Skip to content

21 CFR 801.40: Form of a unique device identifier

FDA's inspection data lists 8 Form 483 citations of 21 CFR 801.40 at 8 facilities, from February 2024 to September 17, 2026.

Counts: FDA Data Dashboard citations, last synced Nov 23, 2026, 04:25 UTC, and FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Regulation text: eCFR. FDAgent is independent of FDA.

Form 483 citations
8
At 8 inspections
Facilities cited
8
Distinct FEIs
Most recent citation
Sep 17, 2026
BSN Medical, Inc.
Warning letters
0
State a violation of it

The regulation

The current text of 21 CFR 801.40 from the Electronic Code of Federal Regulations.

§ 801.40 Form of a unique device identifier.

(a) Every unique device identifier (UDI) must meet the technical requirements of § 830.20 of this chapter. The UDI must be presented in two forms:

(1) Easily readable plain-text, and

(2) Automatic identification and data capture (AIDC) technology.

(b) The UDI must include a device identifier segment. Whenever a device label includes a lot or batch number, a serial number, a manufacturing date, an expiration date, or for a human cell, tissue, or cellular or tissue-based product (HCT/P) regulated as a device, a distinct identification code as required by § 1271.290(c) of this chapter, the UDI must include a production identifier segment that conveys such information.

(c) If the AIDC technology is not evident upon visual examination of the label or device package, the label or device package must disclose the presence of AIDC technology.

(d) A class I device that bears a Universal Product Code (UPC) on its label and device packages is deemed to meet all requirements of subpart B of this part. The UPC will serve as the unique device identifier required by § 801.20.

21 CFR Part 801: Labeling, Subpart BRead on eCFR

Citations by year

Form 483 citations of 21 CFR 801.40 by the year the inspection ended. FDA's citation data starts with inspections ending in October 2008, and the current year runs to the latest sync, so both are partial years (lighter bars).

Peak: 5 in 2026

Form 483 citations of 21 CFR 801.40 by inspection year
YearForm 483 citations
20242
20251
20265

How FDA words this citation

FDA's short descriptions and citation text on these 8 citations, most used first. These are FDA's standard summaries in its citation data, not the text of any one Form 483.

Short descriptions

  • Form of a unique device identifier8 citations

Citation text

  • “The medical device(s) label did not have a unique device identifier (UDI) present in easily readable plain-text and automatic identification and data capture (AIDC) technology.”5 citations
  • “The medical device(s) label did not have a unique device identifier (UDI) present in.”2 citations
  • “The medical device(s) label did not have a unique device identifier (UDI) present in easily readable plain-text.”1 citations

Facilities cited most recently

The latest inspections where FDA cited 21 CFR 801.40, by inspection end date, with FDA's short description. Each links to the facility's FDA record on FDAgent.

Show 2 more

Often cited at the same inspection

Of the 8 inspections where FDA cited 21 CFR 801.40, the share where it also cited each of these sections.

Ask FDAgent about 21 CFR 801.40

FDAgent reads FDA's Form 483 citations, observations and warning letters and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • Citation counts come from FDA's Data Dashboard, which lists the regulation behind each Form 483 observation for inspections since October 2008.
  • Observation excerpts come from Form 483 PDFs in FDA's electronic reading room, tied to this section only by exact wording as described above.
  • A warning letter counts for this section when one of its numbered violation sentences cites 21 CFR 801.40; FDAgent quotes that sentence exactly.
  • FDAgent rebuilds this page every week from FDA's public data and is independent of FDA.