NuVasive Specialized Orthopedics, Inc.
101 Enterprise Ste 100, Aliso Viejo, California, United States
FEI 3006179046
FDA has 2 inspections on file for NuVasive Specialized Orthopedics, Inc. in Aliso Viejo, California (FEI 3006179046), from 2014 to 2024. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 12 citations from this inspection: 11 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 803, the device adverse event reporting rules. They are listed below. One FOIA request has asked FDA for this inspection's 483.
No warning letter, court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 2, 2014–2024
- 0 OAI · 1 VAI · 1 NAI
- Inspections with 483 observations shown
- 1 of 2
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 1 request, latest Aug 22, 2024
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices12 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design transfer - Lack of or inadequate procedures
Procedures for design transfer have not been adequately established.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been adequately established.
Show 7 more citations
Design validation - production units
The design was not validated.
Design verification - output does not meet input requirement
Design verification does not confirm that design output meets design input requirements.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Acceptance status
The acceptance status of product was not maintained.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Design output - Lack of or inadequate procedures
Procedures for design output have not been adequately established.
Design Validation - Risk analysis not performed/inadequate
Risk analysis is inadequate.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
One request to FDA under the Freedom of Information Act named this site, on Aug 22, 2024. It asked for a Form 483.
Inspection ending May 6, 2024VAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Aug 22, 2024FDA-FOIA-2024-7441Form 483· 1 of 3 sitesClosed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 7 sections of 21 CFR in FDA’s published citations for 1 of this facility’s 2 inspections.
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?1 of 1
- 21 CFR 820.25Personnel1 of 1
- 21 CFR 820.30Design controls1 of 1
- 21 CFR 820.75Process validation1 of 1
- 21 CFR 820.86Acceptance status1 of 1
- 21 CFR 820.90Nonconforming product1 of 1
- 21 CFR 820.100Corrective and preventive action1 of 1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3006179046 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3006179046. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.