Tesla Biohealing, Inc.
111 Mccoy St, Milford, Delaware, United States
FEI 3023229437
FDA has 2 inspections on file for Tesla Biohealing, Inc. in Milford, Delaware (FEI 3023229437), from 2023 to 2026. The latest ended on and was classified OAI, official action indicated.
OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.
FDA published 9 citations from this inspection. They are listed below. Three of the four FOIA requests for this site have asked FDA for this inspection's 483.
FDA issued a warning letter on (CDRH · Division of Medical Device and Radiological Health Operations East). No court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Warning letter ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- OAI
- Inspections on file
- 2, 2023–2026
- 2 OAI · 0 VAI · 0 NAI
- Inspections with 483 observations shown
- 2 of 2
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 4 requests, latest Jul 7, 2026
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices9 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
- ISO 13485:2016 Clause 7.3.9
Design and development changes not identified
Design and development changes were not identified.
GUDID Information required for unique device identification.
The correct unique device identifier (UDI) information regarding device identifier portion of the UDI assigned to the version or model, proprietary, trade, or brand name of the device as it appears on the label, version or model or similar reference that appears on the label of the device, size of the particular version or model given the device is available in more than one size, the type of production identifiers that appears on the label of the device and the Global Medical Device Nomenclature (GMDN) term or code for the device was not provided to the FDA via the Global Unique Device Identi
- ISO 13485:2016 Clause 8.2.2
Complaint handling procedures lack required elements
Procedures for timely complaint handling do not adequately include requirements and responsibilities for receiving and recording information, evaluating information to determine if the feedback constitutes a complaint, investigating complaints, determining the need to report the information to FDA and determining the need to initiate corrections or corrective actions.
- ISO 13485:2016 Clause 5.5.1
Interrelation of personnel not documented
Top management has not documented the interrelation of all personnel who manage, perform, and verify work affecting quality.
- ISO 13485:2016 Clause 7.6
Software used for monitoring and measurement not validated
Software used for the monitoring and measurement of requirements was not validated prior to initial use.
Show 4 more citations
- ISO 13485:2016 Clause 7.1
Product realization risk management processes not documented
One or more processes for risk management in product realization have not been documented.
- ISO 13485:2016 Clause 7.3.7
Design and development validation plans not adequate
Design and development validation plans did not include methods and acceptance criteria.
- ISO 13485:2016 Clause 7.4.1
Criteria for the evaluation and selection of suppliers not established
Criteria for the evaluation and selection of suppliers were not established.
- ISO 13485:2016 Clause 8.5.2
Corrective action procedures not documented
Corrective action procedures defining the requirements for reviewing nonconformities, determining the causes of nonconformities, evaluating the need for action to ensure that nonconformities do not recur, planning and documenting action needed and implementing such action, verifying that the corrective action does not adversely affect the ability to meet applicable regulatory requirements or the safety and performance of the medical device and reviewing the effectiveness of corrective action taken have not been adequately documented.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Evaluation and Selection, Suppliers, Contractors, etc.
Potential suppliers were not evaluated and selected based on their ability to meet specified requirements.
Lack of or inadequate organizational structure
The organizational structure has not been established and maintained to ensure that devices are designed and produced in accordance with 21 CFR 820.
Design validation - user needs and intended uses
Design validation did not ensure the device conforms to defined user needs and intended uses.
Equipment suitability & capability
Certain inspection, measuring and test equipment is not suitable for its intended purposes and capable of producing valid results.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Four requests to FDA under the Freedom of Information Act named this site, most recently on Jul 7, 2026. Each one asked for a Form 483.
Inspection ending Apr 16, 2026OAI
3 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Sep 14, 2026. FDAgent holds no copy that FDA posted.
- Jul 7, 2026FDA-FOIA-2026-6453Form 483· 1 of 3 sitesRecords released
- Jul 7, 2026FDA-FOIA-2026-6451Form 483· 1 of 3 sitesMerged with another request
- Jun 18, 2026FDA-FOIA-2026-5923Form 483· 1 of 4 sitesRecords released
Inspection ending Mar 23, 2023OAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Oct 3, 2023FDA-FOIA-2023-8470Form 483Closed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 5 sections of 21 CFR in FDA’s published citations for this facility’s 2 inspections, with how many of them cited each.
- 21 CFR 820.20Management responsibility1 of 2
- 21 CFR 820.30Design controls1 of 2
- 21 CFR 820.50Purchasing controls1 of 2
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 2
- 21 CFR 830.310Information required for unique device identification1 of 2
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · Division of Medical Device and Radiological Health Operations East · DevicesLetter on fda.govFDA record
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3023229437 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3023229437. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.