Skip to content

Consolidated Research, Inc.

26250 Euclid Ave Ste 709, Euclid, Ohio, United States

FEI 3009770872

FDA has 3 inspections on file for Consolidated Research, Inc. in Euclid, Ohio (FEI 3009770872), from 2013 to 2026. The latest ended on and was classified NAI, no action indicated.

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

FDA published 7 citations from this inspection. They are listed below. Two FOIA requests have asked FDA for this inspection's 483.

No warning letter, court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
3, 2013–2026
0 OAI · 0 VAI · 3 NAI
Inspections with 483 observations shown
1 of 3
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
2 requests, latest Aug 13, 2026
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices7 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. ISO 13485:2016 Clause 4.1.2

      Risk-based approach not applied to QMS processes

      A risk-based approach to the control of the appropriate processes needed for the quality management system has not been adequately applied.

    2. ISO 13485:2016 Clause 8.5.2

      Corrective action procedures not documented

      Corrective action procedures defining the requirements for evaluating the need for action to ensure that nonconformities do not recur and planning and documenting action needed and implementing such action have not been documented.

    3. GUDID Information required for unique device identification.

      The correct unique device identifier (UDI) information regarding device identifier portion of the UDI assigned to the version or model was not provided to the FDA via the Global Unique Device Identification Database (GUDID).

    4. ISO 13485:2016 Clause 8.3.4

      Rework records not maintained

      Records of rework were not adequately maintained.

    5. ISO 13485:2016 Clause 7.4.1

      Records of supplier (re)evaluation, selection, and monitoring not maintained

      Records of the results of selection, monitoring and re-evaluation of supplier capability or performance were not adequately maintained.

    Show 2 more citations
    1. ISO 13485:2016 Clause 7.1

      Records of risk management activities in product realization not maintained

      Records of risk management activities in product realization were not adequately maintained.

    2. ISO 13485:2016 Clause 4.2.1

      Quality management system documentation not adequate

      The quality management system documentation does not include a quality manual.

  2. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  3. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Two requests to FDA under the Freedom of Information Act named this site, most recently on Aug 13, 2026. Each one asked for a Form 483.

  • Inspection ending Mar 25, 2026NAI

    2 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Aug 24, 2026. FDAgent holds no copy that FDA posted.

    1. Aug 13, 2026FDA-FOIA-2026-7843Form 483Records released
    2. Jun 23, 2026FDA-FOIA-2026-6012Form 483· 1 of 5 sitesRecords released

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Regulations FDA cited here

The 1 section of 21 CFR in FDA’s published citations for 1 of this facility’s 3 inspections.

  • 21 CFR 830.310Information required for unique device identification1 of 1

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3009770872 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3009770872. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.