FIGUR8, INC.
2 Park Plz Ste 605, Boston, Massachusetts, United States
FEI 3034676730
FDA has one inspection on file for FIGUR8, INC. in Boston, Massachusetts (FEI 3034676730). It ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 6 citations from this inspection. They are listed below. One FOIA request has asked FDA for this inspection's 483.
No warning letter, court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 1, in 2026
- Inspections with 483 observations shown
- 1 of 1
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 1 request, latest Aug 20, 2026
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
- ISO 13485:2016 Clause 7.3.3
Design and development inputs not determined
Design and development inputs related to product requirements were not adequately determined.
- ISO 13485:2016 Clause 8.2.2
Complaint handling procedures not documented
Procedures for timely complaint handling have not been adequately documented.
- ISO 13485:2016 Clause 8.3.1
Procedure for nonconforming product not documented
Procedures to define the controls and related responsibilities and authorities for the identification, documentation, segregation, evaluation, and disposition of nonconforming product were not documented.
- ISO 13485:2016 Clause 4.1.5
External party controls do not include written quality agreements
Control over outsourced processes that affect product conformity to requirements do not include adequate written quality agreements.
UDI was not recorded for each medical device or batch
Unique device identification (UDI) was not adequately recorded for each medical device or batch of medical devices.
Show 1 more citation
- ISO 13485:2016 Clause 8.5.2
Corrective actions were unduly delayed
Corrective actions taken to eliminate the cause of nonconformities to prevent recurrence were unduly delayed.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
One request to FDA under the Freedom of Information Act named this site, on Aug 20, 2026. It asked for a Form 483.
Inspection ending Jul 21, 2026VAI
1 request. In the inspection history above.
- Aug 20, 2026FDA-FOIA-2026-8077Form 483· 1 of 5 sitesOpen
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 1 section of 21 CFR in FDA’s published citations for this facility’s 1 inspection.
- 21 CFR 820.35Control of records1 of 1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3034676730 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3034676730. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.