FDA has 3 inspections on file for Jazz Imaging in San Jose, California (FEI 3013503781), from 2019 to 2026. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 5 citations from this inspection. They are listed below. One FOIA request has asked FDA for this inspection's 483.

No warning letter, court action or import refusal is filed under this FEI.

Six regulations were cited at more than one inspection; they are listed below.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
3, 2019–2026
0 OAI · 2 VAI · 1 NAI
Inspections with 483 observations shown
3 of 3
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
1 request, latest Oct 28, 2026
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. ISO 13485:2016 Clause 7.5.1

      Production controls not adequate

      Production controls do not include implementation of monitoring and measurement of process parameters and product characteristics.

    2. Lack of Written MDR Procedures

      Written MDR procedures have not been developed.

    3. ISO 13485:2016 Clause 6.2

      Process(es) for personnel competence, training and awareness not documented

      Process(es) for providing needed training have not been documented.

    4. ISO 13485:2016 Clause 8.2.2

      Complaint handling procedures lack required elements

      Procedures for timely complaint handling do not adequately include requirements and responsibilities for determining the need to report the information to FDA.

    5. Records of complaint investigations do not include required information

      Records of complaint investigations do not include any unique device identifier (UDI) or universal product code (UPC) and any other device identification(s).

  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices13 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Records of complaint investigation

      Records of complaint investigations do not include required information.

    2. Quality Audit/Reaudit - conducted

      Quality audits have not been performed.

    3. Statistical techniques - Lack of or inadequate procedures

      Procedures for identifying valid statistical techniques required for establishing, controlling, and verifying the acceptability of process capability and product characteristics have not been adequately established.

    4. Documentation

      Corrective and preventive action activities and/or results have not been adequately documented.

    5. Documented evaluation

      The evaluation of potential suppliers and contractors was not documented.

    Show 8 more citations
    1. Nonconforming product control

      Products that do not conform to specifications are not adequately controlled.

    2. Equipment Installation, Placement, Specified Requirements

      The equipment used in the manufacturing process does not meet specified requirements.

    3. Calibration, Inspection, etc. Procedures Lack of or Inadequ

      Procedures to ensure equipment is routinely calibrated, inspected, checked and maintained have not been established.

    4. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been established.

    5. Training - Lack of or inadequate procedures

      Procedures for training and identifying training needs have not been adequately established.

    6. Documentation

      Process validation activities have not been documented and approved.

    7. Management review - defined interval, sufficient frequency

      Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system at defined intervals and with sufficient frequency.

    8. Management representative implementing the quality system

      The management representative has not ensured that the quality system requirements are effectively established and maintained.

  3. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices10 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Documented evaluation

      The evaluation of potential suppliers, contractors and consultants was not documented.

    2. Documentation

      Corrective and preventive action activities and/or results have not been documented.

    3. Management review - defined interval, sufficient frequency

      Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system at defined intervals and with sufficient frequency.

    4. Quality Audit/Reaudit - conducted

      Quality audits and reaudits have not been performed.

    5. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    Show 5 more citations
    1. Remedial action

      When test/measurement equipment was found to not meet accuracy and precision limits, no action was taken to evaluate whether there was any adverse effect on the device's quality.

    2. Product rework documentation, DHR {see also 820.184}

      Rework and reevaluation activities have not been documented in the device history record.

    3. Product rework procedures, Lack of or inadequate procedures

      Procedures for rework of nonconforming product have not been established.

    4. Lack of Written MDR Procedures

      Written MDR procedures have not been developed and implemented.

    5. Rationale documented for no investigation

      Records for complaints where no investigation was made do not include required information.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

One request to FDA under the Freedom of Information Act named this site, on Oct 28, 2026. It asked for a Form 483.

  • Inspection ending Oct 1, 2026VAI

    1 request. In the inspection history above.

    1. Oct 28, 2026FDA-FOIA-2026-10439Form 483· 1 of 4 sitesOpen

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Regulations FDA cited here

The 13 sections of 21 CFR in FDA’s published citations for this facility’s 3 inspections, with how many of them cited each.

Show all 13

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3013503781 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3013503781. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.