21 CFR 801.50: Labeling requirements for stand-alone software

FDA's inspection data lists 4 Form 483 citations of 21 CFR 801.50 at 4 facilities, from April 2024 to December 12, 2025.

Counts: FDA Data Dashboard citations, last synced Nov 23, 2026, 04:25 UTC, and FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Regulation text: eCFR. FDAgent is independent of FDA.

Form 483 citations
4
At 4 inspections
Facilities cited
4
Distinct FEIs
Most recent citation
Dec 12, 2025
NeuroCareAI, Inc. dba SaveLife AI, Inc.
Warning letters
0
State a violation of it

The regulation

The current text of 21 CFR 801.50 from the Electronic Code of Federal Regulations.

§ 801.50 Labeling requirements for stand-alone software.

(a) Stand-alone software that is not distributed in packaged form (e.g., when downloaded from a Web site) is deemed to meet the UDI labeling requirements of this subpart if it complies with the requirements of paragraph (b) of this section and conveys the version number in its production identifier.

(b) Regardless of whether it is or is not distributed in packaged form, stand-alone software regulated as a medical device must provide its unique device identifier through either or both of the following:

(1) An easily readable plain-text statement displayed whenever the software is started;

(2) An easily readable plain-text statement displayed through a menu command (e.g., an “About * * *” command).

(c) Stand-alone software that is distributed in both packaged form and in a form that is not packaged (e.g., when downloaded from a Web site) may be identified with the same device identifier.

21 CFR Part 801: Labeling, Subpart BRead on eCFR

Citations by year

Form 483 citations of 21 CFR 801.50 by the year the inspection ended. FDA's citation data starts with inspections ending in October 2008, and the current year runs to the latest sync, so both are partial years (lighter bars).

Peak: 3 in 2024

Form 483 citations of 21 CFR 801.50 by inspection year
YearForm 483 citations
20243
20251

How FDA words this citation

FDA's short descriptions and citation text on these 4 citations, most used first. These are FDA's standard summaries in its citation data, not the text of any one Form 483.

Short descriptions

  • Labeling requirements for stand-alone software4 citations

Citation text

  • “The stand-alone software that is regulated as a medical device did not provide its unique device identifier through either an easily readable plain-text statement displayed whenever the software is started or an easily readable plain-text statement displayed through a menu command.”4 citations

Facilities cited most recently

The latest inspections where FDA cited 21 CFR 801.50, by inspection end date, with FDA's short description. Each links to the facility's FDA record on FDAgent.

Often cited at the same inspection

Of the 4 inspections where FDA cited 21 CFR 801.50, the share where it also cited each of these sections.

Ask FDAgent about 21 CFR 801.50

FDAgent reads FDA's Form 483 citations, observations and warning letters and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • Citation counts come from FDA's Data Dashboard, which lists the regulation behind each Form 483 observation for inspections since October 2008.
  • Observation excerpts come from Form 483 PDFs in FDA's electronic reading room, tied to this section only by exact wording as described above.