributes. As an example, it is acceptable for failures of 'major' functional requirements to occur such as delivered volume outside of the specification of (b) (4) - (b) (4) mL.
- Observation 8
Inadequate design verification procedures
Procedures for design verification have not been adequately established. Specifically, A. Your firm has not developed design verification testing that adequately tests that the design outputs meet the design inputs, specifically in regards to the reliability of the device at the end of the labeled date of expiry. No verification testing exists that is reflective of the design input requirement that the drug delivery performance must be acceptable at the end of product expiry. B. The test methods for design verification testing are inadequate. (b) (4) of the final product is not done (b) (4) prior to functional verification testing according to your document PR-14719 titled 'Final Report - Component and Functional Qualification of the NGA EpiPen Sr. Auto-Injector using (b) (4) Plungers from (b) (4)' for the (b) (4). The same is true for the completed verification testing of the EpiPen and EpiPen Jr. products. C. Verification testing demonstrated failures of design requirements according to PR-14719; however, these failures were deemed acceptable within the test report conclusions. Your firm confirmed that the same test failures would not be acceptable during lot release testing and would trigger the initiation of an investigation. For example, in PR-14719 one unit of the (b) (4) units tested during the (b) (4) test was noted to have a delivered volume of (b) (4) ml, which is outside of the delivered volume specification of (b) (4) - (b) (4) ml. The conclusion of the test was that it was acceptable to have up to (b) (4) delivered volume failures per the specification of (b) (4) - (b) (4) ml and still pass the acceptance criteria. D. Essential performance requirements such as (b) (4), activation force, dispensing time, and dispensing volume are listed to have a reliability of (b) (4)% in PR-14719 which is not commensurate with the listed design input requirements of PRD/TRD 16-001 Rev 1, nor is it commensurate with the risk associated with the intended use of the product.
- Observation 9
Inadequate design validation procedures
Procedures for design validation have not been adequately established. Specifically, A. Your firm has not completed any validation testing of the design of the device in order to ensure that the devices conform to the defined intended uses. No validation testing was conducted to establish the device specifications conform with the intended use of the product to a degree of reliability commensurate with the risk associated with the device (e.g. no failure rate at product expiry has been established through design validation). Actual or simulated use conditions (e.g. activation orientation, environmental temperature, injection through clothing, etc.) are not part of the design validation plan. Validation testing has not been conducted on initial production units, lots, or batches for the EpiPen, EpiPen Jr., or (b) (4) products. B. No validation testing has been conducted on the user interface of the EpiPen, EpiPen Jr., or (b) (4) products. C. No risk analysis was conducted for the 'basic unit' subassembly of the EpiPen product line. Also, the document titled 'NGA Truject Autoinjector Risk Assessment' reference 8637935n FMEA Report has not been reviewed since its issuance in 2009.
- Observation 10
Deficient laboratory controls for functionality testing
Laboratory controls do not include the establishment of scientifically sound and appropriate specifications, standards, sampling plans and test procedures designed to assure that components, drug product containers, closures, in-process materials, labeling and drug products conform to appropriate standards of identity, strength, quality and purity. Specifically, we witnessed functionality testing of EpiPen and observed the following: A. The lot release test fixtures, which are used to perform the functionality testing of the final finished product, does not include assurance that all test surfaces are level. We also observed that the employee conducting the testing did not (b) (4) to ensure that the (b) (4) was not affected by the environment during testing. The (b) (4) the products during testing appeared damaged and currently there is no standard operating procedure for conducting preventive maintenance on this equipment. B. We observed that drug product was accumulated on the (b) (4). We observed that the drug product inside the (b) (4) from previous testing can be transferred to the final finished product's (b) (4) after dispensing the drug product and prior to weighing the unit for the delivered dose measurement. C. (b) (4) entry of (b) (4) and dispensed time relies on the operator to read the (b) (4) (b) (4) from a (b) (4) whereby the measurement of (b) (4) occurs (b) (4) (b) (4) of the device occurs. D. Measurement results can be altered if a (b) (4) applied different forces to measurement tools. For example, the (b) (4) could be increased by applying more force to the device while (b) (4). E. There have been no studies conducted to establish inter-operator and inter-test stand variation of the functional test results. Currently, your firm has (b) (4) and up to (b) (4) operators who are trained to conduct these tests. F. The operator conducting functional testing on a production lot of final finished product was not listed as an operator trained to complete the functionality testing of NGA Auto-injectors or Auto-injectors according to your firm's records in document SOP-QLC-SQC-00307 & 00394. Additionally, the same operator is not included in the training record for conducting testing of Power Pak. G. There is no risk assessment, failure mode analysis, or determination of analytical variation conducted for the functional lot release test procedure and testing process.
- Observation 11
QC unit lacks authority over sterility failure response
The quality control unit lacks the responsibility and authority to approve and reject all components, drug product containers, closures, in process materials, packaging material and drug products. Specifically, Quality Assurance lacked a sufficient response to the positive sterility test of EpiPen (Epinephrine Injection, 0.3 mg) Lot 7GM063 that occurred on Friday 17 FEB 2017. The microorganism found was Bacillus cereus, a spore-former. For example, as of Thursday 24 FEB 2017: A. (b) (4) batch operations continued on the Epinephrine line and other lines, with no extra controls mandated by QA to increase scrutiny of the aseptic processing operation. B. Quality Assurance did not mandate a temporary suspension of batch release while the scope and possible cause of a sterility failure was being investigated. C. Your procedure SOP-LAB-MIC-00416 Investigation of a Sterility Test Positive, Tested in the (b) (4), Version 9.0, only requires an evaluation of whether "(b) (4) lots and (b) (4)" should be rejected, rather than evaluating if a larger issue exists on the line that could have broader scope. There is no requirement that your firm considers whether other lots manufactured on the line and in the facility may have been affected by the route of contamination.
- Observation 12
Aseptic and media fill procedures not established
Procedures designed to prevent microbiological contamination of drug products purporting to be sterile are not established and followed. Specifically, A. We observed breaches in aseptic technique on the Epinephrine and ATNAA/DuoDote lines. For example: 1. On 21 FEB 2017 during the filling of Epinephrine Injection 0.3mg, Lot 7GM163, an operator reached over exposed sterile barrels. 2. Your sterile operators introduce sterile barrels to the Epinephrine filling in an area with no barrier protection and completely open to the surrounding room, classified as Grade B. In addition, your smoke studies demonstrated turbulence in this area. 3. During filling operations of ATNAA Lot 7M1131 on 21 FEB 2017, your operators made interventions with their upper torso over units on the ATNAA line in two instances. B. Regarding your media fill program: 1. For ATNAA, an operation that requires (b) (4), the number of units were (b) (4) as seen in the table below. The media fill for the Epinephrine line is also a (b) (4). Product Target Fill Media Fill Media Fill Media Fill ATNAA (b) (4) units (b) (4) units 06 JAN 2017 (b) (4) units 11 NOV 2016 (b) (4) 21 DEC 2015 Epinephrine (b) (4) units (b) (4) units 05 JAN 2017 (b) (4) units 18 NOV 2016 (b) (4) units 22 JAN 2015 2. The number of media fills is not commensurate with the (b) (4) Although your firm operates on (b) (4), only (b) (4) media fills are required (b) (4) for both the Epinephrine and ATNAA lines. Your firm uses an approach that (b) (4) and is intended to cover (b) (4) 3. Your firm aborted media fill batches for ATNAA without the clear justification as required by procedure SOP-QLA-VAL-00020 Media Challenge of Aseptic Processes which states in part, "*** A media fill run shall be aborted (invalidated) only under circumstances in which written procedures require Supporting data and justification shall documented [sic] in such cases***". Your firm had no production procedures that specified the conditions that were used to justify invalidate media fill batches on the (ATNAA) line.
- Observation 13
Production and process control procedures not followed
Written production and process control procedures are not followed in the execution of production and process control functions and documented at the time of performance. Specifically, A. Epinephrine, as stated in the appropriate master batch records, is required to be (b) (4), however, controls within the inspection and assembly area ((b) (4)) are deficient since: 1) Units were observed in the inspection and assembly area being (b) (4) and were not (b) (4) during breaks. 2) Your firm is not currently calculating the (b) (4) during all phases of production nor are you currently comparing the (b) (4) against a specified criteria. 3) Your firm has not measured the impact of (b) (4) and the duration and frequency of (b) (4) in the inspection and assembly area from: i. (b) (4) used as (b) (4) as part of the (b) (4) process for the (b) (4) ii. (b) (4) used as (b) (4) as part of the (b) (4) process in (b) (4) iii. (b) (4) in (b) (4) iv. (b) (4) from (b) (4) of (b) (4) that is used as (b) (4) in order to aid in the inspection at the (b) (4) v. (b) (4) from (b) (4) from (b) (4) vi. (b) (4) from (b) (4) in Assembly Area in (b) (4) 4) Your firm is not currently evaluating the potential impact of epinephrine (b) (4) as part of the overall investigation process. 1. Based upon review of production data, Epinephrine Batch 6GM059 had been inspected (b) (4) in (b) (4) ((b) (4)) and therefore (b) (4) (b) (4) (all units were (b) (4) This batch was subsequently released by the firm; and during commercial distribution received a complaint related to Liquids Atypical Color / Cloudy. The subsequent compliant investigation (MMT Complaint Number 2016-07-0037004-US) did not consider the atypical nature of the inspection process nor the potential (b) (4) as a root cause factor. B. "SOP-QLA-GEN-00802, version 10.0, Management of Changes for Equipment, Facilities and Manufacturing Processes, Effective Date: 23 MAR 2016" fails to require documentation of effectiveness checks for completed change controls. Process validations are completed (b) (4) but are not evaluated or trended to determine impact to the overall manufacturing process; additionally, complete process validations for autoinjector manufacturing processes such as EpiPen and ATNAA/DuoDote have not been revalidated since the initial validations. Risk assessments are not required to be performed for every change. The aforementioned procedure does not define criteria or provide examples indicating when risk assessments are necessary to be completed for the change.
- Observation 14
Lack of required training for visual inspection
Employees engaged in the manufacture, processing, packing and holding of a drug product lack the training required to perform their assigned functions. Specifically, your firm did not conduct training for employees involved in the visual inspection for (b) (4) used in EpiPen manufacturing. For example, your Site Lead reported there is no formal OJT for the visual (b) (4) evaluation (inspection) before the (b) (4) is introduced into the process. There are no job aides to assist operators in identifying defects and there are no visual examples of defects in the procedure. There is also not a defined time frame that the (b) (4) must be evaluated for the following defects: (b) (4) and (b) (4) The evaluation of the (b) (4) is not timed. The operators are not qualified for visual acuity or accuracy in discovering defects for this inspection before performing the (b) (4) evaluation task.
- Not classified yetFDA posted this Form 483; FDAgent has not matched it to a Data Dashboard inspectionNo inspection matchedForm 483 pagePDF
- Not classified yetFDA posted this Form 483; FDAgent has not matched it to a Data Dashboard inspectionNo inspection matchedForm 483 pagePDF
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssuranceCitations not published
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices13 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Investigations of discrepancies, failures
There is a failure to thoroughly review any unexplained discrepancy and the failure of a batch or any of its components to meet any of its specifications whether or not the batch has been already distributed.
Training--operations, GMPs, written procedures
Employees are not given training in the particular operations they perform as part of their function and written procedures required by current good manufacturing practice regulations.
Procedures not in writing, fully followed
The responsibilities and procedures applicable to the quality control unit are not fully followed.
Authority lacking to review records, investigate errors
The quality control unit lacks authority to review production records to assure that no errors have occurred and fully investigate errors that have occurred.
Items to cover on annual reviews
Written procedures are not followed for evaluations done at least annually and including provisions for a review of complaints and investigations conducted for each drug product.
Show 8 more citations
Written record of investigation incomplete
Written records of investigations into unexplained discrepancies and the failure of a batch or any of its components to meet specifications do not always include the conclusions and follow-up.
Procedures for sterile drug products
Procedures designed to prevent microbiological contamination of drug products purporting to be sterile are not followed.
Extent of discrepancy, failure investigations
Investigations of an unexplained discrepancy and a failure of a batch or any of its components to meet any of its specifications did not extend to other drug products that may have been associated with the specific failure or discrepancy.
Control procedures to monitor and validate performance
Control procedures are not established which monitor the output and validate the performance of those manufacturing processes that may be responsible for causing variability in the characteristics of in-process material and the drug product.
Written program not followed
The written stability testing program is not followed.
Complaint Investigation/Follow-Up Findings
Complaint records are deficient in that they do not include the findings of the investigation and follow-up.
Contamination, chemical or physical change, deterioration
An NDA-Field Alert Report was not submitted within three working days of receipt of information concerning significant chemical, physical, or other change or deterioration in a distributed drug product.
SOPs not followed / documented
Written production and process control procedures are not followed in the execution of production and process control functions and documented at the time of performance.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: DevicesCitations not published
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: DevicesCitations not published
- OAISignificant objectionable conditions; FDA recommended regulatory actionDrugs · Drug Quality AssuranceCitations not published
- OAISignificant objectionable conditions; FDA recommended regulatory actionDrugs · Drug Quality AssuranceCitations not published
- OAISignificant objectionable conditions; FDA recommended regulatory actionDrugs · Drug Quality AssuranceCitations not published
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssuranceCitations not published
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Following/documenting laboratory controls
Established laboratory control mechanisms are not followed.
Procedures for sterile drug products
Procedures designed to prevent microbiological contamination of drug products purporting to be sterile are not written and followed.
SOPs not followed / documented
Written production and process control procedures are not followed in the execution of production and process control functions and documented at the time of performance.
Written Procedures Not Followed
Written procedures are not followed for the handling of components.
- NAINo objectionable conditions that need follow-upDrugs · Drug Quality AssuranceNone
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssuranceCitations not published
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Buildings not maintained in good state of repair
Buildings used in the manufacturing and processing of a drug product are not maintained in a good state of repair.
SOPs not followed / documented
Written production and process control procedures are not followed in the execution of production and process control functions.
Procedures for sterile drug products
Procedures designed to prevent microbiological contamination of drug products purporting to be sterile are not followed.
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Procedures for sterile drug products
Procedures designed to prevent microbiological contamination of drug products purporting to be sterile are not followed.
SOPs not followed / documented
Written production and process control procedures are not followed in the execution of production and process control functions and documented at the time of performance.
Personnel dating/signing equipment log
The persons double-checking the cleaning and maintenance are not dating the equipment cleaning and use log.
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssuranceCitations not published
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Five requests to FDA under the Freedom of Information Act named this site, most recently on Aug 12, 2026. Each one asked for a Form 483.
Inspection ending Jun 30, 2026VAI
1 request. In the inspection history above. FDA released records in answer to a request for its Form 483 on Sep 3, 2026. FDAgent holds no copy that FDA posted.
- Aug 12, 2026FDA-FOIA-2026-7776Form 483· 1 of 5 sitesRecords released
Inspection ending Jun 5, 2024VAI
2 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Jun 9, 2026. FDAgent holds no copy that FDA posted.
- Jun 5, 2026FDA-FOIA-2026-5584Form 483Records released
- Jul 24, 2024FDA-FOIA-2024-6517Form 483· 1 of 5 sitesClosed
Inspection date not given
2 requests. FDA released records in answer to a request for its Form 483 + EIR on Jul 20, 2026. FDAgent holds no copy that FDA posted.
- Jul 6, 2026FDA-FOIA-2026-6440Form 483 + EIRRecords released
- Date not givenFDA-FOIA-2026-7591Form 483Records released
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 23 sections of 21 CFR in FDA’s published citations for 7 of this facility’s 19 inspections, with how many of them cited each.
- 21 CFR 211.113Control of microbiological contamination6 of 7
- 21 CFR 211.100Written procedures; deviations4 of 7
- 21 CFR 211.110Sampling and testing of in-process materials and drug products3 of 7
- 21 CFR 211.160General requirements3 of 7
- 21 CFR 211.192Production record review3 of 7
- 21 CFR 211.166Stability testing2 of 7
- 21 CFR 211.198Complaint files2 of 7
- 21 CFR 314.81Other postmarketing reports2 of 7
Show all 23
- 21 CFR 820.30Design controls2 of 7
- 21 CFR 820.50Purchasing controls2 of 7
- 21 CFR 820.100Corrective and preventive action2 of 7
- 21 CFR 211.22Responsibilities of quality control unit1 of 7
- 21 CFR 211.25Personnel qualifications1 of 7
- 21 CFR 211.58Maintenance1 of 7
- 21 CFR 211.68Automatic, mechanical, and electronic equipment1 of 7
- 21 CFR 211.80General requirements1 of 7
- 21 CFR 211.165Testing and release for distribution1 of 7
- 21 CFR 211.180General requirements1 of 7
- 21 CFR 211.182Equipment cleaning and use log1 of 7
- 21 CFR 820.75Process validation1 of 7
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 7
- 21 CFR 820.90Nonconforming product1 of 7
- 21 CFR 820.198Complaint files1 of 7
Observations read from posted 483 PDFs carry no CFR section of FDA’s, so they are not counted here.
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDER · DrugsFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
FDA drug establishment registration
FDA registration search- Registered operations
- Analysis
- Registration runs through
- Dec 31, 2027
- Registrant
- Meridian Medical Technologies, LLCD-U-N-S 049504624
From FDA’s Drug Establishments Current Registration Site, as of Nov 21, 2026. A registration says what a site told FDA it does. It is not an inspection result. FDA’s registration search looks firms up by name.
Other sites with the same registrant
FDA’s registration file names Meridian Medical Technologies, LLC (D-U-N-S 049504624) as the registrant for this site and the one other below. That exact D-U-N-S number is the only link. A registrant is the firm that filed the registration, which is not always the owner.
- Meridian Medical Technologies, LLCBridgeton, Missouri · FEI 3033729014Registered; no FDA inspection on file
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. This site’s records are filed under FEI 1950222 in FDA’s data; other sites under its registrant are listed in summary only. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every inspection and enforcement record on this page is filed under FEI 1950222. Other sites filed under the same registrant D-U-N-S are listed in summary only. Records filed under another FEI or name are otherwise not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.