Skip to content

IVIMEA LLC

251 Pennington Lawrenceville Rd, Pennington, New Jersey, United States

FEI 3013986441

FDA has 2 inspections on file for IVIMEA LLC in Pennington, New Jersey (FEI 3013986441), from 2019 to 2026. The latest ended on and was classified OAI, official action indicated.

OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.

FDA published 12 citations from this inspection. They are listed below. One FOIA request has asked FDA for this inspection's 483.

No warning letter, court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
OAI
Inspections on file
2, 2019–2026
1 OAI · 0 VAI · 1 NAI
Inspections with 483 observations shown
1 of 2
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
1 request, latest Jun 1, 2026
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices12 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Records for complaints involving failure to meet specifications not maintained

      Records of the review, evaluation, and investigation for any complaints involving the possible failure of a device, labeling, or packaging to meet any of its specifications were not maintained.

    2. ISO 13485:2016 Clause 4.1.2

      Risk-based approach not applied to QMS processes

      A risk-based approach to the control of the appropriate processes needed for the quality management system has not been adequately applied.

    3. UDI was not recorded for each medical device or batch

      Unique device identification (UDI) was not recorded for each medical device or batch of medical devices.

    4. Lack of Written MDR Procedures

      Written MDR procedures have not been developed, maintained and implemented.

    5. ISO 13485:2016 Clause 7.4.1

      Supplier performance not monitored

      Supplier performance in meeting requirements for the purchased product was not monitored.

    Show 7 more citations
    1. GUDID Information required for unique device identification.

      The correct unique device identifier (UDI) information regarding proprietary, trade, or brand name of the device as it appears on the label was not provided to the FDA via the Global Unique Device Identification Database (GUDID).

    2. ISO 13485:2016 Clause 7.3.7

      Design and development validation not performed as planned

      Design and development validation was not performed in accordance with planned and documented arrangements to ensure that the resulting product is capable of meeting the requirements for the specified application or intended use.

    3. Release of labeling for use not documented

      The release of labeling was not documented.

    4. ISO 13485:2016 Clause 4.2.5

      Records of conformity and QMS operations not maintained

      Records to provide evidence of conformity to requirements and the effective operation of the quality management system have not been maintained.

    5. ISO 13485:2016 Clause 7.6(a)

      Measuring equipment not adequately calibrated or verified

      Measuring equipment was not calibrated at specified intervals or prior to use.

    6. ISO 13485:2016 Clause 7.3.7

      Records of design and development validation not maintained

      Records of the results and conclusions of design and development validation and necessary actions were not adequately maintained.

    7. ISO 13485:2016 Clause 5.6.1

      QMS not reviewed by top management at documented planned intervals

      Top management did not adequately review the quality management system at documented planned intervals to ensure its continued suitability, adequacy, and effectiveness.

  2. NAINo objectionable conditions that need follow-upDevices · Radiation Control and Health Safety ActNone

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

One request to FDA under the Freedom of Information Act named this site, on Jun 1, 2026. It asked for a Form 483.

  • Inspection ending Mar 31, 2026OAI

    1 request. In the inspection history above. FDA released records in answer to a request for its Form 483 on Sep 14, 2026. FDAgent holds no copy that FDA posted.

    1. Jun 1, 2026FDA-FOIA-2026-5409Form 483· 1 of 5 sitesRecords released

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Regulations FDA cited here

The 4 sections of 21 CFR in FDA’s published citations for 1 of this facility’s 2 inspections.

  • 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 1
  • 21 CFR 820.35Control of records1 of 1
  • 21 CFR 830.310Information required for unique device identification1 of 1
  • 21 CFR 820.45Device labeling and packaging controls1 of 1

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3013986441 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3013986441. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.