Skip to content

Elite Orthopedics, LLC Form 483, July 2025

FDA issued this Form 483 to Elite Orthopedics, LLC in Columbia, Missouri (FEI 3011803392) at the end of a biologics inspection that ended July 15, 2025. It lists 5 observations. FDA classified the inspection Voluntary Action Indicated (VAI).

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: FDA's Electronic Reading Room, posted Oct 30, 2026. Form 483 PDF on fda.gov. FDAgent is independent of FDA.

Inspection ended
Jul 15, 2025
Biologics · Human Cellular, Tissue, and Gene Therapies
FDA's outcome
VAI
Voluntary Action Indicated
Posted by FDA
Oct 30, 2026
472 days after the inspection
Facility
FEI 3011803392
Columbia, Missouri

The 5 observations on this Form 483

Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent's.

  1. 1

    Outdated CGTP procedures in use

    Procedures for core CTGP requirements were not available to personnel in the area where operations are performed, or in a nearby area when such availability is impractical. Specifically, On 07/15/2025, your firm was observed using an outdated procedure, titled, Human Tissue Product Distribution Guidance Manual - Version 1, Effective Date 2.1.16 to ensure the prevention, transmission, or spread of communicable diseases and disease agents during storage and distribution. Below list versions of the procedure, Human Tissue Product Distribution Guidance Manual, with effective date created, however not made available. • Version 2, Effective Date 01/18/2017 • Version 3, Effective Date 04/01/2018 • Version 4, Effective Date 11/06/2018 • Version 6, Effective Date 02/28/2025 • Version 7, Effective Date 06/23/2025

  2. 2

    Equipment maintenance documentation missing

    Documentation of equipment maintenance, cleaning, sanitization, and calibration was not maintained. Specifically, Since 06/09/2023, your firm did not document pre-installation information, preventative maintenance, and control programming for the (b) (4) use to store (b) (4). As instructed in the (b) (4) Instructions User Manual for the following elements, • Unit requirements • As needed preventative maintenance, such as battery alarm check and cleaning • (b) (4) calibration to meet reference temperature standard

  3. 3

    HCT/P storage temperatures not recorded

    Storage temperatures of HCT/Ps were not recorded and maintained. Specifically, From (b) (4), your firm distributed (b) (4) tissue products and failed to record (b) (4) temperature of storage areas and perform a (b) (4) review of the (b) (4) Temperature Log. Also, did not maintained (b) (4) calibrations records for the (b) (4) thermometer.

  4. 4

    Records not concurrent with operations

    Records were not maintained concurrently with the performance of each step. Specifically, From 08/28/2018, your firm failed to document on the Tissue Tracking Log, as instructed in the procedure, titled, Human Tissue Product Distribution Guidance, Manual, Version 1.0, Effective 2.1.16. the following acceptance criteria of received tissue. • Label verification • Package Integrity, such as no tears, rips, or holes in packaging • Temperature indicator, if compromised or not

  5. 5

    Records management system not followed

    A records management system relating to core CGTP requirements was not followed. Specifically, Your firm failed to follow procedure, titled, Human Tissue Product Distribution Guidance Manual - Version 7, Effective Date 6.23.25 to document and maintain the following records relating to core Current Good Tissue Practices. • Packing List • Tissue Tracking Form • Incident Report Form • Corrective Action Request Form • Weekly Freezer Maintenance Form • Tissue Tracking Log

Ask FDAgent about this Form 483

FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
  • The observations are the text FDAgent read from this PDF by OCR, shown only when every printed observation passed FDAgent’s checks against the scanned text.
  • FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
  • FDAgent is independent of FDA.