FDA has 5 inspections on file for Collaborative Care Diagnostics, LLC. in Eagan, Minnesota (FEI 3002619777), from 2009 to 2026. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
FDA posted the Form 483 from that inspection. FDA published 4 citations from this inspection. They are listed below. Four FOIA requests have asked FDA for this inspection's 483.
FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Form 483 PDF ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 5, 2009–2026
- 0 OAI · 1 VAI · 4 NAI
- Inspections with 483 observations shown
- 3 of 5
- Requested under FOIA
- 4 requests, latest May 19, 2026
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
5 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesPosted 483 · 4 citations
Read this Form 483 on its own page →Form 483 PDF on fda.gov
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
- ISO 13485:2016 Clause 7.1
Product realization risk management processes not documented
One or more processes for risk management in product realization have not been adequately documented.
- ISO 13485:2016 Clause 7.3.9
Review of design and development do not include evaluation of change effects
Review of design and development changes did not include evaluation of the effect of the changes on constituent parts and product in process or already delivered, inputs or outputs of risk management and product realization processes.
Devices subject to device identification GUDID data submission requirements.
The information required by the Global Unique Device Identification Database for each medical device version or model required to bear a unique device identifier (UDI) was not provided.
Form of a unique device identifier
The medical device(s) label did not have a unique device identifier (UDI) present in easily readable plain-text and automatic identification and data capture (AIDC) technology.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design input - Lack of or inadequate procedures
Procedures for design input have not been adequately established.
Design validation- Lack of or inadequate procedures
Procedures for design validation have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Compliance: Devices1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Four requests to FDA under the Freedom of Information Act named this site, most recently on May 19, 2026. Each one asked for a Form 483.
Inspection ending Feb 9, 2026NAI
4 requests. In the inspection history above. FDA has posted its Form 483.
- May 19, 2026FDA-FOIA-2026-5145Form 483· 1 of 3 sitesRecords released
- May 5, 2026FDA-FOIA-2026-4451Form 483Records released
- Apr 14, 2026FDA-FOIA-2026-3726Form 483· 1 of 3 sitesRecords released
- Apr 13, 2026FDA-FOIA-2026-3667Form 483· 1 of 5 sitesWithdrawn
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 4 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 5 inspections, with how many of them cited each.
- 21 CFR 820.30Design controls1 of 3
- 21 CFR 820.100Corrective and preventive action1 of 3
- 21 CFR 830.300Devices subject to device identification data submission requirements1 of 3
- 21 CFR 801.40Form of a unique device identifier1 of 3
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3002619777 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3002619777. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.