Skip to content

Fiberoptics Technology, Inc.

1 Quassett Road, Pomfret, Connecticut, United States

FEI 1000121058

FDA has 3 inspections on file for Fiberoptics Technology, Inc. in Pomfret, Connecticut (FEI 1000121058), from 2017 to 2026. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 6 citations from this inspection. They are listed below. Three FOIA requests have asked FDA for this inspection's 483.

No warning letter, court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
3, 2017–2026
0 OAI · 2 VAI · 1 NAI
Inspections with 483 observations shown
2 of 3
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
3 requests, latest Oct 16, 2026
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. ISO 13485:2016 Clause 7.4.1

      Criteria for the evaluation and selection of suppliers not established

      Criteria for the evaluation and selection of suppliers were not adequately established.

    2. UDI was not recorded for each medical device or batch

      Unique device identification (UDI) was not recorded for each medical device or batch of medical devices.

    3. ISO 13485:2016 Clause 8.2.6

      Evidence of conformity to the acceptance criteria not maintained

      Evidence of conformity to the acceptance criteria for product monitoring and measurement was not maintained.

    4. ISO 13485:2016 Clause 7.3.10

      Design and development file does not include or reference required records

      A design and development file does not include or reference records generated to demonstrate conformity to the requirements for design and development and records for design and development changes.

    5. ISO 13485:2016 Clause 8.2.2

      Complaint handling procedures lack required elements

      Procedures for timely complaint handling do not adequately include requirements and responsibilities for receiving and recording information, evaluating information to determine if the feedback constitutes a complaint, investigating complaints, determining the need to report the information to FDA, handling of complaint-related product and determining the need to initiate corrections or corrective actions.

    Show 1 more citation
    1. ISO 13485:2016 Clause 7.1

      Product realization risk management processes not documented

      One or more processes for risk management in product realization have not been adequately documented.

  2. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    2. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Three requests to FDA under the Freedom of Information Act named this site, most recently on Oct 16, 2026. Each one asked for a Form 483.

  • Inspection ending Jul 24, 2026VAI

    3 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Sep 22, 2026. FDAgent holds no copy that FDA posted.

    1. Oct 16, 2026FDA-FOIA-2026-10084Form 483· 1 of 5 sitesOpen
    2. Aug 26, 2026FDA-FOIA-2026-8267Form 483· 1 of 5 sitesRecords released
    3. Aug 19, 2026FDA-FOIA-2026-8011Form 483Records released

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Regulations FDA cited here

The 3 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 3 inspections, with how many of them cited each.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 1000121058 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 1000121058. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.