Skip to content

21 CFR 830.310: Information required for unique device identification

FDA's inspection data lists 29 Form 483 citations of 21 CFR 830.310 at 28 facilities, from December 2023 to September 17, 2026.

Counts: FDA Data Dashboard citations, last synced Nov 23, 2026, 04:25 UTC, and FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Regulation text: eCFR. FDAgent is independent of FDA.

Form 483 citations
29
At 29 inspections
Facilities cited
28
Distinct FEIs
Most recent citation
Sep 17, 2026
RLS Interventional, Inc.
Warning letters
0
State a violation of it

The regulation

The current text of 21 CFR 830.310 from the Electronic Code of Federal Regulations.

§ 830.310 Information required for unique device identification.

The contact for device identification designated under § 830.320(a) shall provide FDA with the following information concerning each version or model of a device required to bear a unique device identifier (UDI) on its label:

(a) Concerning the labeler:

(1) The name of the labeler;

(2) A telephone number or email address that will allow FDA to communicate with the contact for device identification designated under § 830.320(a); and

(3) The name of each issuing agency whose system is used by the labeler to assign UDIs used by the labeler.

(b) Concerning each version or model of a device with a UDI on its label:

(1) The device identifier portion of the UDI assigned to the version or model;

(2) When reporting a substitution of a new device identifier that will be used in lieu of a previously reported identifier, the device identifier that was previously assigned to the version or model;

(3) If § 801.45 of this chapter requires the device to bear a UDI as a permanent marking on the device itself, either:

(i) A statement that the device identifier that appears as a permanent marking on the device is identical to that reported under paragraph (b)(1) of this section, or

(ii) The device identifier portion of the UDI that appears as a permanent marking on the device;

(4) The proprietary, trade, or brand name of the device as it appears on the label of the device;

(5) Any version or model number or similar reference that appears on the label of the device;

(6) If the device is labeled as sterile, a statement to that effect;

(7) If the device is labeled as containing natural rubber latex that contacts humans, or is labeled as having packaging containing natural rubber latex that contacts humans, as described by §§ 801.437(b)(1), 801.437(b)(3), and 801.437(f) of this chapter, a statement to that effect;

(8) Whether a patient may be safely exposed to magnetic resonance imaging, nuclear magnetic resonance imaging, or magnetic resonance tomography while using the device, or while the device is implanted in patient.

(9) If the device is available in more than one size, the size of the particular version or model, together with the unit of measure, as it appears on the label of the device;

(10) The type of production identifiers that appear on the label of the device;

(11) The FDA premarket submission number of a cleared or approved device, or a statement that FDA has by regulation exempted the device from premarket notification;

(12) The FDA listing number assigned to the device;

(13) The Global Medical Device Nomenclature (GMDN) term or code for the device;

(14) The total number of individual devices contained in the device package.

21 CFR Part 830: Unique Device Identification, Subpart ERead on eCFR

Citations by year

Form 483 citations of 21 CFR 830.310 by the year the inspection ended. FDA's citation data starts with inspections ending in October 2008, and the current year runs to the latest sync, so both are partial years (lighter bars).

Peak: 12 in 2026

Form 483 citations of 21 CFR 830.310 by inspection year
YearForm 483 citations
20231
20245
202511
202612

How FDA words this citation

FDA's short descriptions and citation text on these 29 citations, most used first. These are FDA's standard summaries in its citation data, not the text of any one Form 483.

Short descriptions

  • GUDID Information required for unique device identification.29 citations

Citation text

  • “The correct unique device identifier (UDI) information regarding device identifier portion of the UDI assigned to the version or model was not provided to the FDA via the Global Unique Device Identification Database (GUDID).”17 citations
  • “The correct unique device identifier (UDI) information regarding device identifier portion of the UDI assigned to the version or model, permanent marking requirements and proprietary, trade, or brand name of the device as it appears on the label was not provided to the FDA via the Global Unique Device Identification Database (GUDID).”1 citations
  • “The correct unique device identifier (UDI) information regarding device identifier portion of the UDI assigned to the version or model, proprietary, trade, or brand name of the device as it appears on the label, size of the particular version or model given the device is available in more than one size and the FDA premarket submission number of a cleared or approved device was not provided to the FDA via the Global Unique Device Identification Database (GUDID).”1 citations
  • “The correct unique device identifier (UDI) information regarding device identifier portion of the UDI assigned to the version or model, proprietary, trade, or brand name of the device as it appears on the label, size of the particular version or model given the device is available in more than one size, the type of production identifiers that appears on the label of the device, the FDA premarket submission number of a cleared or approved device and the Global Medical Device Nomenclature (GMDN) term or code for the device was not provided to the FDA via the Global Unique Device Identification Database (GUDID).”1 citations
Show 2 more
  • “The correct unique device identifier (UDI) information regarding device identifier portion of the UDI assigned to the version or model, proprietary, trade, or brand name of the device as it appears on the label, size of the particular version or model given the device is available in more than one size, the type of production identifiers that appears on the label of the device, the FDA premarket submission number of a cleared or approved device, the Global Medical Device Nomenclature (GMDN) term or code for the device and the total number of individual devices contained in the device package was not provided to the FDA via the Global Unique Device Identification Database (GUDID).”1 citations
  • “The correct unique device identifier (UDI) information regarding device identifier portion of the UDI assigned to the version or model, proprietary, trade, or brand name of the device as it appears on the label, version or model or similar reference that appears on the label of the device and the type of production identifiers that appears on the label of the device was not provided to the FDA via the Global Unique Device Identification Database (GUDID).”1 citations

Facilities cited most recently

The latest inspections where FDA cited 21 CFR 830.310, by inspection end date, with FDA's short description. Each links to the facility's FDA record on FDAgent.

  • RLS Interventional, Inc.2026-09-17 · FEI 3001237677GUDID Information required for unique device identification.
  • FKA Distributing Co., LLC2026-09-01 · FEI 1832894GUDID Information required for unique device identification.
  • GUIDEMIA TECHNOLOGIES LLC2026-07-06 · FEI 3013634985GUDID Information required for unique device identification.
  • QUANTA DIALYSIS TECHONLOGIES LIMITED2026-06-11 · FEI 3013536188GUDID Information required for unique device identification.
  • INFLAMMATIX INC2026-04-30 · FEI 3022977716GUDID Information required for unique device identification.
  • Tesla Biohealing, Inc.2026-04-16 · FEI 3023229437GUDID Information required for unique device identification.
Show 4 more
  • ConMed Corporation2026-04-06 · FEI 3004491689GUDID Information required for unique device identification.
  • IVIMEA LLC2026-03-31 · FEI 3013986441GUDID Information required for unique device identification.
  • Consolidated Research, Inc.2026-03-25 · FEI 3009770872GUDID Information required for unique device identification.
  • Limber Prosthetics and Orthotics, Inc2026-03-20 · FEI 3030088168GUDID Information required for unique device identification.

Often cited at the same inspection

Of the 29 inspections where FDA cited 21 CFR 830.310, the share where it also cited each of these sections.

Ask FDAgent about 21 CFR 830.310

FDAgent reads FDA's Form 483 citations, observations and warning letters and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • Citation counts come from FDA's Data Dashboard, which lists the regulation behind each Form 483 observation for inspections since October 2008.
  • Observation excerpts come from Form 483 PDFs in FDA's electronic reading room, tied to this section only by exact wording as described above.
  • A warning letter counts for this section when one of its numbered violation sentences cites 21 CFR 830.310; FDAgent quotes that sentence exactly.
  • FDAgent rebuilds this page every week from FDA's public data and is independent of FDA.