Skip to content

Fenwal International, Inc.

Rd 357, Km 0.8, Maricao, Puerto Rico, United States

FEI 2627511DUNS 091164590

Company: Fenwal, Inc. · Linked by exact D-U-N-S match (794519020) in FDA registration data

FDA has 12 inspections on file for Fenwal International, Inc. in Maricao, Puerto Rico (FEI 2627511), from 2009 to 2026. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 9 citations from this inspection, all under 21 CFR Part 211, the manufacturing rules for finished drugs. They are listed below. Two of the six FOIA requests for this site have asked FDA for this inspection's 483.

The most recent OAI inspection ended on Sep 24, 2021. FDA has issued 4 warning letters under this FEI, the latest on (CBER · Division of Biological Products Operations I). No court action or import refusal is filed under this FEI.

12 regulations were cited at more than one inspection; they are listed below.

As of Nov 21, 2026, the site is registered with FDA as a drug establishment for manufacturing, packing, labeling, analysis and sterilization, through Dec 31, 2026. Its registrant, Fenwal, Inc. (D-U-N-S 794519020), has registered 2 other sites.

Sources:FDA firm profile ↗Warning letter ↗FDA drug establishment registrations ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
12, 2009–2026
3 OAI · 4 VAI · 5 NAI
Inspections with 483 observations shown
10 of 12
Form 483s posted by FDA
↗
Requested under FOIA
6 requests, latest Jul 1, 2026
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Drug establishment registration
Through

Inspection history

12 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products9 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Buildings not maintained in good state of repair

      Buildings used in the processing of a drug product are not maintained in a good state of repair.

    2. Equipment construction - reactive surfaces

      Equipment surfaces that contact components and in-process materials are reactive, additive or absorptive so as to alter the safety, identity, strength, quality, or purity of the drug product beyond the official or other established requirements.

    3. Scientifically sound laboratory controls

      Laboratory controls do not include the establishment of scientifically sound and appropriate test procedures designed to assure that components, in-process materials and drug products conform to appropriate standards of identity, strength, quality and purity.

    4. Following/documenting laboratory controls

      Established test procedures and laboratory control mechanisms are not followed.

    5. Mfg / Processing Operations Area

      Separate or defined areas to prevent contamination or mix-ups are deficient regarding the manufacturing and processing operations.

    Show 4 more citations
    1. Authority lacking to review records, investigate errors

      The quality control unit lacks authority to fully investigate errors that have occurred.

    2. SOPs not followed / documented

      Written production and process control procedures are not followed in the execution of production and process control functions.

    3. Representative Number of Batches for Annual Review

      The procedures for the annual quality standards record evaluation are deficient in that they do not address a review of a representative number of approved and rejected batches.

    4. Absence of Written Procedures

      Your firm failed to establish adequate written procedures for production and process controls designed to assure that the drug products have the identity, strength, purity, and quality that they are purported or represented to possess.

  2. VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood ProductsPosted 483 · 7 citations

    Read this Form 483 on its own page →Form 483 PDF on fda.gov

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Equipment Design, Size and Location

      Equipment used in the manufacture, processing, packing or holding of drug products is not of appropriate design to facilitate operations for its intended use.

    2. Cleaning / Sanitizing / Maintenance

      Equipment and utensils are not maintained at appropriate intervals to prevent contamination that would alter the safety, identity, strength, quality or purity of the drug product.

    3. Records reviewed annually

      Records are not maintained so that data therein can be reviewed at least annually to evaluate the quality standards of each drug product to determine the need for changes in specifications or manufacturing or control procedures.

    4. Equipment for Environmental Control

      Equipment for adequate control over air pressure is not provided when appropriate for the manufacture, processing, packing or holding of a drug product.

    5. Washing and toilet facilities are deficient

      Washing and toilet facilities lack hot and cold water.

    Show 2 more citations
    1. Buildings not maintained in good state of repair

      Buildings used in the of a drug product are not maintained in a good state of repair.

    2. Scientifically sound laboratory controls

      Laboratory controls do not include the establishment of scientifically sound and appropriate test procedures designed to assure that in-process materials conform to appropriate standards of identity, strength, quality and purity.

  3. NAINo objectionable conditions that need follow-upBiologics · Vaccines and Allergenic Products9 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Environmental Monitoring System

      Aseptic processing areas are deficient regarding the system for monitoring environmental conditions.

    2. Absence of Written Procedures

      Your firm failed to establish written procedures for production and process controls designed to assure that the drug products have the identity, strength, purity, and quality that they are purported or represented to possess.

    3. Scientifically sound laboratory controls

      Laboratory controls do not include the establishment of scientifically sound and appropriate test procedures designed to assure that conform to appropriate standards of identity, strength, quality and purity.

    4. Equipment Design, Size and Location

      Equipment used in the manufacture, processing, packing or holding of drug products is not of appropriate design to facilitate operations for its.

    5. Testing and release for distribution

      Testing and release of drug product for distribution do not include appropriate laboratory determination of satisfactory conformance to the final specifications prior to release.

    Show 4 more citations
    1. Authority lacking to review records, investigate errors

      The quality control unit lacks authority to fully investigate errors that have occurred.

    2. Cleaning System

      Aseptic processing areas are deficient regarding the system for cleaning and disinfecting the room and equipment to produce aseptic conditions.

    3. Procedures for sterile drug products

      Procedures designed to prevent microbiological contamination of drug products purporting to be sterile are not established.

    4. Cleaning / Sanitizing / Maintenance

      Equipment and utensils are not cleaned and maintained at appropriate intervals to prevent contamination that would alter the safety, identity, strength, quality or purity of the drug product.

  4. OAISignificant objectionable conditions; FDA recommended regulatory actionBiologics · Blood and Blood Products4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Written record of investigation incomplete

      Written records of investigations into unexplained discrepancies do not always include the conclusions and follow-up.

    2. Procedures for sterile drug products

      Procedures designed to prevent microbiological contamination of drug products purporting to be sterile are not established and followed.

    3. Control procedures to monitor and validate performance

      Control procedures are not established which validate the performance of those manufacturing processes that may be responsible for causing variability in the characteristics of in-process material and the drug product.

    4. Complaint Investigation/Follow-Up Findings

      Complaint records are deficient in that they do not include the findings of the investigation and follow-up.

  5. OAISignificant objectionable conditions; FDA recommended regulatory actionBiologics · Blood and Blood Products6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Identification of containers, lines, equipment

      All compounding and storage containers used during the production of a batch of drug product is not properly identified at all times to indicate contents.

    2. Complete test data included in records

      Laboratory records do not include complete data derived from all tests, examinations and assay necessary to assure compliance with established specifications and standards.

    3. Procedures to be written and followed

      Procedures describing the handling of all written and oral complaints regarding a drug product are not followed.

    4. Objectionable microbiological contamination

      Each lot of a component liable to objectionable microbiological contamination is deficiently subjected to microbiological tests before use.

    5. Equipment Design, Size and Location

      Equipment used in the manufacture, processing, packing or holding of drug products is not of appropriate design to facilitate operations for its cleaning and maintenance.

    Show 1 more citation
    1. Procedures for sterile drug products

      Procedures designed to prevent microbiological contamination of drug products purporting to be sterile are not established.

  6. NAINo objectionable conditions that need follow-upDevices · Compliance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Absence of Written Procedures

      There are no written procedures for production and process controls designed to assure that the drug products have the identity, strength, quality, and purity they purport or are represented to possess.

    2. Investigations of discrepancies, failures

      There is a failure to thoroughly review any unexplained discrepancy whether or not the batch has been already distributed.

    3. Sanitation--buildings not clean, free of infestation

      Buildings used in the manufacture, processing, packing or holding of drug products are not free of infestation by rodents, birds insects, and other vermin.

    4. Failure to meet specifications

      An NDA-Field Alert Report was not submitted within three working days of receipt of information concerning a failure of one or more distributed batches of a drug to meet the specifications established for it in the application.

  7. NAINo objectionable conditions that need follow-upDevices · Compliance: Devices2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Cleaning / Sanitizing / Maintenance

      Equipment and utensils are not maintained at appropriate intervals to prevent contamination that would alter the safety, identity, strength, quality or purity of the drug product.

    2. Reserve samples identified, representative, stored

      Reserve drug product samples are not retained and stored under conditions consistent with product labeling.

  8. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  9. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices7 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Complaints reviewed by Quality Control Unit

      Written procedures describing the handling of complaints do not include provisions for a determination as to the need for an investigation of any unexplained discrepancy and explaining the reasons for the failure of the batch or any of its components to meet specifications.

    2. Control procedures to monitor and validate performance

      Control procedures are not established which monitor the output and validate the performance of those manufacturing processes that may be responsible for causing variability in the characteristics of in-process material and the drug product.

    3. Procedures not in writing, fully followed

      The responsibilities and procedures applicable to the quality control unit are not fully followed.

    4. Failure to meet specifications

      An NDA-Field Alert Report was not submitted within three working days of receipt of information concerning a failure of one or more distributed batches of a drug to meet the specifications established for it in the application.

    5. Procedures are written, and followed

      Procedures designed to assure that correct labels, labeling, and packaging materials are used for drug products are not written and followed.

    Show 2 more citations
    1. Written procedures not established/followed

      Written procedures are not established and followed for the cleaning and maintenance of equipment, including utensils, used in the manufacture, processing, packing or holding of a drug product.

    2. Prepared for each batch, include complete information

      Batch production and control records do not include complete information relating to the production and control of each batch.

  10. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Mix-up

      An NDA-Field Alert Report was not submitted within three working days of receipt of information concerning an incident that caused a drug product or its labeling to be mistaken for another article.

    2. Control procedures to monitor and validate performance

      Control procedures are not established which monitor the output of those manufacturing processes that may be responsible for causing variability in the characteristics of in-process material and the drug product.

    3. Extent of discrepancy, failure investigations

      Investigations of an unexplained discrepancy and a failure of a batch or any of its components to meet any of its specifications did not extend to other batches of the same drug product and other drug products that may have been associated with the specific failure or discrepancy.

  11. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Complaint Handling Procedure

      Procedures describing the handling of written and oral complaints related to drug products are deficiently written or followed.

    2. SOPs not followed / documented

      Written production and process control procedures are not followed in the execution of production and process control functions.

    3. Written record of investigation incomplete

      Written records of investigations into unexplained discrepancies and the failure of a batch or any of its components to meet specifications do not always include the conclusions and follow-up.

  12. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Six requests to FDA under the Freedom of Information Act named this site, most recently on Jul 1, 2026. Each one asked for a Form 483.

  • Inspection ending Feb 5, 2026VAI

    2 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on May 1, 2026. FDAgent holds no copy that FDA posted.

    1. Jul 1, 2026FDA-FOIA-2026-6267Form 483· 1 of 2 sitesRecords released
    2. Apr 10, 2026FDA-FOIA-2026-3642Form 483· 1 of 2 sitesRecords released
  • Inspection ending Sep 18, 2024VAI

    3 requests. In the inspection history above. FDA has posted its Form 483.

    1. May 5, 2026FDA-FOIA-2026-4424Form 483· 1 of 5 sitesRecords released
    2. Jul 22, 2025FDA-FOIA-2025-9057Form 483· 1 of 2 sitesClosed
    3. Dec 12, 2024FDA-FOIA-2024-10943Form 483· 1 of 2 sitesRecords released
  • Inspection date not given

    1 request. None of FDA’s log entries for these requests say records were released.

    1. Jun 8, 2026FDA-FOIA-2026-5631Form 483 + EIRClosed

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.

Regulations FDA cited here

The 24 sections of 21 CFR in FDA’s published citations for 10 of this facility’s 12 inspections, with how many of them cited each.

Show all 24

Warning letters and compliance actions

FDA Data Dashboard compliance actions

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched some of these, so the link opens FDA’s record of the action.

FDA drug establishment registration

FDA registration search
Registered operations
Manufacturing, packing, labeling, analysis, sterilization
Registration runs through
Dec 31, 2026
Registrant
Fenwal, Inc.D-U-N-S 794519020

From FDA’s Drug Establishments Current Registration Site, as of Nov 21, 2026. A registration says what a site told FDA it does. It is not an inspection result. FDA’s registration search looks firms up by name.

Other sites with the same registrant

FDA’s registration file names Fenwal, Inc. (D-U-N-S 794519020) as the registrant for this site and the 2 others below. That exact D-U-N-S number is the only link. A registrant is the firm that filed the registration, which is not always the owner.

Other FEIs recorded as Fenwal International, Inc.

Matched on the exact firm name FDA recorded. Sites recorded under other names are not listed, and a shared name does not show common ownership.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. This site’s records are filed under FEI 2627511 in FDA’s data; other sites under its registrant are listed in summary only. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every inspection and enforcement record on this page is filed under FEI 2627511. Other sites filed under the same registrant D-U-N-S are listed in summary only. Records filed under another FEI or name are otherwise not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.