Skip to content

Superior Labs Inc

4236 Wilkinson Blvd, Charlotte, North Carolina, United States

FEI 3034286079

FDA has one inspection on file for Superior Labs Inc in Charlotte, North Carolina (FEI 3034286079). It ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 9 citations from this inspection. They are listed below. One FOIA request has asked FDA for this inspection's 483.

No warning letter, court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
1, in 2026
Inspections with 483 observations shown
1 of 1
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
1 request, latest Sep 24, 2026
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices9 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. ISO 13485:2016 Clause 7.4.3

      Records of purchased product verification not maintained

      Records of the verification of purchased product were not maintained.

    2. UDI was not recorded for each medical device or batch

      Unique device identification (UDI) was not recorded for each medical device or batch of medical devices.

    3. ISO 13485:2016 Clause 7.5.1

      Production and service provision not adequate

      Production and service provision was not planned, carried out, monitored, and controlled to ensure that product conforms to specification.

    4. ISO 13485:2016 Clause 7.1

      Planning of product realization not consistent with other QMS processes

      Planning of product realization was not consistent with the requirements of other quality management system processes.

    5. Lack of Written MDR Procedures

      Written MDR procedures have not been developed and implemented.

    Show 4 more citations
    1. ISO 13485:2016 Clause 4.1.1

      QMS not documented and maintained adequately

      The quality management system was not documented in accordance with applicable regulatory requirements.

    2. ISO 13485:2016 Clause 7.3.2

      Design and development of product not planned and controlled

      Design and development of product was not planned and controlled.

    3. ISO 13485:2016 Clause 8.2.2

      Complaint handling procedures not documented

      Procedures for timely complaint handling have not been documented.

    4. ISO 13485:2016 Clause 7.6

      Equipment calibration or verification not performed

      Equipment calibration or verification was not performed according to documented procedures.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

One request to FDA under the Freedom of Information Act named this site, on Sep 24, 2026. It asked for a Form 483.

  • Inspection ending Aug 11, 2026VAI

    1 request. In the inspection history above.

    1. Sep 24, 2026FDA-FOIA-2026-9250Form 483· 1 of 5 sitesOpen

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Regulations FDA cited here

The 2 sections of 21 CFR in FDA’s published citations for this facility’s 1 inspection.

  • 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 1
  • 21 CFR 820.35Control of records1 of 1

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3034286079 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3034286079. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.