Skip to content

NeuroCareAI, Inc. dba SaveLife AI, Inc.

1740 Lonesome Dove Dr, Prosper, Texas, United States

FEI 3034656245

FDA has one inspection on file for NeuroCareAI, Inc. dba SaveLife AI, Inc. in Prosper, Texas (FEI 3034656245). It ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 4 citations from this inspection: 3 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 801, the device labeling rules. They are listed below. Two FOIA requests have asked FDA for this inspection's 483.

No warning letter, court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
1, in 2025
Inspections with 483 observations shown
1 of 1
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
2 requests, latest Apr 1, 2026
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Evaluation of suppliers, contractors, etc., requirements

      Requirements that must be met by suppliers, contractors and consultants have not been adequately established.

    2. Labeling requirements for stand-alone software

      The stand-alone software that is regulated as a medical device did not provide its unique device identifier through either an easily readable plain-text statement displayed whenever the software is started or an easily readable plain-text statement displayed through a menu command.

    3. Quality Audit/Reaudit - conducted

      Quality audits and reaudits have not been performed.

    4. Lack of or inadequate instructions

      Installation instructions, inspection instructions and test procedures have not been adequately established.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Two requests to FDA under the Freedom of Information Act named this site, most recently on Apr 1, 2026. Each one asked for a Form 483.

  • Inspection ending Dec 12, 2025VAI

    2 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Apr 29, 2026. FDAgent holds no copy that FDA posted.

    1. Apr 1, 2026FDA-FOIA-2026-3156Form 483Records released
    2. Feb 23, 2026FDA-FOIA-2026-1869Form 483· 1 of 5 sitesRecords released

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Regulations FDA cited here

The 4 sections of 21 CFR in FDA’s published citations for this facility’s 1 inspection.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3034656245 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3034656245. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.