Skip to content

Metropolis International LLC

3813 10th St, Long Island City, New York, United States

FEI 3005877899

FDA has 4 inspections on file for Metropolis International LLC in Long Island City, New York (FEI 3005877899), from 2014 to 2026. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 8 citations from this inspection. They are listed below. Two FOIA requests have asked FDA for this inspection's 483.

No warning letter, court action or import refusal is filed under this FEI.

Two regulations were cited at more than one inspection: 21 CFR 803.17 and 21 CFR 820.100(a).

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
4, 2014–2026
0 OAI · 4 VAI · 0 NAI
Inspections with 483 observations shown
4 of 4
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
2 requests, latest Sep 24, 2026
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices8 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of Written MDR Procedures

      Written MDR procedures have not been developed and maintained.

    2. ISO 13485:2016 Clause 5.6.1

      Management review procedures not documented

      Procedures for management review have not been documented.

    3. ISO 13485:2016 Clause 8.5.3

      Preventive action procedures not documented

      Preventive action procedures describing the requirements for determining potential nonconformities and their causes, evaluating the need for action to prevent occurrence of potential nonconformities, planning and documenting action needed and implementing such action, verifying that the action does not adversely affect the ability to meet applicable regulatory requirements or the safety and performance of the medical device and reviewing the effectiveness of the preventive action taken have not been documented.

    4. ISO 13485:2016 Clause 7.5.4

      Servicing procedures and reference materials and measurements not documented

      The servicing procedures, reference materials and reference measurements for performing servicing activities and verifying that product requirements are met were not documented.

    5. Records of servicing activities do not include required information

      Records of servicing activities do not include any test and inspection data and any unique device identifier (UDI) or universal product code (UPC), and any other device identification(s).

    Show 3 more citations
    1. ISO 13485:2016 Clause 8.2.2

      Complaint handling procedures lack required elements

      Procedures for timely complaint handling do not include requirements and responsibilities for receiving and recording information, evaluating information to determine if the feedback constitutes a complaint, investigating complaints, determining the need to report the information to FDA and handling of complaint-related product.

    2. ISO 13485:2016 Clause 8.5.2

      Corrective action procedures not documented

      Corrective action procedures defining the requirements for reviewing nonconformities, determining the causes of nonconformities, evaluating the need for action to ensure that nonconformities do not recur, planning and documenting action needed and implementing such action, verifying that the corrective action does not adversely affect the ability to meet applicable regulatory requirements or the safety and performance of the medical device and reviewing the effectiveness of corrective action taken have not been documented.

    3. ISO 13485:2016 Clause 8.2.4

      Internal audit procedures not documented

      Procedures to describe the responsibilities and requirements for planning and conducting internal audits and recording and reporting audit results were not documented.

  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Process control procedures, Lack of or inadequate procedures

      Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.

    2. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.

    3. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been established.

    4. Complaints

      Complaint files are not adequately maintained.

  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of Written MDR Procedures

      Written MDR procedures have not been developed.

    2. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been established.

    3. Equipment control activity documentation

      Equipment calibrations have not been documented.

  4. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been established.

    2. Lack of Written MDR Procedures

      Written MDR procedures have not been.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Two requests to FDA under the Freedom of Information Act named this site, most recently on Sep 24, 2026. Each one asked for a Form 483.

  • Inspection ending Aug 6, 2026VAI

    2 requests. In the inspection history above.

    1. Sep 24, 2026FDA-FOIA-2026-9250Form 483· 1 of 5 sitesOpen
    2. Sep 22, 2026FDA-FOIA-2026-9179Form 483· 1 of 5 sitesOpen

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Regulations FDA cited here

The 6 sections of 21 CFR in FDA’s published citations for this facility’s 4 inspections, with how many of them cited each.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3005877899 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3005877899. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.