Skip to content

21 CFR 820.45: Device labeling and packaging controls

FDA's inspection data lists 1 Form 483 citation of 21 CFR 820.45 at 1 facility, from March 2026 to March 31, 2026.

Counts: FDA Data Dashboard citations, last synced Nov 23, 2026, 04:25 UTC, and FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Regulation text: eCFR. FDAgent is independent of FDA.

Form 483 citations
1
At 1 inspection
Facilities cited
1
Distinct FEIs
Most recent citation
Mar 31, 2026
IVIMEA LLC
Warning letters
0
State a violation of it

The regulation

The current text of 21 CFR 820.45 from the Electronic Code of Federal Regulations.

§ 820.45 Device labeling and packaging controls.

In addition to the requirements of Clause 7.5.1 of ISO 13485 (incorporated by reference, see § 820.7), Control of production and service provision, each manufacturer must document and maintain procedures that provide a detailed description of the activities to ensure the integrity, inspection, storage, and operations for labeling and packaging, during the customary conditions of processing, storage, handling, distribution, and, as appropriate, use of the device.

(a) The manufacturer must ensure labeling and packaging has been examined for accuracy prior to release or storage where applicable, to include the following:

(1) The correct unique device identifier (UDI) or universal product code (UPC), or any other device identification(s);

(2) Expiration date;

(3) Storage instructions;

(4) Handling instructions; and

(5) Any additional processing instructions.

(b) The release of the labeling for use must be documented in accordance with Clause 4.2.5 of ISO 13485.

(c) The manufacturer must ensure labeling and packaging operations have been established and maintained to prevent mixups, including, but not limited to, inspection of the labeling and packaging before use to assure that all devices have correct labeling and packaging, as specified in the medical device file. Results of such labeling inspection must be documented in accordance with Clause 4.2.5 of ISO 13485.

21 CFR Part 820: Quality Management System Regulation, Subpart BRead on eCFR

Citations by year

Form 483 citations of 21 CFR 820.45 by the year the inspection ended. FDA's citation data starts with inspections ending in October 2008, and the current year runs to the latest sync, so both are partial years (lighter bars).

Peak: 1 in 2026

Form 483 citations of 21 CFR 820.45 by inspection year
YearForm 483 citations
20261

How FDA words this citation

FDA's short descriptions and citation text on these 1 citation, most used first. These are FDA's standard summaries in its citation data, not the text of any one Form 483.

Short descriptions

  • Release of labeling for use not documented1 citations

Citation text

  • “The release of labeling was not documented.”1 citations

Facilities cited most recently

The latest inspections where FDA cited 21 CFR 820.45, by inspection end date, with FDA's short description. Each links to the facility's FDA record on FDAgent.

  • IVIMEA LLC2026-03-31 · FEI 3013986441Release of labeling for use not documented

Often cited at the same inspection

Of the 1 inspection where FDA cited 21 CFR 820.45, the share where it also cited each of these sections.

  • 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 (100%)
  • 21 CFR 820.35Control of records1 (100%)
  • 21 CFR 830.310Information required for unique device identification1 (100%)

Ask FDAgent about 21 CFR 820.45

FDAgent reads FDA's Form 483 citations, observations and warning letters and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • Citation counts come from FDA's Data Dashboard, which lists the regulation behind each Form 483 observation for inspections since October 2008.
  • Observation excerpts come from Form 483 PDFs in FDA's electronic reading room, tied to this section only by exact wording as described above.
  • A warning letter counts for this section when one of its numbered violation sentences cites 21 CFR 820.45; FDAgent quotes that sentence exactly.
  • FDAgent rebuilds this page every week from FDA's public data and is independent of FDA.