September 2026 · 2026-09-01 to 2026-09-30

FDA enforcement, September 2026

In September 2026 (2026-09-01 to 2026-09-30), FDA posted 27 warning letters from its drug, biologics and device offices, and listed 13 Form 483s in its Electronic Reading Room for the first time. FDA's Enforcement Report added 361 drug and device recalls, 38 of them Class I. FDAgent recorded 2 new Official Action Indicated (OAI) classifications in FDA's inspection data. FDA published 28 firm entries on its import alert lists, all of them on a Red List.

Complete month

Counts are FDA's records dated in September as of each source's last sync, which each section states; a late FDA update can still change them.

Warning letters posted
27
On 5 posting dates · 15 in August
Form 483s listed
13
From 13 facilities · 8 in August
Inspections newly classified
244
2 OAI · 67 VAI · 175 NAI
New OAI classifications
2
Newly listed by FDA · 4 in August
Recalled products
361
139 recall events · 38 Class I
Import alert entries
28
28 Red List

Ask FDAgent about this month

FDAgent reads the warning letters FDA posted this month and answers in plain English, with links to the letters.

Edit it or ask as written. 3 questions free, no card needed.

Compared with August 2026 and September 2025

Each measure is counted the same way in every month. A year-over-year figure is shown only where FDAgent holds FDA's own dates for September 2025.

  • Warning letters posted27

    Aug 2026: 15 · +12 (+80%)Sep 2025: 96 · −69 (−72%)

  • Form 483s listed in the reading room13

    Aug 2026: 8 · +5Sep 2025: 10 · +3 (+30%)

  • Inspections newly classified244

    Aug 2026: 309 · −65 (−21%)Sep 2025: —¹

  • New OAI classifications2

    Aug 2026: 4 · −2Sep 2025: —¹

  • Recalled products (drug and device)361

    Aug 2026: 253 · +108 (+43%)Sep 2025: 312 · +49 (+16%)

  • Recall events139

    Aug 2026: 121 · +18 (+15%)Sep 2025: 110 · +29 (+26%)

  • Class I recall events20

    Aug 2026: 16 · +4 (+25%)Sep 2025: 13 · +7 (+54%)

  • Import alert firm entries28

    Aug 2026: 21 · +7 (+33%)Sep 2025: —²

¹ FDAgent has recorded when an inspection first appears in FDA's data only since late January 2026, so months before February 2026 have no count. ² FDA deletes an import alert entry when it removes a firm, and FDAgent keeps removed entries only from November 2026, so earlier months count only the entries FDA still lists, and the year before is not compared. Sources: FDA warning letters page, last synced Nov 22, 2026, 04:45 UTC; Electronic Reading Room, last synced Nov 22, 2026, 15:42 UTC; FDA Data Dashboard inspections, last synced Nov 22, 2026, 04:55 UTC; openFDA Enforcement Report, FDA last updated Nov 11, 2026; FDA import alert pages, last synced Nov 20, 2026, 03:55 UTC.

Inspections FDA newly classified

244 inspections FDAgent first found in FDA's inspection classification data in September, by FDA's classification. FDA gives no date for a classification, so each counts in the month FDAgent first recorded it. When FDA later removes an inspection from its Data Dashboard, it stops counting, so a finished month's count can fall. They ended a median of 41 days before FDAgent first recorded them, and FDA's Data Dashboard lists Form 483 citations for 58 of them. Inspection classifications last synced Nov 22, 2026, 04:55 UTC.

Product typeOAIVAINAITotal
Drugs2343874
Devices0305585
Biologics038285
All267175244

The 2 OAI inspections are listed under New OAI classifications below.

Regulations cited most

The 21 CFR sections behind the Form 483 citations FDA's Data Dashboard lists for the inspections it newly classified in September (53 citations on 28 inspections), and the sections September's warning letters cite in their numbered violations (53 violations in 16 of the 27 letters). Each section links to its page. Inspection citations last synced Nov 22, 2026, 04:55 UTC.

On inspectionsCitations
21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?10
21 CFR 211.67Equipment cleaning and maintenance5
21 CFR 211.22Responsibilities of quality control unit4
21 CFR 211.160General requirements4
21 CFR 211.192Production record review4
21 CFR 820.35Control of records3
21 CFR 211.68Automatic, mechanical, and electronic equipment2
21 CFR 211.166Stability testing2
In warning lettersLetters
21 CFR 211.22Responsibilities of quality control unit7
21 CFR 211.192Production record review7
21 CFR 211.113Control of microbiological contamination6
21 CFR 211.42Design and construction features5
21 CFR 211.84Testing and approval or rejection of components, drug product containers, and closures5
21 CFR 211.100Written procedures; deviations4
21 CFR 211.165Testing and release for distribution3
21 CFR 211.166Stability testing2

Sections with the same count are ordered by section number. A section without its own page is shown without a link. 127 citations name no 21 CFR section and are not counted; device citations can name ISO 13485:2016 clauses instead. FDA can list an inspection's citations weeks after its classification, so these counts can grow.

Form 483s listed in the reading room

13 Form 483s FDA listed for the first time in its Electronic Reading Room in September, dated by the publish date FDA gives each, from 13 facilities. FDA sometimes lists a 483 days after that date, so a finished month can still gain 483s. FDA also listed 1 amended Form 483. Amended 483s don't count as new 483s. FDA posts only some 483s, often long after the inspection: it listed these a median of 41 days after their record date. Each week's page lists every 483 with its PDF. Reading room last synced Nov 22, 2026, 15:42 UTC.

By year of the 483's record date
  • 202613
  • 20250
  • 20240
  • 20230
  • 2022 and earlier0

The most FDA dated to one publish date was 5, on 2026-09-22.

Warning letters

27 letters FDA posted on its warning letters page in September, from its drug, biologics and device offices. 10 were issued in September and 17 earlier; FDA posted them a median of 11 days after issue. Warning letters last synced Nov 22, 2026, 04:45 UTC.

CDER · Drugs
22
CDRH · Devices
3
CBER · Biologics
2
Show 7 more

An FEI links to its facility page only when the letter names exactly one FEI and FDAgent holds FDA records for it.

Recalls by class

Drug and device recalls in FDA's Enforcement Report dated in September, from openFDA, which FDA last updated Nov 11, 2026. Each openFDA record is one recalled product; a recall event can cover many products, so events are counted too.

ClassDrugDeviceEvents
Class I83020
Class II68248112
Class III707
All83278139

The largest Class I recall event, from Medline Industries, LP, covers 7 of the 38 Class I products. The Class I products are listed below. Class II and III recalls are listed on each week's page.

Class I38 recallsShow all
  • Z-3194-2026

    AVID Medical, Inc.

    Halyard IV START KIT NON-STERILE; 48 kits per case. Containing BD ChloraPrep Clear - 1 mL Applicator. Kit Model Number: KRIV26-02.

    Reason: Convenience kits containing BD ChloraPrep Clear - 1 mL Applicator, which may be contaminated with Aspergillus penicillioides.

    Device · initiated 2026-07-22 · Enforcement Report 2026-09-30 · Ongoing

    openFDA record
  • Z-3214-2026

    Boston Scientific Corporation

    Imager II Angiographic Catheters M001314001 IMAGER II/5/ST/65/038 BX5; M001314011 IMAGER II/5/ST/100/038 BX5; M001314021 IMAGER II/5/ST/65/035 BX 5; M001314031 IMAGER II/5/ST/100/035 BX 5; M001314041 IMAGER II/5/BARNH/100/038 BX5; M001314051 IMAGER II/5/BERN/100/038 BX5; M001314…

    Reason: Tip may fracture or detach, leading to device fragments remaining within the patient or migrating through the vasculature. Patient impact may include prolonged procedure or life-threatening embolism.

    Device · initiated 2026-08-24 · Enforcement Report 2026-09-30 · Ongoing

    openFDA record
  • Z-3139-2026

    CooperSurgical, Inc.

    INCA (Infant Nasal Cannulae Assy) Infant Nasal CPAP, Complete Set. Stay-Flex tubing component. 5 units per box. Model Numbers: 44-2707, 44-2709, 44-2710, 44-2712, 44-2715.

    Reason: Potential to exhibit a loose connection between the flexible tubing and the nasal cannula, during assembly or while in use, which may result in loss of CPAP pressure, possibly reducing the effectiveness of treatment and lead to decreased oxygenation until connection is restored.

    Device · initiated 2026-08-11 · Enforcement Report 2026-09-30 · Ongoing

    openFDA record
  • Z-3140-2026

    CooperSurgical, Inc.

    INCA (Infant Nasal Cannulae Assy) Neonatal Nasal CPAP, Replacement Set. Stay-Flex tubing component. 5 units per box. Model Numbers: 44-0712, 44-0707, 44-0709, 44-0715.

    Reason: Potential to exhibit a loose connection between the flexible tubing and the nasal cannula, during assembly or while in use, which may result in loss of CPAP pressure, possibly reducing the effectiveness of treatment and lead to decreased oxygenation until connection is restored.

    Device · initiated 2026-08-11 · Enforcement Report 2026-09-30 · Ongoing

    openFDA record
  • Z-3148-2026

    Abiomed, Inc.

    Automated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 2. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU. 3. Impella Controller, Packaged, IN; Product…

    Reason: The potential for purge cassette recognition issues due to the failure of the purge flag component of the purge pressure sensor assembly.

    Device · initiated 2026-08-12 · Enforcement Report 2026-09-23 · Ongoing

    openFDA record
  • Z-3149-2026

    Abiomed, Inc.

    Impella Optical Controller with the below product descriptions and corresponding Product Codes. 1. Impella Optical Controller, Packaged, AU; Product Code: 0042-0010-AU. 2. Impella Optical Controller, Packaged, EU; Product Code: 0042-0010-EU. 3. Impella Optical Controller, Packag…

    Reason: The potential for purge cassette recognition issues due to the failure of the purge flag component of the purge pressure sensor assembly.

    Device · initiated 2026-08-12 · Enforcement Report 2026-09-23 · Ongoing

    openFDA record
  • Z-3150-2026

    Abiomed, Inc.

    Impella Optical Controller, Automated Impella Controller, Impella Connect Package with the below product descriptions and corresponding Product Codes. 1. Optical, AIC, Impella Connect, Packaged, AU; Product Code: 0042-0040-AU. 2. Optical, AIC, Impella Connect, Packaged, CA; Prod…

    Reason: The potential for purge cassette recognition issues due to the failure of the purge flag component of the purge pressure sensor assembly.

    Device · initiated 2026-08-12 · Enforcement Report 2026-09-23 · Ongoing

    openFDA record
  • Z-3108-2026

    AVID Medical, Inc.

    Halyard IV START KIT NON-STERILE REF: KRIV26-02

    Reason: Convenience kits contain BD ChloraPrep Clear - 1 mL and FREPP Clear 1.5 mL Applicators which are contaminated with Aspergillus penicillioides

    Device · initiated 2026-07-14 · Enforcement Report 2026-09-23 · Ongoing

    openFDA record
  • Z-3151-2026

    AVID Medical, Inc.

    Haylard 18G SINGLE SHOT EPI TRAY; Model/Item Number (Kit): DPGE002-02;

    Reason: Potential for product quality and sterility issues regarding kits containing Sodium Chloride ampules subject to recall by Spectra Medical.

    Device · initiated 2026-08-10 · Enforcement Report 2026-09-23 · Ongoing

    openFDA record
  • Z-3152-2026

    AVID Medical, Inc.

    Haylard CONTINUOUS NERVE BLOCK TRAY; Model/Item Number (Kit): BMCN001-05;

    Reason: Potential for product quality and sterility issues regarding kits containing Sodium Chloride ampules subject to recall by Spectra Medical.

    Device · initiated 2026-08-10 · Enforcement Report 2026-09-23 · Ongoing

    openFDA record
  • Z-3153-2026

    AVID Medical, Inc.

    Haylard BASIC BIOPSY TRAY; Model/Item Number (Kit): PM415;

    Reason: Potential for product quality and sterility issues regarding kits containing Sodium Chloride ampules subject to recall by Spectra Medical.

    Device · initiated 2026-08-10 · Enforcement Report 2026-09-23 · Ongoing

    openFDA record
  • Z-3154-2026

    AVID Medical, Inc.

    Haylard UNIV IMAGING BIOPSY TRUCUSTOM; Model/Item Number (Kit): VARV042-03;

    Reason: Potential for product quality and sterility issues regarding kits containing Sodium Chloride ampules subject to recall by Spectra Medical.

    Device · initiated 2026-08-10 · Enforcement Report 2026-09-23 · Ongoing

    openFDA record
  • Z-2984-2026

    Bard Access Systems, Inc.

    Power-Trialysis Short-Term Straight Dialysis Catheter, REF: 5605150; Power-Trialysis Short-Term Straight Dialysis Catheter, REF: 5605200; Power-Trialysis Short-Term Straight Dialysis Catheter, REF: 5605240; Power-Trialysis Short-Term Straight Dialysis Catheter, REF: 5605300; Pow…

    Reason: this is a downstream recall of class I recall RES 98777

    Device · initiated 2026-08-10 · Enforcement Report 2026-09-23 · Ongoing

    openFDA record
  • Z-2985-2026

    Bard Access Systems, Inc.

    DuoGlide Short-Term Straight Dialysis Catheter, REF: 5625150; DuoGlide Short-Term Straight Dialysis Catheter, REF: 5625200; DuoGlide Short-Term Straight Dialysis Catheter, REF: 5625240; DuoGlide Short-Term Straight Dialysis Catheter, REF: 5625300; DuoGlide Short-Term Curved Exte…

    Reason: this is a downstream recall of class I recall RES 98777

    Device · initiated 2026-08-10 · Enforcement Report 2026-09-23 · Ongoing

    openFDA record
  • Z-3080-2026

    BD SWITZERLAND SARL

    BD Alaris Pump Infusion Sets 10561554* 10885403232565 50885403232563 BD Alaris Pump Infusion Set, 15 Micron Filter, Back Check Valve, 3 SmartSite Y-sites 10015861A 10885403232626 50885403232624 BD Alaris Pump Infusion Set, SmartSite Bag Access Non-Vented, PE Lined Tubing, Back C…

    Reason: Due to discrepancies in specific attributes in the performance data for infusion sets, which may lead to inappropriate pump performance.

    Device · initiated 2026-08-19 · Enforcement Report 2026-09-23 · Ongoing

    openFDA record
  • Z-3121-2026

    Medline Industries, LP

    Medline Convenience Kits: 1) KIT ANESTHESIA L&D A- LINE SET, Catalog Number: DYKMBNDL199A

    Reason: The kits contain certain lots of sterile applicators which may contain fungal contamination that may occur under certain environmental conditions, allowing the growth of Aspergillus penicillioides. Contamination of skin preparation products with Aspergillus penicillioides may le…

    Device · initiated 2026-08-11 · Enforcement Report 2026-09-23 · Ongoing

    openFDA record
  • Z-3122-2026

    Medline Industries, LP

    Medline Convenience Kits: 1) OPEN HEART ADULT, Catalog Number: CDS840396Y; 2) FULL LINE ANEURYSM CDS, Catalog Number: CDS983672O; 3) NON STERILE CHEST TUBE REMOVAL, Catalog Number: DYNDA1080

    Reason: The kits contain certain lots of sterile applicators which may contain fungal contamination that may occur under certain environmental conditions, allowing the growth of Aspergillus penicillioides. Contamination of skin preparation products with Aspergillus penicillioides may le…

    Device · initiated 2026-08-11 · Enforcement Report 2026-09-23 · Ongoing

    openFDA record
  • Z-3123-2026

    Medline Industries, LP

    Medline Convenience Kits: 1) PORT ACCESS KIT UNIV OF ALA, Catalog Number: DYKM1414; 2) ARTERIAL LINE ST JOES SYR PACK, Catalog Number: DYNJ38145F

    Reason: The kits contain certain lots of sterile applicators which may contain fungal contamination that may occur under certain environmental conditions, allowing the growth of Aspergillus penicillioides. Contamination of skin preparation products with Aspergillus penicillioides may le…

    Device · initiated 2026-08-11 · Enforcement Report 2026-09-23 · Ongoing

    openFDA record
  • Z-3124-2026

    Medline Industries, LP

    Medline Convenience Kits: 1) ACC010308A, Catalog Number: ACC010308A; 2) ACC010374A, Catalog Number: ACC010374A; 3) ACC010679, Catalog Number: ACC010679; 4) ACC010726A, Catalog Number: ACC010726A; 5) DYNDB3061, Catalog Number: DYNDB3061; 6) DYNDH1245, Catalog Number: DYNDH1245; 7…

    Reason: The kits contain certain lots of sterile applicators which may contain fungal contamination that may occur under certain environmental conditions, allowing the growth of Aspergillus penicillioides. Contamination of skin preparation products with Aspergillus penicillioides may le…

    Device · initiated 2026-08-11 · Enforcement Report 2026-09-23 · Ongoing

    openFDA record
  • Z-3125-2026

    Medline Industries, LP

    Medline Convenience Kits: 1) DYKM1000B, Catalog Number: DYKM1000B; 2) DYKM2187A, Catalog Number: DYKM2187A

    Reason: The kits contain certain lots of sterile applicators which may contain fungal contamination that may occur under certain environmental conditions, allowing the growth of Aspergillus penicillioides. Contamination of skin preparation products with Aspergillus penicillioides may le…

    Device · initiated 2026-08-11 · Enforcement Report 2026-09-23 · Ongoing

    openFDA record
  • Z-3126-2026

    Medline Industries, LP

    Medline Convenience Kits: 1) DYNDH1193, Catalog Number: DYNDH1193; 2) DYNDH1487A, Catalog Number: DYNDH1487A

    Reason: The kits contain certain lots of sterile applicators which may contain fungal contamination that may occur under certain environmental conditions, allowing the growth of Aspergillus penicillioides. Contamination of skin preparation products with Aspergillus penicillioides may le…

    Device · initiated 2026-08-11 · Enforcement Report 2026-09-23 · Ongoing

    openFDA record
  • Z-3127-2026

    Medline Industries, LP

    Medline Convenience Kits: 1) DYNDB3086D, Catalog Number: DYNDB3086D; 2) DYNDB3100, Catalog Number: DYNDB3100; 3) DYNDH1619A, Catalog Number: DYNDH1619A; 4) DYNDH1620A, Catalog Number: DYNDH1620A; 5) DYNDH1694A, Catalog Number: DYNDH1694A; 6) DYNDH1695, Catalog Number: DYNDH1695;…

    Reason: The kits contain certain lots of sterile applicators which may contain fungal contamination that may occur under certain environmental conditions, allowing the growth of Aspergillus penicillioides. Contamination of skin preparation products with Aspergillus penicillioides may le…

    Device · initiated 2026-08-11 · Enforcement Report 2026-09-23 · Ongoing

    openFDA record
  • Z-3128-2026

    Nova Biomedical Corporation

    StatStrip Glucose Hospital Meter System. Model No: 63685. Glucose test system.

    Reason: A software synchronization/timing issue in the barcode scanning and patient data processing component that causes the device to intermittently retain a previously scanned barcode including patient identification (ID) barcodes and quality control (QC) solution ID barcodes when a…

    Device · initiated 2026-07-29 · Enforcement Report 2026-09-23 · Ongoing

    openFDA record
  • Z-3129-2026

    Nova Biomedical Corporation

    StatStrip Glucose Hospital Meter System. Model No: 66739. Glucose test system.

    Reason: A software synchronization/timing issue in the barcode scanning and patient data processing component that causes the device to intermittently retain a previously scanned barcode including patient identification (ID) barcodes and quality control (QC) solution ID barcodes when a…

    Device · initiated 2026-07-29 · Enforcement Report 2026-09-23 · Ongoing

    openFDA record
  • Z-3130-2026

    Nova Biomedical Corporation

    StatStrip Glucose and B- Ketone meter. Model No: 63683. Glucose test system.

    Reason: A software synchronization/timing issue in the barcode scanning and patient data processing component that causes the device to intermittently retain a previously scanned barcode including patient identification (ID) barcodes and quality control (QC) solution ID barcodes when a…

    Device · initiated 2026-07-29 · Enforcement Report 2026-09-23 · Ongoing

    openFDA record
  • Z-3131-2026

    Nova Biomedical Corporation

    StatStrip Glucose and B- Ketone meter. Model No: 63910. Glucose test system.

    Reason: A software synchronization/timing issue in the barcode scanning and patient data processing component that causes the device to intermittently retain a previously scanned barcode including patient identification (ID) barcodes and quality control (QC) solution ID barcodes when a…

    Device · initiated 2026-07-29 · Enforcement Report 2026-09-23 · Ongoing

    openFDA record
  • Z-3132-2026

    Nova Biomedical Corporation

    StatStrip Glucose and B- Ketone meter. Model No: 66429. Glucose test system.

    Reason: A software synchronization/timing issue in the barcode scanning and patient data processing component that causes the device to intermittently retain a previously scanned barcode including patient identification (ID) barcodes and quality control (QC) solution ID barcodes when a…

    Device · initiated 2026-07-29 · Enforcement Report 2026-09-23 · Ongoing

    openFDA record
  • D-0866-2026

    VITRUVIAS THERAPEUTICS INC

    Thyroid Tablets, USP, 1/2 Grain (30 mg), Each tablet contains: levothyroxine (T4) 19 mcg, liothyronine (T3) 4.5 mcg, 100 Tablets, Rx only, Distributed by: Vitruvias Therapeutics, Auburn, AL 36830, Product of USA, NDC 69680-166-00.

    Reason: Superpotent Drug

    Drug · initiated 2026-08-21 · Enforcement Report 2026-09-23 · Ongoing

    openFDA record
  • D-0865-2026

    American Regent, Inc.

    Epinephrine Injection, USP, 30 mg/30 mL (1mg/mL), 30 mL-Multiple-Dose Vial, Rx Only, For Intravenous Infusion, intramuscular and Subcutaneous Use, AMERICAN REGENT INC., SHIRLEY, NY 11967. NDC 0517-3030-01

    Reason: Presence of Particulate Matter and Lack of Assurance of Sterility. Particulate matter identified as nylon, cellulosic, acrylic, polyethylene and glass and broken and leaking vials.

    Drug · initiated 2026-08-11 · Enforcement Report 2026-09-16 · Ongoing

    openFDA record
  • D-0869-2026

    B BRAUN MEDICAL INC

    0.9% Sodium Chloride Injection, USP, 100 mL Partial Fill in 150 mL PAB Container, Sterile, Rx only, B. Braun Medical Inc., Bethlehem, PA 18018 USA, NDC 0264-1800-32.

    Reason: Presence of Particulate Matter: particulate matter identified as iron oxide, inorganic material, cellulose cotton fiber or polystyrene

    Drug · initiated 2026-08-19 · Enforcement Report 2026-09-16 · Ongoing

    openFDA record
  • D-0850-2026

    Baxter Healthcare Corporation

    Dextrose Injection, USP, 70 %, 2000 mL bags, Rx Only, Baxter Healthcare Corporation Deerfield, IL, 60016, Made in USA, NDC 00338-0719-06

    Reason: Presence of Particulate Matter: particulate matter identified as stainless steel particles in the solution.

    Drug · initiated 2026-08-17 · Enforcement Report 2026-09-16 · Ongoing

    openFDA record
  • Z-3090-2026

    Given Imaging Ltd.

    Bravo CF capsule delivery device, 5-pack, Product Number FGS-0635, Bravo CF capsule delivery device, 1-pack, Product Number FGS-0636. Used to attach a pH monitoring capsule to the esophagus.

    Reason: Capsule may fail to attach to the esophagus or detach from the delivery device. Risks include patient aspiration/inhalation, perforation of esophagus, obstruction of the airway, hemorrhage/blood loss/bleeding, laceration of the esophagus, a delay in diagnosis, and foreign bodies…

    Device · initiated 2026-08-06 · Enforcement Report 2026-09-16 · Ongoing

    openFDA record
  • Z-3106-2026

    Medical Action Industries, Inc. 306

    Kit: Epidural/Pain TQC 20/Cs, REF: 82465 Containing Sodium Chloride Injection, 0.9%, USP 10ml ampules

    Reason: Kits containing Sodium Chloride ampules subject to another recall are recalled due to potentially compromised sterility.

    Device · initiated 2026-07-23 · Enforcement Report 2026-09-16 · Ongoing

    openFDA record
  • Z-3107-2026

    Medical Action Industries, Inc. 306

    Kit: Single Shot Epidural 20/Cs, REF: 59553; Kit: Single Shot Epidural 20/Cs, REF: 83945B. Containing Sodium Chloride Injection, 0.9%, USP 10ml ampules

    Reason: Kits containing Sodium Chloride ampules subject to another recall are recalled due to potentially compromised sterility.

    Device · initiated 2026-07-23 · Enforcement Report 2026-09-16 · Ongoing

    openFDA record
  • D-0849-2026

    Baxter Healthcare Corporation

    0.9% Sodium Chloride Injection USP 500 mL, VIAFLEX Plastic Containers, with 24 units per case. Baxter Healthcare Corporation, Deerfield, IL, 60015, USA, Made in USA, NDC 0338-0049-03

    Reason: Presence of particulate matter: Particulates identified as fiberglass

    Drug · initiated 2026-08-17 · Enforcement Report 2026-09-09 · Ongoing

    openFDA record
  • D-0851-2026

    OPTIMAL BALANCE PHARMACY

    Glutathione 200 mg/mL, 30 mL sterile multidose vial, Each vial contains: Glutathione, Ascorbic Acid, Benzyl Alcohol & sterile water for injection, For IM or IV Injection Use Only, RX only, Optimal Balance Pharmacy, 2204 Cypress Creek Pkwy Suite F, Houston, TX 77090

    Reason: Microbial Contamination of Sterile Products - out of specifications results were obtained for bacterial endotoxin.

    Drug · initiated 2026-08-05 · Enforcement Report 2026-09-09 · Ongoing

    openFDA record
  • D-0832-2026

    B BRAUN MEDICAL INC

    Lactated Ringer's Injection USP, 1000 mL EXCEL container, Rx only, L7500, B. Braun Medical, Inc., Bethlehem, PA 18018-3524 USA, NDC 0264-7750-00

    Reason: Presence of Particulate matter.

    Drug · initiated 2026-08-12 · Enforcement Report 2026-09-02 · Ongoing

    openFDA record
  • D-0827-2026

    Fresenius Kabi USA, LLC

    Tyenne (tocilizumab-aazg) Injection, 400 mg/20 mL (20 mg/mL), 20 mL vial, Rx Only, Manufactured by Fresenius Kabi USA, LLC, Lake Zurich, Illinois, 60047, NDC 65219-594-20.

    Reason: Presence of particulate matter:An internal investigation at the firm found the product to contain glass particles.

    Drug · initiated 2026-08-10 · Enforcement Report 2026-09-02 · Ongoing

    openFDA record

Import alert entries

28 firm entries FDA published on its import alert pages this month, dated by each entry's date published: 28 on a Red List. FDA lists firms by name and address; FDAgent links an entry to a facility only when FDA printed the firm's FEI. 3 entries whose name may be a person's rather than a firm's are not listed or counted here; signed-in users can see them in FDAgent. Import alerts last synced Nov 20, 2026, 03:55 UTC.

66-40Red List4 entriesShow allDetention Without Physical Examination of Drugs From Firms Which Have Not Met Drug GMPsAlert on fda.gov
  • Published 2026-09-30

    Shenzhen Benliju Bio-Tech Co., Ltd.

    CHINA

  • Published 2026-09-21

    Sanrace Biotechnology Co., Ltd.

    CHINA

  • Published 2026-09-19

    Chengdu Shengnuo Biopharm Co., Ltd.

    CHINA

  • Published 2026-09-03

    PRITAM INTERNATIONAL PRIVATE LIMITED

    INDIA

66-41Red List4 entriesShow allDetention Without Physical Examination of Unapproved New Drugs Promoted In The U.S.Alert on fda.gov
  • Published 2026-09-29

    Cosmocos Co., Ltd.

    KOREA (THE REPUBLIC OF)

  • Published 2026-09-25

    LUEN HOP PHARMACEUTICAL COMPANY

    HONG KONG SAR

  • Published 2026-09-25

    Naturo Aid Pharmaceutical Inc.

    CANADA

  • Published 2026-09-11

    EIGHTJUICE - LABORATORIO NUTRACERUTICO, LDA

    PORTUGAL

66-79Red List2 entriesShow allDetention Without Physical Examination of Drugs From Foreign Establishments Refusing FDA InspectionAlert on fda.gov
  • Published 2026-09-23

    Horster Biotek Private Limited

    INDIA

  • Published 2026-09-15

    Jiangsu Grand Xinyi Pharmaceutical Co., Ltd

    CHINA

68-19Red List1 entryShow allDETENTION WITHOUT PHYSICAL EXAMINATION OF UNAPPROVED FINISHED NEW ANIMAL DRUGSAlert on fda.gov
  • Published 2026-09-30

    Shandong Runxin Biotechnology Co., Ltd.

    CHINA

79-01Red List1 entryShow allDetention Without Physical Examination of Topical/External Bandages, Dressings, Gauzes/Sponges and Cotton Pads Due to Microbiological ContaminationAlert on fda.gov
  • Published 2026-09-04

    Zhejiang Longterm Medical Technology Co., LTD

    CHINA

89-08Red List1 entryShow allDetention Without Physical Examination of Devices without Approved PMA's or IDE's and Other Devices Not Substantially Equivalent or Without a 510(k)Alert on fda.gov
  • Published 2026-09-21

    GMP MEDICARE SDN BHD

    MALAYSIA

95-04Red List6 entriesShow allDetention Without Physical Examination Of Laser Pointers, Laser Gunsights, Laser Levels, Laser Light Shows, Laser Pointer Key Chains, & Similar Products That Fail To Comply With Applicable Performance Standards And Reporting Requirements.Alert on fda.gov
  • Published 2026-09-25

    Shanghai Sinostar Technology Co., Ltd.

    CHINA

  • Published 2026-09-25

    Yicheng Junhao Electronic Technology Co., Ltd.

    CHINA

  • Published 2026-09-23

    Foshan Chiding Optoelectronics Co., Ltd.

    CHINA

  • Published 2026-09-21

    Beijing Huitejiayi Technology Development Center

    CHINA

  • Published 2026-09-16

    Comnav Technology Ltd.

    CHINA

  • Published 2026-09-09

    SHENZHEN CHENGYUANXIANG TECHNOLOGY CO.,LTD

    CHINA

95-05Red List6 entriesShow allDetention Without Physical Examination of Electronic Products that Fail to Comply with Performance Standards or to Have Required CertificationAlert on fda.gov
  • Published 2026-09-30

    Shangqiu Yuxin Tools Co., Ltd

    CHINA

  • Published 2026-09-28

    Shanghai Hao Min Digital Technology Co., Ltd.

    CHINA

  • Published 2026-09-21

    LIAOCHENG YONGXIN LASER EQUIPMENT CO.,LTD

    CHINA

  • Published 2026-09-15

    HESING INTERNATIONAL TRADING

    CHINA

  • Published 2026-09-15

    HESING INTERNATIONAL TRADING COLTD

    CHINA

  • Published 2026-09-09

    ZHEJIANG INNOVATION LASER EQUIPMENT CO., LTD

    CHINA

99-34Red List2 entriesShow allDETENTION WITHOUT PHYSICAL EXAMINATION OF DRUGS OR MEDICAL DEVICES FROM FIRMS WITHOUT A VALID DRUG OR MEDICAL DEVICE REGISTRATION AND/OR LISTINGAlert on fda.gov
  • Published 2026-09-17

    BNC Global

    KOREA (THE REPUBLIC OF)

  • Published 2026-09-03

    BIOPHARMA ASSOCIATES

    INDIA

99-39Red List1 entryShow allDetention Without Physical Examination of Imported Food Products That Appear to Be MisbrandedAlert on fda.gov
  • Published 2026-09-23

    Naturo Aid Pharmaceutical Inc.

    CANADA

New OAI classifications

2 inspections classified Official Action Indicated that FDAgent first recorded from FDA's inspection classification data this month, after its reads of that data in the 8 days before had not found them. Older inspections FDAgent adds to its archive are left out. Dates are inspection end dates. Inspection classifications last synced Nov 22, 2026, 04:55 UTC.

  • Inspection ended 2026-07-10

    Playtex Manufacturing IncOAI

    Ormond Beach, Florida, United States · Drugs · Drug Quality Assurance · recorded 2026-09-28

    FEI 1027735
  • Inspection ended 2026-06-12

    Product Society LLCOAI

    North Hollywood, California, United States · Drugs · Drug Quality Assurance · recorded 2026-09-07

    FEI 3023272193

Week by week

Each week's page lists every Form 483 and recall. Weeks run Monday to Sunday, so the first and last can include days of the neighbouring months.

2026-W392026-09-21 to 2026-09-276 letters5 Form 483s98 recalls0 new OAI
2026-W382026-09-14 to 2026-09-201 letter0 Form 483s26 recalls0 new OAI
2026-W372026-09-07 to 2026-09-132 letters5 Form 483s98 recalls1 new OAI
2026-W362026-08-31 to 2026-09-0610 letters1 Form 48369 recalls0 new OAI

Get each week’s page by email every Monday

A free weekly email with each week’s warning letters, new Form 483s, OAI classifications and Class I recalls, counted the same way as this page. Just your email address.

We’ll email once to confirm. Unsubscribe anytime. See last week’s issue →

How this month is counted

Months are calendar months in UTC. Counts are read from FDAgent's copy of each FDA source, which refreshes daily, and FDAgent checks them against FDA's own sources every week. FDA posts some records late and removes others, so a finished month's counts can still change. FDAgent is independent of FDA.

Warning letters posted
Letters from FDA's drug, biologics and device offices, dated by the day FDA posted them, not the day it issued them.
Form 483s listed
Records FDA's Electronic Reading Room types as 483 or Amended 483, dated by FDA's publish date. A firm's 483 response is not counted. Each inspection's 483 counts once, when FDA first listed it: a later listing with the same FEI and record date is shown apart, as a re-post when FDA listed a 483 again under a new media id, or as an amended 483 when FDA types it Amended 483. FDA sometimes lists a 483 days after the publish date it gives it, so a finished month can still gain 483s.
Inspections newly classified
OAI, VAI and NAI inspections counted on the day FDAgent first found them in FDA's classification data, and only when its complete reads in the 8 days before had covered the inspection's end date without finding them. Inspections recorded before FDAgent kept a record of its reads, in November 2026, count when they ended in the 13 weeks before. Backfills and wider reads never count, nor do the Veterinary inspections of FDAgent's first read of them, in November 2026. FDA gives no date for a classification. When FDA removes an inspection from its Data Dashboard, it stops counting, so a finished month's count can fall.
Regulations cited
The FDA Data Dashboard's Form 483 citations on those inspections, by 21 CFR section. FDA often lists an inspection's citations weeks after its classification, so these counts grow after the month ends. For letters, the violation statements that cite a 21 CFR section, each section once per letter.
Recalls
openFDA's drug and device records by Enforcement Report date. Each record is one product; an event is FDA's recall event number.
Import alert entries
Firm entries by the date FDA published them on an import alert. Entries that may name a person are left out. FDA deletes an entry when it removes the firm; entries FDAgent recorded from November 2026 on keep counting in the month FDA dated them.
Months in progress
A month in progress counts the days so far and grows until it ends. Recalls for its latest days appear when openFDA updates.
Comparisons
Every month is counted the same way. A year-over-year figure appears only where FDAgent holds FDA's own dates for the earlier month. A percent change appears only when the earlier count is 10 or more. A month in progress shows no changes until it ends. A month that overlaps a federal government shutdown (Oct 1 to Nov 12, 2025) can have low counts, and month pages say so beside any comparison with one.

Cite this page

Quote a count with its date: the page refreshes daily, and a month's counts can change as FDA posts late records.

FDAgent. "FDA enforcement, September 2026." https://www.fdagent.app/enforcement/2026-09. Counts as of Nov 22, 2026.

About this data

  • Weeks are ISO weeks, Monday to Sunday, in UTC. Each section is dated by its own FDA date: posting date for warning letters, reading-room publish date for Form 483s (each inspection's 483 counts once, when FDA first listed it; FDA's re-posts and amended 483s with the FEI and record date of one it listed before are shown apart), Enforcement Report date for recalls, and the date published FDA prints beside each firm on an import alert. FDA publishes no date for an inspection classification, so an OAI classification counts as new on the day FDAgent first recorded the inspection, and only when FDAgent's complete reads of FDA's inspection data in the 8 days before had covered the inspection's end date without finding it. Older inspections FDAgent adds to its archive, by a backfill or a wider read, never count as new, however late FDA posted them. Inspections FDAgent recorded before it began keeping a record of its reads, in November 2026, count only when they ended in the 13 weeks before, and the Veterinary inspections of FDAgent's first read of them, that same month, never count.
  • Import alert entries come from FDA's import alert pages, where FDA lists firms by name and address, not by FEI. FDAgent links an entry to a facility only when FDA printed the firm's FEI, never by name. Earlier weeks list only entries still on FDA's pages, because FDA deletes an entry when it removes a firm, and FDAgent keeps removed entries only from the day it began recording them. FDAgent lists an entry here only when its name shows it is a firm, through a legal form such as Ltd, Inc or GmbH, a business word such as Laboratories or Trading, or a web address, or when FDA's inspection or drug registration records hold a firm under exactly that name; entries that may name a person are shown only to signed-in users.
  • Import refusals are not listed by week: FDA reports a refusal weeks after it happens, so a week's count would keep changing after the week closed. Each facility page lists its own refusals.
  • Warning letters are those issued by FDA's drug, biologics and device centers and their field and quality offices. Letters from FDA's food, tobacco, veterinary, cosmetics and dietary supplement offices are left out.
  • FDA posts only some Form 483s to its Electronic Reading Room, so a missing 483 does not mean none was issued.
  • FDA publishes an inspection's classification once it is final and updates its inspection data weekly. Source: FDA inspections FAQ.
  • Recalls come from openFDA, which republishes FDA's weekly Enforcement Report; a report's recalls appear here once openFDA publishes it.
  • FDAgent is independent of FDA. Every row links to its FDA source; check it before relying on a record.