ID Biomedical Corporation of Quebec Form 483, May 2022

FDA issued this Form 483 to ID Biomedical Corporation of Quebec in Quebec, Canada (FEI 3004835647) at the end of a biologics inspection that ended May 20, 2022. It lists 5 observations. FDA classified the inspection Voluntary Action Indicated (VAI).

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: FDA's Electronic Reading Room, posted Oct 26, 2026. Form 483 PDF on fda.gov. FDAgent is independent of FDA.

Inspection ended
May 20, 2022
Biologics · Vaccines and Allergenic Products
FDA's outcome
VAI
Voluntary Action Indicated
Posted by FDA
Oct 26, 2026
1,620 days after the inspection
Facility
FEI 3004835647
Quebec, Canada

The 5 observations on this Form 483

Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent's.

  1. 1

    Inadequate investigation of missing endotoxin data

    The following investigation was inadequate. Specifically, on April 25, 2022, event 200926036 was opened when the in-process endotoxin test data performed on 4/22/2022 could not be located on the (b)(4) (b)(4) (b)(4) The system administrator was contacted to retrieve a copy of the [illegible] There was no investigation in order to save the data onto the (b)(4) (b)(4) did not retain the endotoxin test data or a determination as to where the into why the (b)(4) data file went. The event was closed after a work order determined no error occurred with the (b)(4) (b)(4)

  2. 2

    Incomplete deficient syringe defect complaint records

    The following complaint records were found incomplete and/or deficient: a. Between 10/17/2017 and 12/21/2021, your company has received syringe defect complaints associated with multiple lots of FluLaval in which three separate investigations were opened. The complaints included needle connectivity issues to the luer lock, leaking from luer lock, crack at luer lock, and spinning of luer lock on the syringe. The first investigation concluded "incorrect manipulation of the syringe during preparation for injection"; second investigation concluded most probable root cause of "mishandling by the healthcare professional"; and the third was a most probable cause "the defect (LLA crack) was originated in the LLA suppliers process". No "internal root cause at GSK level". A crack in the LLA (luer lock adaptor) on the syringe was confirmed by the supplier for multiple lots of syringes, one of which was provided to your company. FDA was not notified with regards to the defective devices that were released to the market. b. Complaint files do not always include reply to complainant. For example, the complaint files for complaints 1500115512, 1500093176, and 1500115907 do not contain a reply to complainant.

  3. 3

    Inadequate sterile microbial contamination prevention procedures

    Procedures designed to prevent microbial contamination of drug products purporting to be sterile are inadequate. Specifically, (b)(4) styles worn by aseptic operators. The following deficiencies were noted for the (b)(4) worn by the operators working in the aseptic (b)(4): i. [illegible] with the (b)(4) (b)(4) There are (b)(4) on the [illegible] has been specifically evaluated by sterility approximately (b)(4) assurance for the impact to the aseptic environment. ii. The (b)(4) are (b)(4) with no maximum life span. The (b)(4) are wiped with a (b)(4) and are then [illegible] and then (b)(4) for (b)(4) (b)(4) (b)(4) iii. On 5/18/2022, a (b)(4) which had been previously wiped with (b)(4) and [illegible] were opened and (b)(4) for processing in the (b)(4) (b)(4) were observed on the (b)(4) (b)(4) (b)(4) b. The (b)(4) used within the Grade (b)(4) is not sterile. This (b)(4) is [illegible] a grade (b)(4) Particles from the (b)(4) have been recovered within the Grade (b)(4) on more than one occasion. [illegible] and can be used by the [illegible] c. There is a (b)(4) [illegible] the Grade (b)(4) room which (b)(4) operators working in aseptic (b)(4) This (b)(4) is wiped with (b)(4) This (b)(4) has not been [illegible]

  4. 4

    Incomplete preventative maintenance activities program

    Preventative maintenance activities were incomplete. Specifically, the (b)(4) in the (b)(4) on the (b)(4) are not part of a routine preventative maintenance program to include visual inspection. These (b)(4) (b)(4) are used during [illegible]

  5. 5

    Non-representative sampling procedure documented

    The sampling of the (b)(4) is not representative of use. Specifically, [illegible] manufacturing (b)(4) and this is documented on a log. After the (b)(4) (b)(4) Quality control takes a (b)(4) When (b)(4) QC takes the sample, (b)(4) (b)(4)

Ask FDAgent about this Form 483

FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
  • The observations are the text FDAgent read from this PDF by OCR, shown only when every printed observation passed FDAgent’s checks against the scanned text.
  • FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
  • FDAgent is independent of FDA.