Skip to content

SciVac Ltd

13 Gad Road, Rehovot, Israel

FEI 3012695367

FDA has one inspection on file for SciVac Ltd in Rehovot, Israel (FEI 3012695367). It ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA posted the Form 483 from that inspection. FDAgent read its 4 observations from the PDF; they are listed below. One FOIA request has asked FDA for this inspection's 483.

No warning letter, court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Form 483 PDF ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
1, in 2023
Inspections with 483 observations shown
1 of 1
Form 483s posted by FDA
↗
Requested under FOIA
1 request, latest Jul 8, 2025
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Vaccines and Allergenic ProductsPosted 483 · 4 observations

    Read this Form 483 on its own page →Form 483 PDF on fda.gov

    Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent’s.

    1. Observation 1

      Sterile contamination prevention procedures inadequate

      Procedures designed to prevent microbiological contamination of drug products purporting to be sterile are not established, written and followed. Specifically, A. You do not routinely employ the use of sporicidal agent(s) in your Grade filling environments [illegible] various spore forming microbes to include molds and bacillus species over the previous approximate two years. These spore forming recoveries are documented in your open (b)(4) i. You documented Deviation (b)(4) regarding the recovery of (b)(4) of filling operations inside the (b)(4) on (b)(4) sampled (b)(4) Your CAPA does not address (b)(4) door during batch (b)(4) operator error while the investigation includes operator error as a root cause. Sporicide use was not employed on the outside of the (b)(4) doors before filling operations were started for batch (b)(4) dated (b)(4) B. Your environmental and personnel monitoring plan to include sampling locations and times lack scientific support in your Grade [illegible] formulation and filling areas as follows: i. You do not perform personnel monitoring activities immediately after any of your aseptic connections. Personnel monitoring activities are only approximately (b)(4) performed at the end of each series of operations whereby operator(s) who perform aseptic connections sanitize their hands one or more times before personnel monitoring occurs. ii. You do not require aseptic operators who perform Grade interventions are tested to Grade personnel monitoring specifications, I observed your operator perform a routine (b)(4) (b)(4) intervention during batch (b)(4) whereby their head entered your (b)(4) who was later tested to Grade [illegible] specifications regarding head and mask testing. [illegible] are not representative of your dynamic environment during setup [illegible] ii. Your (b)(4) (b)(4) While you have a (b)(4) activities in your Grade filling areas located in room [illegible] area which is separated from your (b)(4) during setup activities, it is located in your (b)(4) filling area by a partition. You perform various setup activities in your filling area to include aseptic connections, filling head placement, etc. [illegible] does not occur proximal to the area where aseptic activities to include aseptic connections are performed. Instead, you surface sample near the (b)(4) which is approximately (b)(4) (b)(4) or more away from your aseptic connections. (b)(4) formulation area are not in or near your [illegible] located in your Grade [illegible] room v. Your (b)(4) work surface and therefore, are not representative of your dynamic / working environment. (b)(4) C. Areas and surfaces of your Grade [illegible] are not cleaned adequately. I observed aseptic open operators cleaning your open (b)(4) during filling operations and after batch (b)(4) was completed. The aseptic operator failed to completely clean the opened (b)(4) door(s) upon completion of an open door (b)(4) intervention. Operator(s) were also observed failing to clean the partition separating the (b)(4) area from the filling area inside your (b)(4)

    2. Observation 2

      Visual inspection training and defect kits deficient

      Employees engaged in the manufacture and processing of a drug product lack the training required to perform their assigned functions. Specifically, A. You lack scientific justification supporting your defect categories and establishment of your defect kits. Cracks are categorized with scratches and cracks are not included in your qualification kits. Furthermore, you have failed to include glass particles in your kits. i. AQLs are not scientifically established per defect or criticality. ii. Statistical analysis of defects by type is not scientifically established. You established trending limits of approximately (b)(4) by defect type when defect rates allow for tighter trending limits. An increase of defects over time was observed regarding [illegible]

    Show observations 3 to 4
    1. Observation 3

      Production deviations not recorded or investigated

      Deviations from written production and process control procedures are not recorded or justified. Specifically, [illegible] glove. A deviation was only opened 3/19/2023 after discussing this failure with your quality management. Media fill batch (b)(4) was filled during the use of this glove. B. You are not required to perform investigations upon failures regarding visual inspections and (b)(4) retraining efforts as well as gowning failures per written procedures. i. Your visual inspection operator failed multiple requalification efforts during the approximate period of 7/3/2022 through 7/28/2022 and no deviation was opened. ii. Your gowning qualification SOP does not require investigations if the operator fails (b)(4) gowning qualification.

    2. Observation 4

      Quality control unit procedures not followed

      The responsibilities and procedures applicable to the quality control unit are not in writing and fully followed. Specifically, A. Procedures are not established and / or followed to manage reconciliation of visual inspection personnel qualification documentation, media fill unit inspection documentation, etc. i. Your visual inspection qualification documentation does not reconcile all vials during operator qualifications. ii. Your media fill inspection vials do not reconcile across various documents included in media fill inspection records. B. Your retain inspection procedures do not review labeled and representative vials. Your retain inspection program include vials which are not labeled and you only inspect vials retained from each lot. Furthermore, you do not inspect your labeled vials.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

One request to FDA under the Freedom of Information Act named this site, on Jul 8, 2025. It asked for a Form 483.

  • Inspection ending Mar 27, 2023VAI

    1 request. In the inspection history above. FDA has posted its Form 483.

    1. Jul 8, 2025FDA-FOIA-2025-8360Form 483· 1 of 3 sitesRecords released

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3012695367 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3012695367. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.