Automated Ophthalmics, Inc.
6671 Santa Barbara Rd Ste F, Elkridge, Maryland, United States
FEI 3005906810
FDA has 6 inspections on file for Automated Ophthalmics, Inc. in Elkridge, Maryland (FEI 3005906810), from 2009 to 2026. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA posted the Form 483 from that inspection. FDA published 5 citations from this inspection. They are listed below. Three FOIA requests have asked FDA for this inspection's 483.
The most recent OAI inspection ended on Mar 5, 2013. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
Two regulations were cited at more than one inspection: 21 CFR 820.90(a) and 21 CFR 820.30(a).
Sources:FDA firm profile ↗Form 483 PDF ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 6, 2009–2026
- 1 OAI · 4 VAI · 1 NAI
- Inspections with 483 observations shown
- 5 of 6
- Form 483s posted by FDA
- ↗
- Requested under FOIA
- 3 requests, latest May 19, 2026
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
6 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: DevicesPosted 483 · 5 citations
Read this Form 483 on its own page →Form 483 PDF on fda.gov
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
- ISO 13485:2016 Clause 7.5.11
Product not protected during processing, storage, handling, or distribution
Product was not protected from alteration, contamination, or damage during processing, storage, handling, and distribution by.
Records of complaint investigations do not include required information
Records of complaint investigations do not include.
- ISO 13485:2016 Clause 8.2.2
Complaint handling records not maintained
Complaint handling records were not adequately maintained.
- ISO 13485:2016 Clause 8.3.1
Nonconforming product not adequately evaluated
Nonconforming product was not adequately evaluated to determine the need for an investigation and notification of any external party responsible for the nonconformity.
Form of a unique device identifier
The medical device(s) label did not have a unique device identifier (UDI) present in.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Environmental control Lack of or inadequate procedures
Procedures to control environmental conditions have not been adequately established.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been adequately established.
Documentation
Corrective and preventive action activities and/or results have not been documented.
Design control - no procedures
Procedures for design control have not been established.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices12 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Nonconforming product control
Products that do not conform to specifications are not adequately controlled.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Management review - Lack of or inadequate procedures
Procedures for management review have not been adequately established.
Quality Audits - defined intervals
Quality audits were not performed to determine whether the quality system activities and results comply with quality system procedures.
Lack of Written MDR Procedures
Written MDR procedures have not been developed.
Show 7 more citations
Design control - no procedures
Procedures for design control have not been established.
Design history file
The design history file was not established.
Documented evaluation
The evaluation of potential suppliers was not documented.
DHR - not or inadequately maintained
A device history record has not been maintained.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.
Documentation
Acceptance activities were not adequately documented.
Sampling plans
Sampling plans are not written and based on valid statistical rationale.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design verification - documentation
The design verification results, including methods, the date, and the individual performing the verification, were not documented and filed in the design history file.
Lack of or inadequate receiving acceptance procedures
Procedures for acceptance or rejection of incoming product were not defined, documented, complete, and implemented.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Three requests to FDA under the Freedom of Information Act named this site, most recently on May 19, 2026. Each one asked for a Form 483.
Inspection ending Feb 13, 2026VAI
3 requests. In the inspection history above. FDA has posted its Form 483.
- May 19, 2026FDA-FOIA-2026-5138Form 483· 1 of 4 sitesRecords released
- Apr 13, 2026FDA-FOIA-2026-3674Form 483· 1 of 5 sitesRecords released
- Mar 5, 2026FDA-FOIA-2026-2264Form 483Records released
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 14 sections of 21 CFR in FDA’s published citations for 5 of this facility’s 6 inspections, with how many of them cited each.
- 21 CFR 820.30Design controls3 of 5
- 21 CFR 820.80Receiving, in-process, and finished device acceptance3 of 5
- 21 CFR 820.90Nonconforming product2 of 5
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 5
- 21 CFR 820.20Management responsibility1 of 5
- 21 CFR 820.22Quality audit1 of 5
- 21 CFR 820.35Control of records1 of 5
- 21 CFR 820.50Purchasing controls1 of 5
Show all 14
- 21 CFR 820.70Production and process controls1 of 5
- 21 CFR 820.75Process validation1 of 5
- 21 CFR 820.100Corrective and preventive action1 of 5
- 21 CFR 820.184Device history record1 of 5
- 21 CFR 820.250Statistical techniques1 of 5
- 21 CFR 801.40Form of a unique device identifier1 of 5
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3005906810 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3005906810. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.