Emerald Supply Inc
68 35th St Unit 5, Brooklyn, New York, United States
FEI 3011714211
FDA has one inspection on file for Emerald Supply Inc in Brooklyn, New York (FEI 3011714211). It ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 8 citations from this inspection. They are listed below. Two FOIA requests have asked FDA for this inspection's 483.
No warning letter, court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 1, in 2026
- Inspections with 483 observations shown
- 1 of 1
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 2 requests, latest Jul 16, 2026
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices8 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Approval, inadequate purchasing data
Purchasing data that clearly describe or reference specified requirements for purchased or otherwise received product and services have not been approved and established.
- ISO 13485:2016 Clause 7.1
Records of risk management activities in product realization not maintained
Records of risk management activities in product realization were not maintained.
- ISO 13485:2016 Clause 5.1
Lack of evidence for top management's commitment to QMS
Top management did not provide evidence of its commitment to the development and implementation of the quality management system and maintenance of its effectiveness by communicating to the organization the importance of meeting requirements, establishing the quality policy, ensuring the quality objectives are established, conducting management reviews and ensuring the availability of resources.
Lack of Written MDR Procedures
Written MDR procedures have not been developed, maintained and implemented.
- ISO 13485:2016 Clause 4.2.1
Quality management system documentation not adequate
The quality management system documentation does not include statements of a quality policy and quality objectives and a quality manual.
Show 3 more citations
Quality System Procedures
Quality system procedures and instructions have not been established.
Records for complaints involving failure to meet specifications not maintained
Records of the review, evaluation, and investigation for any complaints involving the possible failure of a device, labeling, or packaging to meet any of its specifications were not maintained.
Evaluation and Selection, Suppliers, Contractors, etc.
Potential suppliers and contractors were not evaluated and selected based on their ability to meet specified requirements.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Two requests to FDA under the Freedom of Information Act named this site, most recently on Jul 16, 2026. Each one asked for a Form 483.
Inspection ending Apr 16, 2026VAI
2 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Sep 28, 2026. FDAgent holds no copy that FDA posted.
- Jul 16, 2026FDA-FOIA-2026-6764Form 483· 1 of 3 sitesRecords released
- Jul 7, 2026FDA-FOIA-2026-6454Form 483· 1 of 4 sitesRecords released
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 4 sections of 21 CFR in FDA’s published citations for this facility’s 1 inspection.
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 1
- 21 CFR 820.20Management responsibility1 of 1
- 21 CFR 820.35Control of records1 of 1
- 21 CFR 820.50Purchasing controls1 of 1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3011714211 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3011714211. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.