21 CFR 820.35: Control of records

FDA's inspection data lists 19 Form 483 citations of 21 CFR 820.35 at 17 facilities, from February 2026 to October 1, 2026. 1 warning letter states a violation of it.

Counts: FDA Data Dashboard citations, last synced Nov 23, 2026, 04:25 UTC, and FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Regulation text: eCFR. FDAgent is independent of FDA.

Form 483 citations
19
At 17 inspections
Facilities cited
17
Distinct FEIs
Most recent citation
Oct 1, 2026
Jazz Imaging
Warning letters
1
State a violation of it

The regulation

The current text of 21 CFR 820.35 from the Electronic Code of Federal Regulations.

§ 820.35 Control of records.

In addition to the requirements of Clause 4.2.5 in ISO 13485 (incorporated by reference, see § 820.7), Control of Records, the manufacturer must include the following information in certain records:

(a) Records of complaints. In addition to Clause 8.2.2 in ISO 13485, Complaint Handling, the manufacturer shall maintain records of the review, evaluation, and investigation for any complaints involving the possible failure of a device, labeling, or packaging to meet any of its specifications. If an investigation has already been performed for a similar complaint, another investigation is not necessary, and the manufacturer shall maintain records documenting justification for not performing such investigation. For complaints that must be reported to FDA under part 803 of this chapter, complaints that a manufacturer determines must be investigated, and complaints that the manufacturer investigated regardless of those requirements, the manufacturer must record the following information:

(1) The name of the device;

(2) The date the complaint was received;

(3) Any unique device identifier (UDI) or universal product code (UPC), and any other device identification(s);

(4) The name, address, and phone number of the complainant;

(5) The nature and details of the complaint;

(6) Any correction or corrective action taken; and

(7) Any reply to the complainant.

(b) Records of servicing activities. In adhering to Clause 7.5.4 in ISO 13485, Servicing Activities, the manufacturer must record the following information, at a minimum, for servicing activities:

(1) The name of the device serviced;

(2) Any UDI or UPC, and any other device identification(s);

(3) The date of service;

(4) The individual(s) who serviced the device;

(5) The service performed; and

(6) Any test and inspection data.

(c) Unique Device Identification. In addition to the requirements of Clauses 7.5.1, 7.5.8, and 7.5.9 in ISO 13485, the UDI must be recorded for each medical device or batch of medical devices.

(d) Confidentiality. Records deemed confidential by the manufacturer may be marked to aid FDA in determining whether information may be disclosed under the public information regulation in part 20 of this chapter.

21 CFR Part 820: Quality Management System Regulation, Subpart BRead on eCFR

Citations by year

Form 483 citations of 21 CFR 820.35 by the year the inspection ended. FDA's citation data starts with inspections ending in October 2008, and the current year runs to the latest sync, so both are partial years (lighter bars).

Peak: 19 in 2026

Form 483 citations of 21 CFR 820.35 by inspection year
YearForm 483 citations
202619

How FDA words this citation

FDA's short descriptions and citation text on these 19 citations, most used first. These are FDA's standard summaries in its citation data, not the text of any one Form 483.

Short descriptions

  • Records for complaints involving failure to meet specifications not maintained6 citations
  • UDI was not recorded for each medical device or batch6 citations
  • Records of complaint investigations do not include required information4 citations
  • Records of servicing activities do not include required information2 citations
  • Lack of justification for not investigating complaints requiring investigation1 citations

Citation text

  • “Unique device identification (UDI) was not recorded for each medical device or batch of medical devices.”5 citations
  • “Records of the review, evaluation, and investigation for any complaints involving the possible failure of a device, labeling, or packaging to meet any of its specifications were not adequately maintained.”3 citations
  • “Records of the review, evaluation, and investigation for any complaints involving the possible failure of a device, labeling, or packaging to meet any of its specifications were not maintained.”3 citations
  • “Records of complaint investigations do not include any unique device identifier (UDI) or universal product code (UPC) and any other device identification(s).”2 citations
Show 2 more
  • “Justification for not investigating complaint(s) involving the possible failure of a device, labeling, or packaging to meet any of its specifications was not adequately documented.”1 citations
  • “Records of complaint investigations do not include the name of the device, any unique device identifier (UDI) or universal product code (UPC) and any other device identification(s), the name, address, and phone number of the complainant and any reply to the complainant.”1 citations

Facilities cited most recently

The latest inspections where FDA cited 21 CFR 820.35, by inspection end date, with FDA's short description. Each links to the facility's FDA record on FDAgent.

  • Jazz Imaging2026-10-01 · FEI 3013503781Records of complaint investigations do not include required information
  • Walz Elektronik Gmbh2026-09-10 · FEI 1000161819UDI was not recorded for each medical device or batch
  • Superior Labs Inc2026-08-11 · FEI 3034286079UDI was not recorded for each medical device or batch
  • United Dental Resources Corporation2026-08-07 · FEI 3015194291Records of complaint investigations do not include required information
  • Metropolis International LLC2026-08-06 · FEI 3005877899Records of servicing activities do not include required information
  • Fiberoptics Technology, Inc.2026-07-24 · FEI 1000121058UDI was not recorded for each medical device or batch
Show 4 more
  • FIGUR8, INC.2026-07-21 · FEI 3034676730UDI was not recorded for each medical device or batch
  • MODERN MEDICAL SUPPLIES LLC2026-06-12 · FEI 3015374852Records for complaints involving failure to meet specifications not maintained
  • Hugh Steeper Ltd2026-05-28 · FEI 3017697481Records of servicing activities do not include required information
  • Domico Med-Device Llc2026-05-12 · FEI 1825560Records of complaint investigations do not include required information

Warning letters stating a violation of 21 CFR 820.35

1 warning letter FDA posted states a violation of this section in a numbered violation sentence. The most recent are below, each linking to FDAgent's page for the letter.

“Failure to maintain records of the review, evaluation, and investigation for any complaints involving the possible failure of a device, labeling, or packaging to meet any of its specifications, as required by 21 CFR 820.35(a).”

Often cited at the same inspection

Of the 17 inspections where FDA cited 21 CFR 820.35, the share where it also cited each of these sections.

  • 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?7 (41%)
  • 21 CFR 801.20Label to bear a unique device identifier1 (6%)
  • 21 CFR 801.40Form of a unique device identifier1 (6%)
  • 21 CFR 803.40If I am an importer, what reporting requirements apply to me?1 (6%)
  • 21 CFR 820.20Management responsibility1 (6%)
  • 21 CFR 820.45Device labeling and packaging controls1 (6%)

Ask FDAgent about 21 CFR 820.35

FDAgent reads FDA's Form 483 citations, observations and warning letters and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • Citation counts come from FDA's Data Dashboard, which lists the regulation behind each Form 483 observation for inspections since October 2008.
  • Observation excerpts come from Form 483 PDFs in FDA's electronic reading room, tied to this section only by exact wording as described above.
  • A warning letter counts for this section when one of its numbered violation sentences cites 21 CFR 820.35; FDAgent quotes that sentence exactly.
  • FDAgent rebuilds this page every week from FDA's public data and is independent of FDA.