Bionix, Llc
1670 Indian Wood Cir, Maumee, Ohio, United States
FEI 1526854
FDA has 6 inspections on file for Bionix, Llc in Maumee, Ohio (FEI 1526854), from 2010 to 2026. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 8 citations from this inspection. They are listed below. Two FOIA requests have asked FDA for this inspection's 483.
No warning letter, court action or import refusal is filed under this FEI.
Two regulations were cited at more than one inspection: 21 CFR 820.100(a) and 21 CFR 820.198(a).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 6, 2010–2026
- 0 OAI · 4 VAI · 2 NAI
- Inspections with 483 observations shown
- 4 of 6
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 2 requests, latest Oct 16, 2026
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
6 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices8 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
- ISO 13485:2016 Clause 4.1.2
Risk-based approach not applied to QMS processes
A risk-based approach to the control of the appropriate processes needed for the quality management system has not been adequately applied.
Records of MDR Investigation
Investigation records of MDR reportable complaints do not include required information.
- ISO 13485:2016 Clause 8.4
Procedures for analysis of data not documented
Procedures were not adequately documented to determine, collect, and analyze appropriate data to demonstrate the suitability, adequacy, and effectiveness of the quality management system.
- ISO 13485:2016 Clause 7.4.3
Purchased product verification activities not established or implemented
The inspection or other activities necessary for ensuring that purchased product meets specified purchasing requirements were not adequately established and implemented.
- ISO 13485:2016 Clause 7.3.6
Design and development verification not performed
Design and development verification was not adequately performed in accordance with planned and documented arrangements to ensure that design and development outputs met design and development input requirements.
Show 3 more citations
- ISO 13485:2016 Clause 7.1
Records of risk management activities in product realization not maintained
Records of risk management activities in product realization were not adequately maintained.
- ISO 13485:2016 Clause 7.4.1
Supplier evaluation and selection criteria do not include required elements
Criteria for the evaluation and selection of suppliers were not based on the supplier's ability to provide product that meets requirements, based on the performance of the supplier, based on the effect of the purchased product on the quality of the device and proportionate to the risk associated with the medical device.
Records of complaint investigation - UDI
Records of complaint investigations do not include required information, including any unique device identifier (UDI) or universal product code (UPC).
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Training records
Personnel training is not documented.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.
DMR - not or inadequately maintained
A device master record has not been maintained.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been established.
Acceptable supplier records, inadequate records
Records of acceptable consultants have not been established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been established.
Design control - no procedures
Procedures for design control have not been established.
Lack of System for Event Evaluations
The written MDR Procedure does not include an internal system which provides for the timely and effective communication and evaluation of events that may be subject to medical device reporting requirements.
Show 1 more citation
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been established.
Approval, inadequate purchasing data
Purchasing data that clearly describe or reference specified requirements for purchased or otherwise received product and services have not been established.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been established.
Investigation of device failures
Complaints involving the possible failure of to meet any of its specifications were not investigated where necessary.
Lack of System for Determining MDR Events
The written MDR procedure does not include an internal system which provides for a standardized review process/procedure for determining when an event meets the criteria for reporting.
Show 2 more citations
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.
Evaluation and Selection, Suppliers, Contractors, etc.
Potential suppliers and contractors were not evaluated and selected based on their ability to meet specified requirements.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: DevicesCitations not published
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Two requests to FDA under the Freedom of Information Act named this site, most recently on Oct 16, 2026. Each one asked for a Form 483.
Inspection ending Jul 14, 2026VAI
2 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Sep 22, 2026. FDAgent holds no copy that FDA posted.
- Oct 16, 2026FDA-FOIA-2026-10083Form 483· 1 of 5 sitesOpen
- Sep 1, 2026FDA-FOIA-2026-8487Form 483· 1 of 5 sitesRecords released
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 8 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 6 inspections, with how many of them cited each.
- 21 CFR 820.198Complaint files4 of 4
- 21 CFR 820.50Purchasing controls3 of 4
- 21 CFR 820.100Corrective and preventive action3 of 4
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?2 of 4
- 21 CFR 820.30Design controls2 of 4
- 21 CFR 820.25Personnel1 of 4
- 21 CFR 820.75Process validation1 of 4
- 21 CFR 820.181Device master record1 of 4
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1526854 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1526854. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.