Skip to content

Argon Medical Productions & Vertrieb Gesellschaft MBH & CO

Franz-Kirsten-Str. 1, Bingen Am Rhein, Germany

FEI 3011575197

FDA has one inspection on file for Argon Medical Productions & Vertrieb Gesellschaft MBH & CO in Bingen Am Rhein, Germany (FEI 3011575197). It ended on and was classified OAI, official action indicated. FDA issued a warning letter on (CDRH).

OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.

FDA posted the Form 483 from that inspection. FDAgent read its 4 observations from the PDF; they are listed below. One FOIA request has asked FDA for this inspection's 483.

No court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Form 483 PDF ↗Warning letter ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
OAI
Inspections on file
1, in 2024
Inspections with 483 observations shown
1 of 1
Form 483s posted by FDA
↗
Requested under FOIA
1 request, latest Feb 18, 2026
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: DevicesPosted 483 · 4 observations

    Read this Form 483 on its own page →Form 483 PDF on fda.gov

    Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent’s.

    1. Observation 1

      MDR procedures not implemented

      Written MDR procedures have not been implemented. Specifically, the Argon Post Market Surveillance procedure A002 which includes procedures for adverse event notification does not describe criteria for Medical Device Reporting and does not provide details for how MDR reporting will be made.

    2. Observation 2

      MDR reports not submitted within 30 days

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have contributed to a serious injury. Specifically, the following complaints described that the dental implant failed to osseointegrate: Complaint number, 23.058 received on April 23, 2023, implant 40065K3Pro.R/Short s/n [illegible] implant 45065K3Pro.R/Short s/n [illegible] Complaint number, 23.060 received on April 17, 2023, implant 40011K3Pro.R s/n [illegible] Complaint number, 23.062 received on April 19, 2023, implant 40075K3Pro.R s/n [illegible] Failure to osseointegrate results in additional dental procedure to remove the failed implant and install another implant.

    Show observations 3 to 4
    1. Observation 321 CFR 80121 CFR 830

      Labels lack required unique device identifier

      Labels of K3Pro Kronus Dental implants such as article 40b9K3PRO.R do not bear a unique device identifier that meet the requirements of 801 Subpart B and 21 CFR 830. Also, labeling for K3PROXP Line article BSPXP 35020/3 do not include a unique device identifier and the code that retrieves from the labeled QR code cannot be found in the GUDID database.

    2. Observation 4

      Failure to investigate osseointegration complaints

      Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary. Specifically, complaints of failure of dental implants to osseointegrate are not investigated, for

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

One request to FDA under the Freedom of Information Act named this site, on Feb 18, 2026. It asked for a Form 483.

  • Inspection ending Jul 31, 2024OAI

    1 request. In the inspection history above. FDA has posted its Form 483.

    1. Feb 18, 2026FDA-FOIA-2026-1713Form 483· 1 of 2 sitesRecords released

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Warning letters and compliance actions

FDA Data Dashboard compliance actions

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3011575197 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3011575197. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.