Skip to content

Fagron Analytical Services, LLC Form 483, March 2026

FDA issued this Form 483 to Fagron Analytical Services, LLC in Wichita, Kansas (FEI 3033031230) at the end of a drug inspection that ended March 26, 2026. It lists 1 observation. FDA classified the inspection Voluntary Action Indicated (VAI).

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: FDA's Electronic Reading Room, posted Oct 26, 2026. Form 483 PDF on fda.gov. FDAgent is independent of FDA.

Inspection ended
Mar 26, 2026
Drugs · Drug Quality Assurance
FDA's outcome
VAI
Voluntary Action Indicated
Posted by FDA
Oct 26, 2026
214 days after the inspection
Facility
FEI 3033031230
Wichita, Kansas

The 1 observation on this Form 483

Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent's.

  1. 1

    pH meter data backup and access controls

    Failure to maintain a backup file of data entered into the computer or related system. Specifically, The Mettler Toledo pH meter used in the chemistry laboratory creates digital data and results that are not backed up. The instrument can store up to 250 results, but there is no system in place to preserve this electronic data before it is overwritten or lost. Additionally, the pH meter does not have unique usernames and logins for individual operators. Instead, a shared (b) (4) username, "Analyst," is used by personnel. This shared login system does not allow for attribution of specific test results to individual operators and does not provide adequate controls to ensure that only authorized personnel are accessing and using the equipment. There is no assurance that pH test data generated by this equipment is complete, accurate, and attributable to specific individuals, or that historical data can be retrieved for batch record review or investigation purposes. The pH meter is used for release testing on the following products: Product (b) (4) (b) (4)

Ask FDAgent about this Form 483

FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
  • The observations are the text FDAgent read from this PDF by OCR, shown only when every printed observation passed FDAgent’s checks against the scanned text.
  • FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
  • FDAgent is independent of FDA.