Bentley Laboratories LLC
111 Fieldcrest Ave, Edison, New Jersey, United States
FEI 2244819DUNS 068351753
FDA has 6 inspections on file for Bentley Laboratories LLC in Edison, New Jersey (FEI 2244819), from 2009 to 2026. The latest ended on and was classified OAI, official action indicated. FDA issued a warning letter on (CDER).
OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.
FDA posted the Form 483 from that inspection. FDAgent read its 9 observations from the PDF; they are listed below. One of the two FOIA requests for this site has asked FDA for this inspection's 483.
No court action or import refusal is filed under this FEI.
Three regulations were cited at more than one inspection: 21 CFR 211.115(a), 21 CFR 211.192 and 21 CFR 211.84(d)(2).
As of Nov 21, 2026, the site is registered with FDA as a drug establishment for manufacturing, through Dec 31, 2027. Its registrant is Bentley Laboratories, LLC (D-U-N-S 068351753).
Sources:FDA firm profile ↗Form 483 PDF ↗Warning letter ↗FDA drug establishment registrations ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- OAI
- Inspections on file
- 6, 2009–2026
- 1 OAI · 5 VAI · 0 NAI
- Inspections with 483 observations shown
- 4 of 6
- Form 483s posted by FDA
- ↗
- Requested under FOIA
- 2 requests, latest May 18, 2026
- Warning letters
- · CDER ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
- Drug establishment registration
- Through
Inspection history
6 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- OAISignificant objectionable conditions; FDA recommended regulatory actionDrugs · Drug Quality AssurancePosted 483 · 9 observations
Read this Form 483 on its own page →Form 483 PDF on fda.gov
Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent’s.
- Observation 1
Incoming component testing and supplier qualification
Reports of analysis from component suppliers are accepted in lieu of testing each component for conformity with all appropriate written specifications, without performing at least one specific identity test on each component and establishing the reliability of the supplier’s analyses through appropriate validation of the supplier's test results at appropriate intervals. Specifically, 1) Only FTIR identification test is performed on each shipment of each incoming lot of the following ingredients used in manufacturing OTC drug products and the following specific identification tests are not conducted: a. Limit of (b) (4) testing is not conducted on incoming shipments of (b) (4) which is used to manufacture OTC drug products including, but not limited to, (b) (4) (b) (4) + Lot#(b) (4) (Exp: (b)(4)). b. (b) (4) and (b) (4) limit testing is not conducted on incoming shipments of (b) (4) USP, (b) (4) USP, and (b) (4) USP. (b) (4) (b) (4), and (b) (4) are used to manufacture OTC drug products including, but not limited to, (b) (4) (b) (4) (Lot# (b) (4) Exp: (b) (4)), (b) (4) (Lot# (b) (4) Exp (b) (@)) and (b) (4) (Lot#(b) (4) Exp: (b) (4)). respectively. 2) There are no written procedures for qualifying suppliers of raw materials and active pharmaceutical ingredients (APIs). For example, the suppliers for API (b) (4) and raw materials such as (b) (4). [illegible]
- Observation 2
Laboratory release lacking B. cepacia testing
Testing and release of drug product for distribution do not include appropriate laboratory determination of satisfactory conformance to the final specifications, identity and strength of each active ingredient prior to release. Specifically, (b) (4) based non-sterile finished OTC drug products and the (b) (4), used in manufacturing OTC drug products, are not routinely tested for Burkholderia cepacia (B. cepacia). No documented investigations were conducted when B. cepacia was identified during (b) (4) testing of your (b) (4) system on the following dates: July 3, 2024 ((b) (4)) November 4, 2024 ((b) (4)) September 3, 2025 ((b) (4)) September 8, 2025 ((b) (4)) October 24, 2025 ((b) (4)) The following OTC drug products were manufactured with (b) (4) on (b) (4), and (b) (4): respectively, and distributed: (b)(4). Bulk Lot #(b) (4), Finished Good Lot #(b) (4) (Expiration: (b) (4)) (b) (4). Bulk Lot #(b) (4). Finished Good Lot #(b) (4) (Expiration: (b) (4))
Show observations 3 to 9
- Observation 3
Inadequate investigations of OOS and discrepancies
There is a failure to thoroughly review any unexplained discrepancy, the failure of a batch or any of its components to meet any of its specifications whether or not the batch has been already distributed. Specifically, 1) The stability results at P¥°C#%, RH for two lots of (b) (4) demonstrated the following OOS results: Lot # | Time Point | Assay (Specification) | OOS Results | Color/Appearance (Specification) | OOS Result Lot (b) (4) | (b) (4) | 2.44% (Spec: (b) (4) %) | Color: Gray (Spec: (b) (4)) | Appearance: Soft Stick/Cream (Spec: (b) (4)) | i: 5.74% Lot (b) (4) | (b) (4) | 2.67% (b) (4) | Color: Gray | Appearance: Cream (b) (4) 4.36% The investigation (Non-Conformance Report (NCR) # 24-093), initiated on 12/27/2024, for these OOS assay results and appearance failures stated that bulk retains and finished good retains were pulled from pilot to most recent batches and confirmed that all samples showed “minor to major signs of softening”. Despite physical degradation observed across multiple batches, only (b) (4) lots of retain samples were submitted for (b) (4) testing and no Assay testing was conducted on any of the retain samples that exhibited softening. 2) NCR # 24-032 was initiated on 04/01/2024 for failed microbial results for Total Aerobic Microbial Count (TAMC) and Total Yeast and Mold Count (TYMC) (Results: >57,000 CFU/g, Specification: TAMC: (b) (4) CFU/g, TYMC: (b) (4) CFU/g) for (b) (4), Lot#(b) (4). The investigation failed to include a documented review of microbial results of (b) (4) used to manufacture the OTC drug product. Additionally, the root cause was identified as the hoses, used to drop the product from the kettles into storage (b) (4), may not have been cleaned and sanitized thoroughly; however, the Preventive Action only states that hoses will be labeled to have traceability but does not include actions to ensure adequate cleaning and sanitization of the hoses. 3) An Out of Specification (OOS) for (b) (4), Batch #(b) (4), Exp Date: (b) (4), Fill Date: (b) (4) was received for (b) (4) (Result: 5.96; Specification: (b) (4)) at (b) (4) at P8°CPE%RH. There was no investigation or documented review of the formulation or (b) (4) system to identify the underlying root cause for the (b) (4) drift (b) (4) the original specification. Following the OOS result at (b) (4) stability testing, the (b) (4) specification range was broadened to (b) (4) without scientific justification.
- Observation 4
Method suitability not verified
The suitability of all testing methods is not verified under actual conditions of use. Specifically, Failure to ensure that the approved contract laboratory has performed method suitability of compendial methods USP <61> MICROBIOLOGICAL EXAMINATION OF NONSTERILE PRODUCTS: MICROBRIAL ENUMERATION TESTS and USP <62> MICROBIOLOGICAL EXAMINATION OF NONSTERILE PRODUCTS: TESTS FOR SPECIFIED MICROORGANISMS, used for the testing of Total Aerobic Microbial Count (TAMC), Total Yeast and Mold Count (TYMC), and tests for specified microorganisms such as Escherichia coli, Pseudomonas aeruginosa, Staphylococcus aureus, and Salmonella, respectively. For example, no method suitability was conducted for (b) (4) OTC Drug Products which have a (b) (4) (b) (4) in the formulation.
- Observation 5
Stability testing methods not validated
The written stability program for drug products does not include reliable, meaningful, and specific test methods. Specifically, Failure to ensure that the approved contract laboratory, which conducts stability testing for (b) (4) (b) (4) OTC Drug Products, has performed validation of the in-house method for Assay tests and has conducted forced degradation studies to demonstrate the Assay method is stability indicating.
- Observation 6
Insufficient computer and data integrity controls
Appropriate controls are not exercised over computers or related systems to assure that changes in master production and control records or other records are instituted only by authorized personnel. Specifically, 1) Electronic data stored within the (b) (4) computer used to operate the FTIR (b) (4) Software, used for identification testing of raw materials used in the manufacturing of OTC drug products, is not protected against possible deletion or modification. The raw data/spectra are saved on the (b) (4) after analysis; on 03/17/2026, we observed that the delete function on (b) (4) is not disabled, and we observed deleted spectra in the recycle bin. Additionally, the software does not have an audit trail to track the creation, modification, and deletion of data. 2) There is no assurance that all analytical data on the standalone (b) (4) pH meters, used to test raw materials, in-process samples, and finished OTC drug products are retained. On 03/17/2026, we observed a flashing error message on pH Meter # which stated "(b) (4) (b) (4)". The data from the pH meters are not backed up. In addition, results from the two pH Meters (# and #) are not printed out, and no second person verification is conducted at time of testing. Additionally, we observed that the "delete" option on the pH meters are not disabled. 3) The (b) (4) weigh balances (IDs: (b) (4)), used to conduct (b) (4) measurements for in-process bulk batches and finished OTC drug products, lack adequate controls to ensure complete and accurate data capture. The balances do not have the ability to save data electronically and no permanent record of the original instrument reading is maintained. All weight measurements are manually transcribed by an analyst onto (b) (4) and there are no procedures requiring second person verification of results at time of testing. 4) The (b) (4) (IDs: (b) (4)) used in (b) (4) testing of bulk batches used in finished OTC drug products, lack adequate controls to prevent loss and ensure data integrity. The (b) (4) do not have the ability to save data electronically and analysts can clear or delete test data. Additionally, all (b) (4) measurements are manually transcribed by an analyst onto CMS and there are no procedures requiring second person verification of results at time of testing.
- Observation 7
Process changes implemented without revalidation
Your firm failed to establish adequate written procedures for production and process controls designed to assure that the drug products have the identity, strength, purity, and quality that they are purported or represented to possess. Specifically. 1) Changes to manufacturing processes were implemented without undergoing a revalidation to ensure its reproducibility. For example, Customer Concern Communication Report (CCCR # 24-013) was initiated on 02/22/2024 when the customer expressed concerns with bulk manufactured material (b) (4) (b) (4) having a grainy appearance. An evaluation of finished goods and bulk retains was conducted and it was determined that samples showed some variance of grainy appearance. After a review of the validated process, a decision was made that a new process/re-phasing was necessary. The new phasing consisted of (b) (4) as opposed to the validated procedure which included the (b) (4) (b) (4). The change in manufacturing process was implemented on 04/26/2024 without a Change Control per SOP QA-002c "Change Control System" and without re-validation. The following batches of (b) (4) were manufactured and released after the change in manufacturing process: Lot #(b) (4) Expiration: (b) (4) Lot #(b) (4) Expiration: (b) (4) Lot #(b) (4) Expiration: (b) (4) Lot #(b) (4) Expiration: (b) (4) 2) The process validation (PV) for (b) (4), Formula (b) (4) (Protocol No: PV-006-2024, Approved: 12/26/2024) was not completed in accordance with procedure #VA-001a "Validation Master Plan" which requires a process to be validated using (b) (4) batches. (b) (4) bulk batches ((b) (4)) were manufactured; however, validation batch (b) (4) did not proceed to the filling and assembly stage. Additionally, the (b) (4) PV batches were not placed on stability studies.
- Observation 8
Deficient cleaning validation and procedures
Written procedures are not established and followed for the cleaning and maintenance of equipment, including utensils, used in the manufacture, processing, packing or holding of a drug product. Specifically, 1) Cleaning Validation (CV) conducted for (b) (4), Formula (b) (4) (Protocol No: PV-006-2024, Approved: 12/26/2024) is deficient for the following reasons: a. (b) (4) bulk batches ((b) (4)) were manufactured but only (b) (4) were filled and assembled. Validation batch #(b) (4) did not proceed to the filling and assembly stage; as a result, swab samples for detergent and API residue and microbial burden samples were not collected from product contact areas of the filling and assembly equipment for (b) (4) batches in accordance with the firm's procedure #VA-001a "Validation Master Plan" which requires CV to be conducted through (b) (4) product changeovers. b. During the CV, the following results from swab analyses which exceeded specification for residuals of actives (Specification: (b) (4)) were received and no investigations were conducted: [PV Batch No. | Batch No. | Location | Result (ppm/swab)] (b) (4) | (b) (4) | 1 | 14.66 (b) (4) | (b) (4) | 113.26 (b) (4) | (b) (4) | 48.08 (b) (4) | (b) (4) | 123.1 (b) (4) | (b) (4) | 116.78 (b) (4) | (b) (4) | 16.83 (b) (4) | (b) (4) | 75.30 (b) (4) | (b) (4) | 28.17 (b) (4) | (b) (4) | 128.20 (b) (4) | (b) (4) | 179.38 (b) (4) | (b) (4) | 392.28 (b) (4) | (b) (4) | 131.48 (b) (4) | (b) (4) | 130.69 (b) (4) | (b) (4) | 22.13 (b) (4) | (b) (4) | 141.21 (b) (4) | (b) (4) | 103.6 (b) (4) | (b) (4) | 11.68 Additionally, no swab samples were collected for detergent and API residue and microbial burden of product contact surfaces of the bulk manufacturing Kettle and (b) (4) Pump for batch (b) (4). c. The CV Protocol No: PV-006-2024 states "the equipment used during the manufacture, filling, and packaging...will be sampled ...at locations with product contact and perceived of being difficult to clean" areas; however, the protocol does not define the hardest-to-clean areas of the manufacturing Kettle and Filling Hopper and there is no documentation of what areas were swabbed during the CV. Additionally, SOP CS-009f "Cleaning and Sanitization of Manufacturing, Filling Equipment and Parts" prescribes contact time to be maintained with each cleaning and sanitizing agent; however, there is no documentation to show contact time for cleaning agents used during the CV. 2) Procedure SOP CS-009f "Cleaning and Sanitization of Manufacturing, Filling Equipment and Parts" requires the hopper and parts to (b) (4); additionally, the procedure requires all parts (b) (4). During a walkthrough on 03/17/2026, it was observed that there is no documentation to show contact time maintained with cleaning and sanitizing agents used as prescribed in the procedure during routine cleaning.
- Observation 9
Quality unit responsibilities and investigations lacking
The responsibilities and procedures applicable to the quality control unit are not in writing and fully followed. Specifically. 1) No investigation was initiated when the contract testing laboratory did not perform species identification on two Colony Forming Units of Gram-negative rods recovered on 08/07/2023 from (b) (4) of the (b) (4) (b) (4) system. 2) Retain samples were not assessed in response to a Customer Concern Communication Report (CCCR # 23-025) initiated on 10/04/2023 for (b) (4), Formula: (b) (4); the customer ((b) (4)) received complaints from clients of severe allergic reactions linked to lot numbers (b) (4) and (b) (4). The investigation failed to include a documented review of retain samples of the (b) (4) lots as required by SOP QA-034c "Customer Concern Communication (CCCR)" which states that preliminary investigations should include evaluations of retain samples. 3) No investigation was initiated to investigate low percentage yields for the following bulk batches of (b) (4) (b) (4) which were summarized as achieving the following yields in the Annual Product Review (APR) (Approved 04/23/2024): a. Batch (b) (4) (manufactured (b) (4)) achieved a 94.11% yield b. Batch (b) (4) (manufactured (b) (4)) achieved a 94.64% yield The APR states that the allowable theoretical yield for any given OTC drug product will be between (b) % and (b) %,
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Identification of Persons Performing Significant Steps
The batch production and control records are deficient in that they do not include identification of persons each significant step in the operation.
Sampling and testing plans not described
Sampling and testing plans for drug products are not described in written procedures which include the method of sampling and number of units per batch to be tested.
Control procedures to monitor and validate performance
Control procedures are not established which monitor the output and validate the performance of those manufacturing processes that may be responsible for causing variability in the characteristics of in-process material and the drug product.
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssuranceCitations not published
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssuranceCitations not published
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance13 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Quality control unit review of records
Drug product production and control records, are not reviewed and approved by the quality control unit to determine compliance with all established, approved written procedures before a batch is released or distributed.
Procedures not in writing, fully followed
The responsibilities and procedures applicable to the quality control unit are not in writing and fully followed.
Scientifically sound laboratory controls
Laboratory controls do not include the establishment of scientifically sound and appropriate test procedures designed to assure that drug products conform to appropriate standards of identity, strength, quality and purity.
Extent of discrepancy, failure investigations
Investigations of a failure of a batch or any of its components to meet any of its specifications did not extend to other batches of the same drug product and other drug products that may have been associated with the specific failure or discrepancy.
Labeling control records including specimens or copies
Batch production and control records do not include complete labeling control records, including specimens or copies of all labeling used for each batch of drug product produced.
Show 8 more citations
Investigations of discrepancies, failures
There is a failure to thoroughly review any unexplained discrepancy whether or not the batch has been already distributed.
Containers sampled so as to prevent contamination
Containers are not opened and sampled in a manner designed to prevent contamination of their contents.
Procedures for non-sterile drug products
Procedures designed to prevent objectionable microorganisms in drug products not required to be sterile are not established, written, and followed.
Reprocessing procedures not written or followed
Procedures prescribing a system for reprocessing batches to insure that the reprocessed batches will conform with all established standards, specifications, and characteristics are not written and followed.
Protective Apparel Not Worn
Protective apparel is not worn as necessary to protect drug products from contamination.
Reports of Analysis (Components)
Reports of analysis from component suppliers are accepted in lieu of testing each component for conformity with all appropriate written specifications, without performing at least one specific identity test on each component and establishing the reliability of the supplier's analyses through appropriate validation of the supplier's test results at appropriate intervals.
Buildings of Suitable Size, Construction, Location
Buildings used in the manufacture, processing, packing, or holding of a drug product do not have the suitable construction to facilitate cleaning, maintenance, and proper operations.
Lack of written stability program
There is no written testing program designed to assess the stability characteristics of drug products.
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Written record of investigation incomplete
Written records of investigations into the failure of a batch or any of its components to meet specifications do not include the conclusions and follow-up.
Reports of Analysis (Components)
Reports of analysis from component suppliers are accepted in lieu of testing each component for conformity with all appropriate written specifications, without establishing the reliability of the supplier's analyses through appropriate validation of the supplier's test results at appropriate intervals.
Extent of discrepancy, failure investigations
Investigations of a failure of a batch or any of its components to meet any of its specifications did not extend to other batches of the same drug product.
Quality control unit review of records
Drug product production and control records, are not reviewed and approved by the quality control unit to determine compliance with all established, approved written procedures before a batch is released or distributed.
Reprocessing procedures not written or followed
Procedures prescribing a system for reprocessing batches to insure that the reprocessed batches will conform with all established standards, specifications, and characteristics are not followed.
Show 1 more citation
Cleaning SOPs/equipment identification
Procedures for the cleaning and maintenance of equipment are deficient regarding the removal or obliteration of the previous batch identification.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Two requests to FDA under the Freedom of Information Act named this site, most recently on May 18, 2026. Each one asked for a Form 483.
Inspection ending Mar 27, 2026OAI
1 request. In the inspection history above. FDA has posted its Form 483.
- May 18, 2026FDA-FOIA-2026-5047Form 483· 1 of 5 sitesRecords released
Inspection ending Aug 17, 2022VAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Oct 24, 2024FDA-FOIA-2024-9438Form 483Closed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 13 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 6 inspections, with how many of them cited each.
- 21 CFR 211.84Testing and approval or rejection of components, drug product containers, and closures2 of 3
- 21 CFR 211.115Reprocessing2 of 3
- 21 CFR 211.188Batch production and control records2 of 3
- 21 CFR 211.192Production record review2 of 3
- 21 CFR 211.22Responsibilities of quality control unit1 of 3
- 21 CFR 211.28Personnel responsibilities1 of 3
- 21 CFR 211.42Design and construction features1 of 3
- 21 CFR 211.67Equipment cleaning and maintenance1 of 3
Show all 13
- 21 CFR 211.110Sampling and testing of in-process materials and drug products1 of 3
- 21 CFR 211.113Control of microbiological contamination1 of 3
- 21 CFR 211.160General requirements1 of 3
- 21 CFR 211.165Testing and release for distribution1 of 3
- 21 CFR 211.166Stability testing1 of 3
Observations read from posted 483 PDFs carry no CFR section of FDA’s, so they are not counted here.
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDER · DrugsLetter on fda.govFDA record
FDA drug establishment registration
FDA registration search- Registered operations
- Manufacturing
- Registration runs through
- Dec 31, 2027
- Registrant
- Bentley Laboratories, LLCD-U-N-S 068351753
From FDA’s Drug Establishments Current Registration Site, as of Nov 21, 2026. A registration says what a site told FDA it does. It is not an inspection result. FDA’s registration search looks firms up by name.
Other FEIs recorded as Bentley Laboratories LLC
Matched on the exact firm name FDA recorded. Sites recorded under other names are not listed, and a shared name does not show common ownership.
Bentley Laboratories LLCFEI 2222033Clifton, New Jersey, United States
Last inspected 2017-02-13NAI
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 2244819 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 2244819. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.