Domico Med-Device Llc
14241 N Fenton Rd, Fenton, Michigan, United States
FEI 1825560
FDA has 2 inspections on file for Domico Med-Device Llc in Fenton, Michigan (FEI 1825560), from 2016 to 2026. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
FDA published 6 citations from this inspection. They are listed below. Four of the five FOIA requests for this site have asked FDA for this inspection's records.
No warning letter, court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 2, 2016–2026
- 0 OAI · 1 VAI · 1 NAI
- Inspections with 483 observations shown
- 2 of 2
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 5 requests, latest Aug 21, 2026
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
- ISO 13485:2016 Clause 8.5.2
Corrective action procedures not documented
Corrective action procedures defining the requirements for determining the causes of nonconformities, planning and documenting action needed and implementing such action, verifying that the corrective action does not adversely affect the ability to meet applicable regulatory requirements or the safety and performance of the medical device and reviewing the effectiveness of corrective action taken have not been adequately documented.
- ISO 13485:2016 Clause 7.1
Product realization risk management processes not documented
One or more processes for risk management in product realization have not been adequately documented.
- ISO 13485:2016 Clause 8.2.2
Complaint handling procedures lack required elements
Procedures for timely complaint handling do not adequately include requirements and responsibilities for receiving and recording information and investigating complaints.
- ISO 13485:2016 Clause 7.6
Monitoring and measurement requirements not determined
Your firm did not adequately determine the monitoring and measurement to be undertaken and the related equipment needed to provide evidence of conformity of product to requirements.
Records of complaint investigations do not include required information
Records of complaint investigations do not include any unique device identifier (UDI) or universal product code (UPC) and any other device identification(s).
Show 1 more citation
- ISO 13485:2016 Clause 7.5.1
Production controls not adequate
Production controls do not include qualification of infrastructure.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely calibrated have not been adequately established.
Quality audits - auditor independence
Individuals who conduct quality audits have direct responsibility for the matters being audited.
Procedures not adequately established or maintained
Document control procedures have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Five requests to FDA under the Freedom of Information Act named this site, most recently on Aug 21, 2026.
Inspection ending May 12, 2026NAI
4 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Aug 26, 2026. FDAgent holds no copy that FDA posted.
- Aug 21, 2026FDA-FOIA-2026-8127Form 483· 1 of 4 sitesOpen
- Aug 3, 2026FDA-FOIA-2026-7389Form 483Records released
- Jul 15, 2026FDA-FOIA-2026-6727Form 483· 1 of 5 sitesRecords released
- Jul 7, 2026FDA-FOIA-2026-6490EIROpen
Inspection ending Jan 1, 2010
1 request. Not listed in FDA’s Data Dashboard yet.
- Jun 15, 2026FDA-FOIA-2026-5815EIR· 1 of 4 sitesOpen
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 5 sections of 21 CFR in FDA’s published citations for this facility’s 2 inspections, with how many of them cited each.
- 21 CFR 820.22Quality audit1 of 2
- 21 CFR 820.35Control of records1 of 2
- 21 CFR 820.40[Reserved]1 of 2
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 2
- 21 CFR 820.100Corrective and preventive action1 of 2
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1825560 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1825560. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.