October 2026 · 2026-10-01 to 2026-10-31

FDA enforcement, October 2026

In October 2026 (2026-10-01 to 2026-10-31), FDA posted 39 warning letters from its drug, biologics and device offices, and listed 426 Form 483s in its Electronic Reading Room for the first time. FDA's Enforcement Report added 412 drug and device recalls, 21 of them Class I. FDAgent recorded 7 new Official Action Indicated (OAI) classifications in FDA's inspection data. FDA published 21 firm entries on its import alert lists, 14 of them on a Red List.

Complete month

Counts are FDA's records dated in October as of each source's last sync, which each section states; a late FDA update can still change them.

Warning letters posted
39
On 4 posting dates · 27 in September
Form 483s listed
426
From 413 facilities · 13 in September
Inspections newly classified
246
7 OAI · 57 VAI · 182 NAI
New OAI classifications
7
Newly listed by FDA · 2 in September
Recalled products
412
131 recall events · 21 Class I
Import alert entries
21
14 Red List · 7 Green List

Ask FDAgent about this month

FDAgent reads the warning letters FDA posted this month and answers in plain English, with links to the letters.

Edit it or ask as written. 3 questions free, no card needed.

Compared with September 2026 and October 2025

Each measure is counted the same way in every month. A year-over-year figure is shown only where FDAgent holds FDA's own dates for October 2025.

  • Warning letters posted39

    Sep 2026: 27 · +12 (+44%)Oct 2025: 19 · +20 (+105%)

  • Form 483s listed in the reading room426

    Sep 2026: 13 · +413 (+3,177%)Oct 2025: 7 · +419

  • Inspections newly classified246

    Sep 2026: 244 · +2 (+1%)Oct 2025: —¹

  • New OAI classifications7

    Sep 2026: 2 · +5Oct 2025: —¹

  • Recalled products (drug and device)412

    Sep 2026: 361 · +51 (+14%)Oct 2025: 478 · −66 (−14%)

  • Recall events131

    Sep 2026: 139 · −8 (−6%)Oct 2025: 135 · −4 (−3%)

  • Class I recall events7

    Sep 2026: 20 · −13 (−65%)Oct 2025: 21 · −14 (−67%)

  • Import alert firm entries21

    Sep 2026: 28 · −7 (−25%)Oct 2025: —²

¹ FDAgent has recorded when an inspection first appears in FDA's data only since late January 2026, so months before February 2026 have no count. ² FDA deletes an import alert entry when it removes a firm, and FDAgent keeps removed entries only from November 2026, so earlier months count only the entries FDA still lists, and the year before is not compared. October 2025 fell during the Oct 1 to Nov 12, 2025 federal shutdown, so its counts may be low. Sources: FDA warning letters page, last synced Nov 22, 2026, 04:45 UTC; Electronic Reading Room, last synced Nov 22, 2026, 15:42 UTC; FDA Data Dashboard inspections, last synced Nov 22, 2026, 04:55 UTC; openFDA Enforcement Report, FDA last updated Nov 11, 2026; FDA import alert pages, last synced Nov 20, 2026, 03:55 UTC.

Inspections FDA newly classified

246 inspections FDAgent first found in FDA's inspection classification data in October, by FDA's classification. FDA gives no date for a classification, so each counts in the month FDAgent first recorded it. When FDA later removes an inspection from its Data Dashboard, it stops counting, so a finished month's count can fall. They ended a median of 39 days before FDAgent first recorded them, and FDA's Data Dashboard lists Form 483 citations for 50 of them. Inspection classifications last synced Nov 22, 2026, 04:55 UTC.

Product typeOAIVAINAITotal
Drugs5355494
Devices2183858
Biologics049094
All757182246

The 7 OAI inspections are listed under New OAI classifications below.

Regulations cited most

The 21 CFR sections behind the Form 483 citations FDA's Data Dashboard lists for the inspections it newly classified in October (75 citations on 33 inspections), and the sections October's warning letters cite in their numbered violations (66 violations in 19 of the 39 letters). Each section links to its page. Inspection citations last synced Nov 22, 2026, 04:55 UTC.

On inspectionsCitations
21 CFR 211.22Responsibilities of quality control unit7
21 CFR 211.67Equipment cleaning and maintenance7
21 CFR 211.192Production record review7
21 CFR 211.100Written procedures; deviations5
21 CFR 211.84Testing and approval or rejection of components, drug product containers, and closures4
21 CFR 211.160General requirements3
21 CFR 212.20What activities must I perform to ensure drug quality?3
21 CFR 211.110Sampling and testing of in-process materials and drug products2
In warning lettersLetters
21 CFR 211.22Responsibilities of quality control unit10
21 CFR 211.84Testing and approval or rejection of components, drug product containers, and closures6
21 CFR 211.192Production record review6
21 CFR 211.113Control of microbiological contamination4
21 CFR 211.42Design and construction features3
21 CFR 211.160General requirements3
21 CFR 211.100Written procedures; deviations2
21 CFR 211.110Sampling and testing of in-process materials and drug products2

Sections with the same count are ordered by section number. A section without its own page is shown without a link. 59 citations name no 21 CFR section and are not counted; device citations can name ISO 13485:2016 clauses instead. FDA can list an inspection's citations weeks after its classification, so these counts can grow.

Form 483s listed in the reading room

426 Form 483s FDA listed for the first time in its Electronic Reading Room in October, dated by the publish date FDA gives each, from 413 facilities. FDA sometimes lists a 483 days after that date, so a finished month can still gain 483s. FDA also re-posted 12 Form 483s it had listed before. Re-posts don't count as new 483s. FDA posts only some 483s, often long after the inspection: it listed these a median of 538 days after their record date. Each week's page lists every 483 with its PDF. Reading room last synced Nov 22, 2026, 15:42 UTC.

By year of the 483's record date
  • 2026117
  • 2025175
  • 202472
  • 202327
  • 2022 and earlier35

The most FDA dated to one publish date was 153, on 2026-10-26.

Warning letters

39 letters FDA posted on its warning letters page in October, from its drug, biologics and device offices. 26 were issued in October and 13 earlier; FDA posted them a median of 12 days after issue. Warning letters last synced Nov 22, 2026, 04:45 UTC.

CDER · Drugs
29
CDRH · Devices
6
CBER · Biologics
3
The Office of Prescription Drug Promotion (OPDP)
1
Show 19 more

An FEI links to its facility page only when the letter names exactly one FEI and FDAgent holds FDA records for it.

Recalls by class

Drug and device recalls in FDA's Enforcement Report dated in October, from openFDA, which FDA last updated Nov 11, 2026. Each openFDA record is one recalled product; a recall event can cover many products, so events are counted too.

ClassDrugDeviceEvents
Class I3187
Class II67308117
Class III4116
Not yet classified101
All75337131

The largest Class I recall event, from Ventec Life Systems, Inc., covers 12 of the 21 Class I products. The Class I products are listed below. Class II and III recalls are listed on each week's page.

Class I21 recallsShow all
  • D-0084-2027

    Greenwich Rx

    Glutathione Injection (Glutathione-L, Ascorbic Acid, Sodium Hydroxide, Bacteriostatic Water), 200 mg/mL, 10 mL multi-dose vial, Rx only, Subcutaneous Only, Greenwich Rx.

    Reason: Microbial Contamination of Sterile Products

    Drug · initiated 2026-10-01 · Enforcement Report 2026-10-28 · Ongoing

    openFDA record
  • D-0082-2027

    Par Health USA, LLC

    Argatroban Injection, USP, 250 mg/2.5mL (100 mg/mL), 2.5 mL x 1 Single Dose Vial, Rx only, Manufactured for: Endo USA, Malvern,PA 19355/Argatroban Injection, 250 mg/2.5mL (100 mg/mL), 2.5 mL x 1 Single Dose Vial, Rx only, Distributed by: Par Pharmaceutical, Chestnut Ridge, NY 10…

    Reason: Particulate matter: particulate matter identified as cellulose, polystyrene, polyethylene teraphalate, stainless steel, or stopper material in the product solution.

    Drug · initiated 2026-10-13 · Enforcement Report 2026-10-28 · Ongoing

    openFDA record
  • D-0083-2027

    Par Health USA, LLC

    Argatroban Injection, USP, 250 mg/2.5mL (100 mg/mL), 2.5 mL x 1 Single Dose Vial, Rx only, Manufactured for: Endo USA, Malvern,PA 19355/ Argatroban Injection, 250 mg/2.5mL (100 mg/mL), 2.5 mL x 1 Single Dose Vial, Rx only, Distributed by: Par Pharmaceutical, Cranbury, NJ 08512,…

    Reason: Particulate matter: particulate matter identified as cellulose, polystyrene, polyethylene teraphalate, stainless steel, or stopper material in the product solution.

    Drug · initiated 2026-10-13 · Enforcement Report 2026-10-28 · Ongoing

    openFDA record
  • Z-0152-2027

    Philips Ultrasound, LLC

    Philips Lumify Diagnostic Ultrasound System

    Reason: Ultrasound system may: 1) Not recognize previous transducer registration, prevent scanning, require re-registration with Wi-Fi/cellular access, which may be unavailable outside of the hospital (ambulances/aircraft/field); 2) At connection between transducer USB cable to tablet/p…

    Device · initiated 2026-09-24 · Enforcement Report 2026-10-28 · Ongoing

    openFDA record
  • Z-0122-2027

    Ventec Life Systems, Inc.

    VOCSN+Pro, Model Number:PRT-00490-001, Package PN:PRT-00853-000. used for ventilatory support.

    Reason: Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.

    Device · initiated 2026-09-17 · Enforcement Report 2026-10-28 · Ongoing

    openFDA record
  • Z-0123-2027

    Ventec Life Systems, Inc.

    VOCSN+Pro (Japan), Model Number: PRT-00490-100, Package PN:PRT-00853-100. used for ventilatory support.

    Reason: Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.

    Device · initiated 2026-09-17 · Enforcement Report 2026-10-28 · Ongoing

    openFDA record
  • Z-0124-2027

    Ventec Life Systems, Inc.

    VC+Pro Model Number: PRT-00900-001, Package PN:PRT-00973-000. used for ventilatory support.

    Reason: Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.

    Device · initiated 2026-09-17 · Enforcement Report 2026-10-28 · Ongoing

    openFDA record
  • Z-0125-2027

    Ventec Life Systems, Inc.

    VC+Pro (Japan) Model Number: PRT-00900-100, Package PN:PRT-00973-100. used for ventilatory support.

    Reason: Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.

    Device · initiated 2026-09-17 · Enforcement Report 2026-10-28 · Ongoing

    openFDA record
  • Z-0126-2027

    Ventec Life Systems, Inc.

    VOCSN Model Number: PRT-01100-000, Package PN:PRT-01126-000. used for ventilatory support.

    Reason: Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.

    Device · initiated 2026-09-17 · Enforcement Report 2026-10-28 · Ongoing

    openFDA record
  • Z-0127-2027

    Ventec Life Systems, Inc.

    VOCSN (Japan) Model Number: PRT-01100-100, Package PN:PRT-01126-100. used for ventilatory support.

    Reason: Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.

    Device · initiated 2026-09-17 · Enforcement Report 2026-10-28 · Ongoing

    openFDA record
  • Z-0128-2027

    Ventec Life Systems, Inc.

    VCSN+Pro Model Number: PRT-01184-000, Package PN:PRT-01172-000. used for ventilatory support.

    Reason: Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.

    Device · initiated 2026-09-17 · Enforcement Report 2026-10-28 · Ongoing

    openFDA record
  • Z-0129-2027

    Ventec Life Systems, Inc.

    V+Pro Model Number: PRT-01185-000, Package PN:PRT-01198-000; Model Number: PRT-01185-001, Package PN: N/A; Model Number: PRT-01185-002, Package PN:N/A. used for ventilatory support.

    Reason: Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.

    Device · initiated 2026-09-17 · Enforcement Report 2026-10-28 · Ongoing

    openFDA record
  • Z-0130-2027

    Ventec Life Systems, Inc.

    VC Model Number: PRT-01098-000, Package PN:PRT-01125-000. used for ventilatory support.

    Reason: Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.

    Device · initiated 2026-09-17 · Enforcement Report 2026-10-28 · Ongoing

    openFDA record
  • Z-0131-2027

    Ventec Life Systems, Inc.

    VC(Japan) Model Number: PRT-01098-100, Package PN:PRT-01125-100. used for ventilatory support.

    Reason: Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.

    Device · initiated 2026-09-17 · Enforcement Report 2026-10-28 · Ongoing

    openFDA record
  • Z-0132-2027

    Ventec Life Systems, Inc.

    V+ Pro Emergency, Model Number: PRT-01201-000, Package PN:PRT-01173-000,PRT-01222-000. used for ventilatory support.

    Reason: Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.

    Device · initiated 2026-09-17 · Enforcement Report 2026-10-28 · Ongoing

    openFDA record
  • Z-0133-2027

    Ventec Life Systems, Inc.

    V*Home, Model Number: PRT-01444-000, Package PN:PRT-01445-000. used for ventilatory support.

    Reason: Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.

    Device · initiated 2026-09-17 · Enforcement Report 2026-10-28 · Ongoing

    openFDA record
  • Z-0001-2027

    Instrumentation Laboratory

    HemoslL AcuStar HIT-lgG(PF-4). Part Number: 0009800127. in vitro diagnostic

    Reason: Potential for invalid calibration curve or negative mean bias of up to 47% with patient samples.

    Device · initiated 2026-09-03 · Enforcement Report 2026-10-21 · Ongoing

    openFDA record
  • Z-0002-2027

    Instrumentation Laboratory

    HemoslL AcuStar HIT-lgG(PF-4). Part Number: 0009802028. in vitro diagnostic (OUS only)

    Reason: Potential for invalid calibration curve or negative mean bias of up to 47% with patient samples.

    Device · initiated 2026-09-03 · Enforcement Report 2026-10-21 · Ongoing

    openFDA record
  • Z-3252-2026

    Draeger, Inc.

    VentStar Resus Neo. Neonatal. 25 Units per Box. Part Number: MP00310 Disposable breathing circuit for use with a resuscitation module.

    Reason: Potential cracks in the hose, which could lead to leakage or the hose falling off.

    Device · initiated 2026-09-03 · Enforcement Report 2026-10-07 · Ongoing

    openFDA record
  • Z-3227-2026

    Medline Industries, LP

    Medline Basic Homecare Beds and Medlite Homecare Beds and Accessories, Hand Pendants: MDR510094 PENDANT F/ MDR107002E MDR520094 PENDANT F/ MDR107003E MDR530094 PENDANT, W/ 9V BATTERY BACKUP, F/ 2E BED MDR540094 PENDANT, W/ 9V BATTERY BACKUP, F/ 3E BED

    Reason: Potential fire hazard involving the hand control pendant and associated wiring used with affected beds. Pendant or pendant wiring may overheat, potentially leading to smoking, sparking, melting, burning, or fire. Additional products added to 11/26/2025 recall.

    Device · initiated 2026-08-14 · Enforcement Report 2026-10-07 · Ongoing

    openFDA record
  • Z-3228-2026

    Medline Industries, LP

    Medline Basic Homecare Beds and Medlite Homecare Beds and Accessories MDR107000P BED, MANUAL, MEDLITE

    Reason: Potential fire hazard involving the hand control pendant and associated wiring used with affected beds. Pendant or pendant wiring may overheat, potentially leading to smoking, sparking, melting, burning, or fire. Additional products added to 11/26/2025 recall.

    Device · initiated 2026-08-14 · Enforcement Report 2026-10-07 · Ongoing

    openFDA record

Import alert entries

21 firm entries FDA published on its import alert pages this month, dated by each entry's date published: 14 on a Red List, 7 on a Green List. FDA lists firms by name and address; FDAgent links an entry to a facility only when FDA printed the firm's FEI. Import alerts last synced Nov 20, 2026, 03:55 UTC.

66-40Red List2 entriesShow allDetention Without Physical Examination of Drugs From Firms Which Have Not Met Drug GMPsAlert on fda.gov
  • Published 2026-10-27

    COHESION BIOTEC PRIVATE LIMITED

    INDIA

  • Published 2026-10-01

    Nanjing Chengong Pharmaceutical Co., Ltd.

    CHINA

66-41Red List2 entriesShow allDetention Without Physical Examination of Unapproved New Drugs Promoted In The U.S.Alert on fda.gov
  • Published 2026-10-09

    Shantou North Moon Biotechnology Co., Ltd.

    CHINA

  • Published 2026-10-01

    NINGBO CHENGMEI MEDICAL PRODUCTSCO., LTD

    CHINA

76-01Green List7 entriesShow allDetention Without Physical Examination Of Medical Instruments from PakistanAlert on fda.gov
  • Published 2026-10-22

    Surgitrace Instruments

    PAKISTAN

  • Published 2026-10-22

    Washington Surgical Intl. (Private) Limited

    PAKISTAN

  • Published 2026-10-20

    Khadim Surgical Industry

    PAKISTAN

  • Published 2026-10-15

    Benelux Surgical

    PAKISTAN

  • Published 2026-10-15

    Scal Surgical

    PAKISTAN

  • Published 2026-10-09

    Revlis Enterprise

    PAKISTAN

  • Published 2026-10-07

    TAHIRON CORPORATION

    PAKISTAN

89-04Red List3 entriesShow allDetention Without Physical Examination of Devices from Firms that Have not met Device Quality System RequirementsAlert on fda.gov
  • Published 2026-10-19

    AOMORI OLYMPUS CO., LTD. Kuroishi Site

    JAPAN

  • Published 2026-10-19

    OLYMPUS MEDICAL SYSTEMS CORP. Hinode Plant

    JAPAN

  • Published 2026-10-19

    Shirakawa Olympus Co., Ltd.

    JAPAN

89-08Red List1 entryShow allDetention Without Physical Examination of Devices without Approved PMA's or IDE's and Other Devices Not Substantially Equivalent or Without a 510(k)Alert on fda.gov
  • Published 2026-10-07

    CHANGSHA RENJI MEDICAL EQUIPMENT CO., LTD.

    CHINA

95-05Red List5 entriesShow allDetention Without Physical Examination of Electronic Products that Fail to Comply with Performance Standards or to Have Required CertificationAlert on fda.gov
  • Published 2026-10-29

    JINAN CONSURE ELECTRONIC TECHNOLOGY CO LTD

    CHINA

  • Published 2026-10-23

    ZHONGSHAN POURINGLIGHT TECHNOLOGY CO., LTD

    CHINA

  • Published 2026-10-22

    Wuhan Juyang Technology Co., Ltd.

    CHINA

  • Published 2026-10-07

    DEMARK (WUHAN) TECHNOLOGY CO., LTD

    CHINA

  • Published 2026-10-02

    QINGDAO MEKTEX ELECTRONIC TECHNOLOGY CO., LTD

    CHINA

99-34Red List1 entryShow allDETENTION WITHOUT PHYSICAL EXAMINATION OF DRUGS OR MEDICAL DEVICES FROM FIRMS WITHOUT A VALID DRUG OR MEDICAL DEVICE REGISTRATION AND/OR LISTINGAlert on fda.gov
  • Published 2026-10-13

    SHRIYAN HEALTHCARE

    INDIA

New OAI classifications

7 inspections classified Official Action Indicated that FDAgent first recorded from FDA's inspection classification data this month, after its reads of that data in the 8 days before had not found them. Older inspections FDAgent adds to its archive are left out. Dates are inspection end dates. Inspection classifications last synced Nov 22, 2026, 04:55 UTC.

  • Inspection ended 2026-08-19

    Pharmalogic New York City, LLCOAI

    Bronx, New York, United States · Drugs · Drug Quality Assurance · recorded 2026-10-26

    FEI 3029514624
  • Inspection ended 2026-07-23

    BIOPSYBELL S.R.L.OAI

    Mirandola, Italy · Devices · Compliance: Devices · recorded 2026-10-19

    FEI 3003272754
  • Inspection ended 2026-07-21

    Endovascular Engineering, IncOAI

    Menlo Park, California, United States · Devices · Bioresearch Monitoring · recorded 2026-10-12

    FEI 3026567963
  • Inspection ended 2026-07-14

    Aspen SA Sterile Operations (Pty) Ltd.OAI

    Port Elizabeth, South Africa · Drugs · Drug Quality Assurance · recorded 2026-10-12

    FEI 3007490529
  • Inspection ended 2026-07-22

    Medline Industries, LPOAI

    Mundelein, Illinois, United States · Drugs · Drug Quality Assurance · recorded 2026-10-12

    FEI 1000138386
  • Inspection ended 2026-07-23

    Fuller Industries IncOAI

    Great Bend, Kansas, United States · Drugs · Drug Quality Assurance · recorded 2026-10-12

    FEI 1922337
  • Inspection ended 2026-07-13

    Conopco Inc. d/b/a UnileverOAI

    Raeford, North Carolina, United States · Drugs · Drug Quality Assurance · recorded 2026-10-05

    FEI 1047784

Week by week

Each week's page lists every Form 483 and recall. Weeks run Monday to Sunday, so the first and last can include days of the neighbouring months.

2026-W442026-10-26 to 2026-11-0110 letters224 Form 483s120 recalls1 new OAI
2026-W432026-10-19 to 2026-10-257 letters198 Form 483s63 recalls1 new OAI
2026-W422026-10-12 to 2026-10-186 letters0 Form 483s85 recalls4 new OAI
2026-W412026-10-05 to 2026-10-1116 letters2 Form 483s144 recalls1 new OAI
2026-W402026-09-28 to 2026-10-048 letters4 Form 483s70 recalls1 new OAI

Get each week’s page by email every Monday

A free weekly email with each week’s warning letters, new Form 483s, OAI classifications and Class I recalls, counted the same way as this page. Just your email address.

We’ll email once to confirm. Unsubscribe anytime. See last week’s issue →

How this month is counted

Months are calendar months in UTC. Counts are read from FDAgent's copy of each FDA source, which refreshes daily, and FDAgent checks them against FDA's own sources every week. FDA posts some records late and removes others, so a finished month's counts can still change. FDAgent is independent of FDA.

Warning letters posted
Letters from FDA's drug, biologics and device offices, dated by the day FDA posted them, not the day it issued them.
Form 483s listed
Records FDA's Electronic Reading Room types as 483 or Amended 483, dated by FDA's publish date. A firm's 483 response is not counted. Each inspection's 483 counts once, when FDA first listed it: a later listing with the same FEI and record date is shown apart, as a re-post when FDA listed a 483 again under a new media id, or as an amended 483 when FDA types it Amended 483. FDA sometimes lists a 483 days after the publish date it gives it, so a finished month can still gain 483s.
Inspections newly classified
OAI, VAI and NAI inspections counted on the day FDAgent first found them in FDA's classification data, and only when its complete reads in the 8 days before had covered the inspection's end date without finding them. Inspections recorded before FDAgent kept a record of its reads, in November 2026, count when they ended in the 13 weeks before. Backfills and wider reads never count, nor do the Veterinary inspections of FDAgent's first read of them, in November 2026. FDA gives no date for a classification. When FDA removes an inspection from its Data Dashboard, it stops counting, so a finished month's count can fall.
Regulations cited
The FDA Data Dashboard's Form 483 citations on those inspections, by 21 CFR section. FDA often lists an inspection's citations weeks after its classification, so these counts grow after the month ends. For letters, the violation statements that cite a 21 CFR section, each section once per letter.
Recalls
openFDA's drug and device records by Enforcement Report date. Each record is one product; an event is FDA's recall event number.
Import alert entries
Firm entries by the date FDA published them on an import alert. Entries that may name a person are left out. FDA deletes an entry when it removes the firm; entries FDAgent recorded from November 2026 on keep counting in the month FDA dated them.
Months in progress
A month in progress counts the days so far and grows until it ends. Recalls for its latest days appear when openFDA updates.
Comparisons
Every month is counted the same way. A year-over-year figure appears only where FDAgent holds FDA's own dates for the earlier month. A percent change appears only when the earlier count is 10 or more. A month in progress shows no changes until it ends. A month that overlaps a federal government shutdown (Oct 1 to Nov 12, 2025) can have low counts, and month pages say so beside any comparison with one.

Cite this page

Quote a count with its date: the page refreshes daily, and a month's counts can change as FDA posts late records.

FDAgent. "FDA enforcement, October 2026." https://www.fdagent.app/enforcement/2026-10. Counts as of Nov 22, 2026.

About this data

  • Weeks are ISO weeks, Monday to Sunday, in UTC. Each section is dated by its own FDA date: posting date for warning letters, reading-room publish date for Form 483s (each inspection's 483 counts once, when FDA first listed it; FDA's re-posts and amended 483s with the FEI and record date of one it listed before are shown apart), Enforcement Report date for recalls, and the date published FDA prints beside each firm on an import alert. FDA publishes no date for an inspection classification, so an OAI classification counts as new on the day FDAgent first recorded the inspection, and only when FDAgent's complete reads of FDA's inspection data in the 8 days before had covered the inspection's end date without finding it. Older inspections FDAgent adds to its archive, by a backfill or a wider read, never count as new, however late FDA posted them. Inspections FDAgent recorded before it began keeping a record of its reads, in November 2026, count only when they ended in the 13 weeks before, and the Veterinary inspections of FDAgent's first read of them, that same month, never count.
  • Import alert entries come from FDA's import alert pages, where FDA lists firms by name and address, not by FEI. FDAgent links an entry to a facility only when FDA printed the firm's FEI, never by name. Earlier weeks list only entries still on FDA's pages, because FDA deletes an entry when it removes a firm, and FDAgent keeps removed entries only from the day it began recording them. FDAgent lists an entry here only when its name shows it is a firm, through a legal form such as Ltd, Inc or GmbH, a business word such as Laboratories or Trading, or a web address, or when FDA's inspection or drug registration records hold a firm under exactly that name; entries that may name a person are shown only to signed-in users.
  • Import refusals are not listed by week: FDA reports a refusal weeks after it happens, so a week's count would keep changing after the week closed. Each facility page lists its own refusals.
  • Warning letters are those issued by FDA's drug, biologics and device centers and their field and quality offices. Letters from FDA's food, tobacco, veterinary, cosmetics and dietary supplement offices are left out.
  • FDA posts only some Form 483s to its Electronic Reading Room, so a missing 483 does not mean none was issued.
  • FDA publishes an inspection's classification once it is final and updates its inspection data weekly. Source: FDA inspections FAQ.
  • Recalls come from openFDA, which republishes FDA's weekly Enforcement Report; a report's recalls appear here once openFDA publishes it.
  • FDAgent is independent of FDA. Every row links to its FDA source; check it before relying on a record.