Guam Memorial Hosp Authority
850 Gov. C. G. Camacho Rd, Tamuning, Guam, United States
FEI 1000523688
FDA has 4 inspections on file for Guam Memorial Hosp Authority in Tamuning, Guam (FEI 1000523688), from 2011 to 2026. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA posted the Form 483 from that inspection. FDAgent read its 3 observations from the PDF; they are listed below. Two FOIA requests have asked FDA for this inspection's 483.
No warning letter, court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Form 483 PDF ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 4, 2011–2026
- 0 OAI · 3 VAI · 1 NAI
- Inspections with 483 observations shown
- 2 of 4
- Form 483s posted by FDA
- ↗
- Requested under FOIA
- 2 requests, latest Aug 15, 2026
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood ProductsPosted 483 · 6 observations
Read this Form 483 on its own page →Form 483 PDF on fda.gov
Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent’s.
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
- Observation 1
Failure to obtain 8-week contact address
Failure to obtain from the donor on the day of donation a postal address where the donor may be contacted for 8 weeks after donation. Specifically, your firm has failed to obtain confirmation from donors on the day of donation regarding whether they will be at their current address for 8 weeks following donation. Neither your printed donor educational materials nor your Medical Questionnaire review process addresses verification of a postal address where prospective donors may be contacted during the 8-week period after donation. Additionally, SOP "Donor Interview and Processing" D105.02, Rev. 9, effective July 9, 2007, and SOP "Donor Notification of Abnormal Results" D117.01, effective January 5, 2000, do not include the requirement for 8-week postal address verification. Since 2023, your firm has had approximately (b) (4) registered donors and one (1) abnormal result.
- Observation 2
Failure to notify deferred hepatitis B donor
Failure to make reasonable attempts to notify a donor who has been deferred based on the results of tests for evidence of infection with a relevant transfusion-transmitted infection(s). Specifically, donor was tested positive for hepatitis B core Antibody on (b) (6), (b) (7)(C) and you failed to send a letter of notification within 8 weeks to inform donor with abnormal results. According to your Standard Operative Procedure for Donor Notification of Abnormal Result, January 05, 2000 D117.01, the Donor Interview card with abnormal result must be brought to the Blood Bank Medical Director for review. Attached letter of notification of abnormal results is sent to the donor. Receipt for certified mail will be attached to the letter and return receipt will follow. Letters returned to GMH Blood Bank because of change of address and/or foreign donors with incomplete or inaccurate addresses will be further investigated. Outside agencies may be relied on for this duty whenever possible if accepted by said agency. If all efforts fail for acquiring correct information or locating donor, handle as a (b) (4) and file appropriate.
Show 4 more observations
- Observation 3
SOPs for donor notification not followed
Written standard operating procedures including all steps to be followed in the collection and distribution of blood and blood components for allogeneic transfusion were not always followed. Specifically, donor tested positive for hepatitis B core antibody on (b) (6), (b) (7)(C), and you failed to follow Standard Operating Procedure for Donor Notification of Abnormal Results, D117.01, effective January 5, 2000, which requires: 1) The Donor Interview card with abnormal result must be brought to the Blood Bank Medical Director for review. Attached letter of notification of abnormal results is sent to the donor. Receipt for certified mail will be attached to the letter and return receipt will follow. 2) Letters returned to GMH Blood Bank because of change of address and/or foreign donors with incomplete or inaccurate addresses will be further investigated. Outside agencies may be relied on for this duty whenever possible if accepted by said agency. If all efforts fail for acquiring correct information or locating donor, handle as a (b) (4) and file appropriately. 3) Physician test results report can only be released to the Physician upon the donor's request. Be sure a copy of the release request is kept with donor card. Note date and obtain a donor signature on donor card to verify that results were received, as donor picks (b) (4) up letter and/or test results.
Establish, maintain and follow manufacturing SOPs
Written standard operating procedures including all steps to be followed in the collection and distribution of blood and blood components for allogeneic transfusion were not always followed.
Notification
Failure to make reasonable attempts to notify a donor who has been deferred based on the results of tests for evidence of infection with a relevant transfusion-transmitted infection(s).
Proof Of Identity
Failure to obtain from the donor on the day of donation a postal address where the donor may be contacted for 8 weeks after donation.
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood ProductsCitations not published
- NAINo objectionable conditions that need follow-upBiologics · Blood and Blood ProductsNone
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Establishment of spec., standards, and test procedures
Failure to establish scientifically sound and appropriate specifications, standards and test procedures to assure that blood and blood components are safe, pure, potent and effective.
Controlling storage temperatures
The standard operating procedure fails to include a written description of the storage temperatures and methods of controlling storage temperatures for all blood products and reagents
Who must report - manufacturer
Failure to submit biological deviation reports.
Labeling requirements for emergency shipment
Failure to include the statement, ?FOR EMERGENCY USE ONLY BY __________? on blood or blood components shipped in an emergency prior to completion of required tests.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Two requests to FDA under the Freedom of Information Act named this site, most recently on Aug 15, 2026. Each one asked for a Form 483.
Inspection ending Mar 6, 2026VAI
2 requests. In the inspection history above. FDA has posted its Form 483.
- Aug 15, 2026FDA-FOIA-2026-7903Form 483· 1 of 4 sitesRecords released
- May 27, 2026FDA-FOIA-2026-5295Form 483Records released
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 6 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 4 inspections, with how many of them cited each.
- 21 CFR 606.100Standard operating procedures2 of 2
- 21 CFR 600.14Reporting of biological product deviations by licensed manufacturers1 of 2
- 21 CFR 606.121Container label1 of 2
- 21 CFR 606.140Laboratory controls1 of 2
- 21 CFR 630.10General donor eligibility requirements1 of 2
- 21 CFR 630.40Requirements for notifying deferred donors1 of 2
Observations read from posted 483 PDFs carry no CFR section of FDA’s, so they are not counted here.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1000523688 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1000523688. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.