Skip to content

21 CFR 211.22: Responsibilities of quality control unit

FDA's inspection data lists 3,495 Form 483 citations of 21 CFR 211.22 at 2,404 facilities, from October 2008 to September 23, 2026. 275 warning letters state a violation of it.

Counts: FDA Data Dashboard citations, last synced Nov 23, 2026, 04:25 UTC, and FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Regulation text: eCFR. FDAgent is independent of FDA.

Form 483 citations
3,495
At 3,213 inspections
Facilities cited
2,404
Distinct FEIs
Most recent citation
Sep 23, 2026
Baxter Healthcare Corporation
Warning letters
275
State a violation of it

The regulation

The current text of 21 CFR 211.22 from the Electronic Code of Federal Regulations.

§ 211.22 Responsibilities of quality control unit.

(a) There shall be a quality control unit that shall have the responsibility and authority to approve or reject all components, drug product containers, closures, in-process materials, packaging material, labeling, and drug products, and the authority to review production records to assure that no errors have occurred or, if errors have occurred, that they have been fully investigated. The quality control unit shall be responsible for approving or rejecting drug products manufactured, processed, packed, or held under contract by another company.

(b) Adequate laboratory facilities for the testing and approval (or rejection) of components, drug product containers, closures, packaging materials, in-process materials, and drug products shall be available to the quality control unit.

(c) The quality control unit shall have the responsibility for approving or rejecting all procedures or specifications impacting on the identity, strength, quality, and purity of the drug product.

(d) The responsibilities and procedures applicable to the quality control unit shall be in writing; such written procedures shall be followed.

21 CFR Part 211: Current Good Manufacturing Practice for Finished Pharmaceuticals, Subpart BRead on eCFR

Citations by year

Form 483 citations of 21 CFR 211.22 by the year the inspection ended. FDA's citation data starts with inspections ending in October 2008, and the current year runs to the latest sync, so both are partial years (lighter bars).

Peak: 269 in 2025

Form 483 citations of 21 CFR 211.22 by inspection year
YearForm 483 citations
200832
2009203
2010225
2011247
2012212
2013179
2014178
2015169
2016185
2017257
2018231
2019232
202060
2021126
2022181
2023175
2024226
2025269
2026108

How FDA words this citation

FDA's short descriptions and citation text on these 3,495 citations, most used first. These are FDA's standard summaries in its citation data, not the text of any one Form 483.

Short descriptions

  • Procedures not in writing, fully followed2,419 citations
  • Lack of quality control unit474 citations
  • Approve or reject procedures or specs175 citations
  • Authority lacking to review records, investigate errors160 citations
  • Approve or reject components, products139 citations
Show 2 more
  • Adequate lab facilities not available65 citations
  • Contract drug products--lack of responsibility63 citations

Citation text

  • “The responsibilities and procedures applicable to the quality control unit are not fully followed.”992 citations
  • “The responsibilities and procedures applicable to the quality control unit are not in writing and fully followed.”941 citations
  • “There is no quality control unit.”474 citations
  • “The responsibilities and procedures applicable to the quality control unit are not in writing.”466 citations
Show 2 more
  • “The quality control unit lacks responsibility to approve and reject all procedures or specifications impacting on the identity, strength, quality and purity of drug products.”105 citations
  • “The quality control unit lacks authority to fully investigate errors that have occurred.”71 citations

Facilities cited most recently

The latest inspections where FDA cited 21 CFR 211.22, by inspection end date, with FDA's short description. Each links to the facility's FDA record on FDAgent.

Show 4 more

What the observations say

Excerpts from Form 483s that FDA posted in its reading room. A Form 483 usually doesn't name the regulation, so FDAgent ties an observation to 21 CFR 211.22 only when it opens with one of FDA's standard citation texts for this section, word for word apart from case, spacing and punctuation. 94 observations tie this way. The highlighted sentence is the matched wording.

  • Glenmark Pharmaceuticals Limited2025-02-14 · Observation 5

    The quality control unit lacks responsibility to approve all procedures or specifications impacting on the quality of drug products. Specifically, Since 01 Jan 2023, approximately [illegible] US commercial samples [illegible] completed in a timely manner. These testing delays at [illegible] stations as shown below: This delay is potentially causing several commercial batches to be in the market without having concurrently [illegible] supporting …

    This observation opens with FDA's standard wording for 21 CFR 211.22(c).FDA also lists this citation for this inspection.

  • Hikma Pharmaceuticals USA Inc.2024-06-12 · Observation 3

    The responsibilities and procedures applicable to the quality control unit are not fully followed. Specifically, your firm did not follow the instructions outlined in your procedure, SOP CHH15028 Manufacturing Investigation Report (MIR) System. According to section 7.4.2.4 of SOP CHH15028, "when performing an investigation, if a review of previous lot data, stability and retain samples, customer complaints, Annual Product Reviews, and other …

    This observation opens with FDA's standard wording for 21 CFR 211.22(d).FDA also lists this citation for this inspection.

  • LeeSar Inc2026-09-23 · Observation 3

    The responsibilities and procedures applicable to the quality control unit are not in writing and fully followed. Specifically, A. Your firm's Quality Unit (QU) failed to exercise adequate oversight and did not implement appropriate corrective and preventive actions in response to microbial contamination recovered within the (b)(4) ISO 5 containment cell. On 10/25/2025, surface sample (b)(4) from (b)(4) yielded 1 CFU identified as Serratia …

    This observation opens with FDA's standard wording for 21 CFR 211.22(d).

Show 2 more
  • IntegraDose Compounding Services LLC2026-08-12 · Observation 2

    The responsibilities and procedures applicable to the quality control unit are not in writing and fully followed. Specifically, A. Your firm qualifies visual inspectors and AQL inspectors on (b)(4) basis. The following deficiencies in inspector qualification were identified during the inspection: 1. Following the start of operations at your new facility on February 2, 2026, your firm did not perform requalification of visual inspectors or AQL …

    This observation opens with FDA's standard wording for 21 CFR 211.22(d).

  • Cascade Specialty Pharmacy LLC2026-07-17 · Observation 1

    The responsibilities and procedures applicable to the quality control unit are not fully followed. Specifically, Your firm's written procedure designates the pharmacist as responsible for ensuring the accuracy of compounding records and performing quality checks requiring pharmacist verification. However, your firm's compounding record does not specify container and closure to use to ensure they are adequate for its intended pharmaceutical …

    This observation opens with FDA's standard wording for 21 CFR 211.22(d).

Warning letters stating a violation of 21 CFR 211.22

275 warning letters FDA posted state a violation of this section in a numbered violation sentence. The most recent are below, each linking to FDAgent's page for the letter.

“Your firm’s quality unit failed to exercise its responsibility to ensure drug products manufactured are in compliance with CGMP and meet established specifications for identity, strength, quality, and purity (21 CFR 211.22).”

Each letter below states the violation in these words.

“Your firm’s quality control unit failed to exercise its responsibility to ensure drug products manufactured are in compliance with CGMP, and meet established specifications for identity, strength, quality, and purity (21 CFR 211.22).”
“Your firm failed to establish an adequate quality control unit with the responsibility and authority to approve or reject all components, drug product containers, closures, in-process materials, packaging materials, labeling, and drug products (21 CFR 211.22(a)).”

Often cited at the same inspection

Of the 3,213 inspections where FDA cited 21 CFR 211.22, the share where it also cited each of these sections.

Ask FDAgent about 21 CFR 211.22

FDAgent reads FDA's Form 483 citations, observations and warning letters and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • Citation counts come from FDA's Data Dashboard, which lists the regulation behind each Form 483 observation for inspections since October 2008.
  • Observation excerpts come from Form 483 PDFs in FDA's electronic reading room, tied to this section only by exact wording as described above.
  • A warning letter counts for this section when one of its numbered violation sentences cites 21 CFR 211.22; FDAgent quotes that sentence exactly.
  • FDAgent rebuilds this page every week from FDA's public data and is independent of FDA.